Dergiler / Journal of Experimental and Clinical Medicine / 2026 / Volume: 43 Issue: 2

Late tibial post fracture following mobile-bearing posterior-stabilized total knee arthroplasty: A case report highlighting the role of malalignment and polyethylene wear

Taner Alıç(Hitit Üniversitesi Tıp Fakültesi Ortopedi ve Travmatoloji Ana Bilim Dalı)0000-0003-3848-8577
Yunus Erol Şahinci(HITIT UNIVERSITY, SCHOOL OF MEDICINE)0009-0002-8264-2818
Soner Gürel(HITIT UNIVERSITY, SCHOOL OF MEDICINE)0009-0005-2782-1401
Sayfa
—
Yayın Tarihi
2026-06-30
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Clinicians should consider polyethylene-related complications in patients presenting with late instability, even in the absence of a clearly defined traumatic event.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Tibial post fracture in mobile-bearing posterior-stabilized total knee arthroplasty (MB-PS TKA) is a rare but clinically significant complication associated with knee instability.
📊
Temel Bulgular & Veri:A 72-year-old female presented with progressive knee pain and instability eight years after primary MB-PS TKA.
💡
Sonuç & Katkı:Clinicians should consider polyethylene-related complications in patients presenting with late instability, even in the absence of a clearly defined traumatic event.
🔬Tür:Case Report
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:total knee arthroplasty · posterior-stabilized · mobile bearing polyethylene

Abstract

Tibial post fracture in mobile-bearing posterior-stabilized total knee arthroplasty (MB-PS TKA) is a rare but clinically significant complication associated with knee instability. A 72-year-old female presented with progressive knee pain and instability eight years after primary MB-PS TKA. Radiographic and clinical evaluation demonstrated varus malalignment and excessive posterior tibial slope. Revision TKA was performed, and intraoperative findings confirmed a fractured polyethylene tibial post with localized wear. Excessive posterior tibial slope may lead to anterior tibial translation and increased cam-post impingement, while varus malalignment results in asymmetric load distribution and accelerated polyethylene wear. Although mobile-bearing designs are intended to reduce stress concentration, these advantages may be negated under conditions of malalignment. Patient-related factors such as elevated body mass index and sustained physical activity may further contribute to implant failure. Proper component alignment and patient-specific risk assessment are essential to prevent tibial post fracture in MB-PS TKA. Clinicians should consider polyethylene-related complications in patients presenting with late instability, even in the absence of a clearly defined traumatic event.

🔖

Bu Makaleye Atıf Yap

Alıç, T., Şahinci, Y. E., Gürel, S. (2026). Late tibial post fracture following mobile-bearing posterior-stabilized total knee arthroplasty: A case report highlighting the role of malalignment and polyethylene wear. Journal of Experimental and Clinical Medicine, Volume: 43 Issue: 2. https://ajindex.com/articles/1422521

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (24)

  1. Cakmak MF, Horoz L. Effects of mobile- and fixed-bearing tibial inserts on clinical result of knee arthroplasty: a retrospective study. J Istanbul Fac Med. 2024;87(1):81-86.
  2. Bachmann M, Bolliger L, Ilchmann T, Clauss M. Long-term survival and radiological results of the Duracon total knee arthroplasty. Int Orthop. 2014;38(4):747-752.
  3. Callaghan JJ, Squire MW, Goetz DD, Sullivan PM, Johnston RC. Cemented rotating-platform total knee replacement: a nine to twelve-year follow-up study. J Bone Joint Surg Am. 2000;82(5):705-711.
  4. Tsushima T, Sasaki E, Sasaki S, et al. Mobile bearing shows larger rollback motion than fixed bearing in total knee arthroplasty using a medial stabilising technique with a navigation system. J Exp Orthop. 2024;11(3):e12053. doi:10.1002/jeo2.12053
  5. Puloski SK, McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB. Tibial post wear in posterior stabilized total knee arthroplasty: an unrecognized source of polyethylene debris. J Bone Joint Surg Am. 2001;83(3):390-397. doi:10.2106/00004623-200103000-00011
  6. Ueyama H, Fukunaga K, Takemura S, Koyanagi J, Yamamura M, Nakagawa S. Simultaneous post and cone fractures of a polyethylene insert in a mobile-bearing posterior-stabilized total knee arthroplasty. Arthroplast Today. 2024;26:101316. doi:10.1016/j.artd.2024.101316
  7. Kumar N, Yadav C, Raj R, Yadav S. Fracture of the polyethylene tibial post in a posterior stabilized knee prosthesis: a case report and review of literature. J Orthop. 2015;12:160.e3. doi:10.1016/j.jor.2015.01.002
  8. Diamond OJ, Howard L, Masri B. Five cases of tibial post fracture in posterior stabilized total knee arthroplasty using Prolong highly cross-linked polyethylene. Knee. 2018;25:657-662. doi:10.1016/j.knee.2018.05.005
  9. Sawaguchi N, Majima T, Ishigaki T, Mori N, Terashima T, Minami A. Mobile-bearing total knee arthroplasty improves patellar tracking and patellofemoral contact stress: in vivo measurements in the same patients. J Arthroplasty. 2010;25:920-925. doi:10.1016/j.arth.2009.07.024
  10. Zingde SM, Leszko F, Sharma A, Mahfouz MR, Komistek RD, Dennis DA. In vivo determination of cam-post engagement in fixed and mobile-bearing TKA. Clin Orthop Relat Res. 2014;472(1):254-262. doi:10.1007/s11999-013-3257-3
  11. Jeyalakshmi P, Ramkumar P. Failure mechanisms and influencing factors on orthopaedic knee implant failure: a comprehensive review. Proc Inst Mech Eng H. 2024. doi:10.1177/13506501241288733
  12. Kuwashima U, Hamai S, Okazaki K, et al. Contact stress analysis of the anterior tibial post in bi-cruciate stabilized and mobile-bearing posterior stabilized total knee arthroplasty designs. J Mech Behav Biomed Mater. 2016;60:460-467. doi:10.1016/j.jmbbm.2016.03.003
  13. Huang CH, Lu YC, Hsu LI, Liau JJ, Chang TK, Huang CH. Effect of material selection on tibial post stresses in posterior-stabilized knee prosthesis. Bone Joint Res. 2020;9(11):768-777. doi:10.1302/2046-3758.911.BJR-2020-0019.R2
  14. D’Angelo F, Marcolli D, Bulgheroni P, Murena L, Congiu T, Cherubino P. Two-stage fracture of a polyethylene post in a 9-year-old posterior-stabilized knee prosthesis: a case report. J Med Case Rep. 2010;4:65. doi:10.1186/1752-1947-4-65
  15. Bal BS, Greenberg D, Li S, Mauerhan D, Schultz L, Cherry K. Tibial post failures in a condylar posterior cruciate-substituting total knee arthroplasty. J Arthroplasty. 2008;23(5):650-655. doi:10.1016/j.arth.2007.08.002
  16. Lachiewicz PF. How to treat a tibial post fracture in total knee arthroplasty? A systematic review. Clin Orthop Relat Res. 2011;469:1709-1715. doi:10.1007/s11999-010-1609-9
  17. Mauerhan DR. Fracture of the polyethylene tibial post in a posterior cruciate-substituting total knee arthroplasty mimicking patellar clunk syndrome: a report of 5 cases. J Arthroplasty. 2003;18(7):942-945. doi:10.1016/s0883-5403(03)00333-4
  18. Lee CS, Chen WM, Kou HC, Lo WH, Chen CL. Early nontraumatic fracture of the polyethylene tibial post in a NexGen LPS-Flex posterior stabilized knee prosthesis. J Arthroplasty. 2009;24(8):1292.e5-1292.e9. doi:10.1016/j.arth.2009.02.013
  19. Mestha P, Shenava Y, D’Arcy JC. Fracture of the polyethylene tibial post in posterior stabilized (Insall Burstein II) total knee arthroplasty. J Arthroplasty. 2000;15(6):814-815. doi:10.1054/arth.2000.6615
  20. Mariconda M, Lotti G, Milano C. Fracture of posterior-stabilized tibial insert in a Genesis knee prosthesis. J Arthroplasty. 2000;15(4):529-530. doi:10.1054/arth.2000.4810
  21. Clarke HD, Math KR, Scuderi GR. Polyethylene post failure in posterior stabilized total knee arthroplasty. J Arthroplasty. 2004;19(5):652-657. doi:10.1016/j.arth.2004.02.026
  22. Chiu YS, Chen WM, Huang CK, Chiang CC, Chen TH. Fracture of the polyethylene tibial post in a NexGen posterior-stabilized knee prosthesis. J Arthroplasty. 2004;19(8):1045-1049. doi:10.1016/j.jarth.2004.04.013
  23. Hendel D, Garti A, Weisbort M. Fracture of the central polyethylene tibial spine in posterior stabilized total knee arthroplasty. J Arthroplasty. 2003;18(5):672-674. doi:10.1016/s0883-5403(03)00192-x
  24. Lim HC, Bae JH, Hwang JH, Kim SJ, Yoon JY. Fracture of a polyethylene tibial post in a Scorpio posterior-stabilized knee prosthesis. Clin Orthop Surg. 2009;1(2):118-121. doi:10.4055/cios.2009.1.2.118.