Specific Determination of Hydrogen Peroxide With A Catalase Biosensor Based on Mercury Thin Film Electrodes

Pages
95–99
Publication Date
2022-11-05
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

The peak current changes upon addition of hydrogen peroxide and gives a linear response in a concentration range of 1-50 x 10-6 M within a response time of 3 minutes.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:A biosensor for specific determination of hydrogen peroxide was developed using catalase (EC 1.11.1.6) in combination with a mercury film electrode.
📊
Key Findings & Data:The biosensor response was monitored by following the reduction peak of dissolved oxygen at around -0.24 V.
💡
Conclusion & Impact:The peak current changes upon addition of hydrogen peroxide and gives a linear response in a concentration range of 1-50 x 10-6 M within a response time of 3 minutes.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Turk. J. Chem. · (2000) · 95-100. Full text: pdf Other articles published in the same issue: Turk. J. Chem.

Abstract

A biosensor for specific determination of hydrogen peroxide was developed using catalase (EC 1.11.1.6) in combination with a mercury film electrode. Catalase was immobilized with gelatine by means of glutaraldehyde on the electrode surface. The biosensor response was monitored by following the reduction peak of dissolved oxygen at around -0.24 V. The peak current changes upon addition of hydrogen peroxide and gives a linear response in a concentration range of 1-50 x 10-6 M within a response time of 3 minutes.

Özet

A biosensor for specific determination of hydrogen peroxide was developed using catalase (EC 1.11.1.6) in combination with a mercury film electrode. Catalase was immobilized with gelatine by means of glutaraldehyde on the electrode surface. The biosensor response was monitored by following the reduction peak of dissolved oxygen at around -0.24 V. The peak current changes upon addition of hydrogen peroxide and gives a linear response in a concentration range of 1-50 x 10-6 M within a response time of 3 minutes.

🔖