ACCURACY OF IMPLANT TORQUE CONTROLLERS FOLLOWING TIME IN CLINICAL SERVICE

Pages
3–9
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Conclusion: The torque output of each individual device deviated in varying degrees from target torque values.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Background and Aim: The purpose of this study was to evaluate the accuracy of ten implant systems torque wrenches in delivering the target torque values following time in clinical service.
📊
Key Findings & Data:Calibration accuracy for the Tohnichi is ± 2% of the scale value.
💡
Conclusion & Impact:Conclusion: The torque output of each individual device deviated in varying degrees from target torque values.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Background and Aim: The purpose of this study was to evaluate the accuracy of ten implant systems torque wrenches in delivering the target torque values following time in clinical service. Materials and Methods: Eleven torque controllers from Nobel Biocare (Nobel Biocare Inc., Yorba Linda, CA), Camlog (Biotechnologies AG, Basel, Switzerland), Friadent (DentsplyFriadent, Mannheim, Germany), Straumann (Straumann AG, Basel, Switzerland), SwissPlus (Zimmer Dental, Carlsbad, CA), Astra Tech (Astra Tech Inc., Waltham, MA), Tapered screwvent (Zimmer Dental, Carlsbad, CA), Bio-Lok (International Inc., Deerfield Beach, FL), Anthogyr (Sallanches, France), BioHorizons, (Maestro Dental Implants, Birmingham, AL), Thommen (Thommen Medical AG, Waldenburg, Switzerland), which had been in clinical service for a maximum one year were tested. An implant, along with an attached abutment and abutment titanium screw, was held firmly in a Tohnichi torque gauge. Calibration accuracy for the Tohnichi is ± 2% of the scale value. Ten torque values (10, 14, 15, 20, 24, 25, 30, 32, 35, and 45 N-cm) were tested. The data for each group were evaluated using a statistical software package with a significance level of .05. Results: Significant differences generally existed between the individual units and the target torque values for all the tested devices except Friadent. The tested devices of the Friadent, Straumann, SwissPlus and Astra Tech implant systems were capable of providing consistent torque at or near to their respective targets although the torque output of each individual device deviated from target torque value in varying degrees. Conclusion: The torque output of each individual device deviated in varying degrees from target torque values.

🔖

Cite This Article

ERSU, B., TULUNOĞLU, İ., ŞENYILMAZ, D. P., MUHTAROĞULLARI, M. (2013). ACCURACY OF IMPLANT TORQUE CONTROLLERS FOLLOWING TIME IN CLINICAL SERVICE. Clinical Dentistry and Research, Cilt: 37 - Sayı: 1, 3–9. https://ajindex.com/articles/166108

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (13)

  1. Krennmair G, Seemann R, Schmidinger S, Ewers R, Piehslinger E. Clinical outcome of root-shaped dental implants of various diameters: 5-year results. Int J Oral Maxillofac Implants 2010; 25: 357-366
  2. Mitrani R, Nicholls JI, Phillips KM, Ma T. Accuracy of electronic implant torque controllers following time in clinical service. Int J Oral Maxillofac Implants 2001; 16: 394-399
  3. Jemt T, Laney WR, Harris D, Henry PJ, Krogh PH Jr, Polizzi G et al. Osseointegrated implants for single tooth replacement: a 1-year report from a multicenter prospective study. Int J Oral Maxillofac Implants 1991; 6: 29-36
  4. Dellinges M, Curtis D. Effects of infection control procedures on the accuracy of a new mechanical torque wrench system for implant restorations. J Prosthet Dent 1996; 75: 93-98
  5. Jemt T. Failures and complications in 391 consecutively inserted fixed prostheses supported by Brånemark implants in edentulous jaws: a study of treatment from the time of prosthesis placement to the first annual checkup. Int J Oral Maxillofac Implants 1991; 6: 270-276
  6. Burguete RL, Johns RB, King T, Patterson EA. Tightening characteristics for screwed joints in osseointegrated dental implants. J Prosthet Dent 1994; 71: 592-599
  7. Gutierrez J, Nicholls JI, Libman WJ, Butson TJ. Accuracy of the implant torque wrench following time in clinical service. Int J Prosthodont 1997; 10: 562-567
  8. Jörnéus L, Jemt T, Carlsson L. Loads and designs of screw joints for single crowns supported by osseointegrated implants. Int J Oral Maxillofac Implants 1992; 7: 353-359
  9. Standlee JP, Caputo AA, Chwu MY, Sun TT. Accuracy of mechanical torque-limiting devices for implants. Int J Oral Maxillofac Implants 2002; 17: 220-224
  10. Vallee MC, Conrad HJ, Basu S, Seong WJ. Accuracy of frictionstyle and spring-style mechanical torque limiting devices for dental implants. J Prosthet Dent 2008; 100: 86-92
  11. Cho SC, Small PN, Elian N. Screw loosening for standard and wide diameter implants in partially edentulous cases: 3- to 7-year longitudinal data. Implant Dent 2004; 13: 245-250 12. McGlumphy EA, Mendel DA, Holloway JA. Implant screw mechanics. Dent Clin North Am 1998; 42: 71-89
  12. Breeding LC, Dixon DL, Nelson EW, Tietge JD. Torque required to loosen single-tooth implant abutment screws before and after simulated function. Int J Prosthodont 1993; 6: 435-439
  13. Kwon JH, Han CH, Kim SJ, Chang JS. The change of rotational freedom following different insertion torques in three implant syst