Dergiler / Turkish Journal of Chemistry / 2000 / Cilt: 24 - Sayı: 1

Specific Determination of Hydrogen Peroxide With A Catalase Biosensor Based on Mercury Thin Film Electrodes

Sayfa
95–99
Yayın Tarihi
2022-11-05
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

The peak current changes upon addition of hydrogen peroxide and gives a linear response in a concentration range of 1-50 x 10-6 M within a response time of 3 minutes.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:A biosensor for specific determination of hydrogen peroxide was developed using catalase (EC 1.11.1.6) in combination with a mercury film electrode.
📊
Temel Bulgular & Veri:The biosensor response was monitored by following the reduction peak of dissolved oxygen at around -0.24 V.
💡
Sonuç & Katkı:The peak current changes upon addition of hydrogen peroxide and gives a linear response in a concentration range of 1-50 x 10-6 M within a response time of 3 minutes.
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Turk. J. Chem. · (2000) · 95-100. Full text: pdf Other articles published in the same issue: Turk. J. Chem.

Özet

A biosensor for specific determination of hydrogen peroxide was developed using catalase (EC 1.11.1.6) in combination with a mercury film electrode. Catalase was immobilized with gelatine by means of glutaraldehyde on the electrode surface. The biosensor response was monitored by following the reduction peak of dissolved oxygen at around -0.24 V. The peak current changes upon addition of hydrogen peroxide and gives a linear response in a concentration range of 1-50 x 10-6 M within a response time of 3 minutes.

Abstract

A biosensor for specific determination of hydrogen peroxide was developed using catalase (EC 1.11.1.6) in combination with a mercury film electrode. Catalase was immobilized with gelatine by means of glutaraldehyde on the electrode surface. The biosensor response was monitored by following the reduction peak of dissolved oxygen at around -0.24 V. The peak current changes upon addition of hydrogen peroxide and gives a linear response in a concentration range of 1-50 x 10-6 M within a response time of 3 minutes.

🔖