Dergiler / Turkish Neurosurgery / 2016 / Cilt: 26 - Sayı: 3

Construction of Finite Element Model for an Artificial Atlanto-Odontoid Joint Replacement and Analysis of Its Biomechanical Properties

Sayfa
430–436
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

The stress distribution in the AAOJ can be successfully analyzed with the FE method

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:AIM: To investigate the stress distribution on artificial atlantoaxial-odontoid joint (AAOJ) components during flexion, extension, lateral bending and rotation of AAOJ model constructed with the finite element (FE) method.
📊
Temel Bulgular & Veri:MATERIAL and METHODS: Human cadaver specimens of normal AAOJ were CT scanned with 1 mm -thickness and transferred into Mimics software to reconstruct the three-dimensional models of AAOJ.
💡
Sonuç & Katkı:The stress distribution in the AAOJ can be successfully analyzed with the FE method
🔬Tür:Modelleme & Sayısal Simülasyon
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Alan Parametreleri

Özet

AIM: To investigate the stress distribution on artificial atlantoaxial-odontoid joint (AAOJ) components during flexion, extension, lateral bending and rotation of AAOJ model constructed with the finite element (FE) method. MATERIAL and METHODS: Human cadaver specimens of normal AAOJ were CT scanned with 1 mm -thickness and transferred into Mimics software to reconstruct the three-dimensional models of AAOJ. These data were imported into Freeform software to place a AAOJ into a atlantoaxial model. With Ansys software, a geometric model of AAOJ was built. Perpendicular downward pressure of 40 N was applied to simulate gravity of a skull, then 1.53 N• m torque was exerted separately to simulate the range of motion of the model. RESULTS: An FE model of atlantoaxial joint after AAOJ replacement was constructed with a total of 103 053 units and 26 324 nodes. In flexion, extension, right lateral bending and right rotation, the AAOJ displacement was 1.109 mm, 3.31 mm, 0.528 mm, and 9.678 mm, respectively, and the range of motion was 1.6°, 5.1°, 4.6° and 22°. CONCLUSION: During all ROM, stress distribution of atlas-axis changed after AAOJ replacement indicating that AAOJ can offload stress. The stress distribution in the AAOJ can be successfully analyzed with the FE method

🔖

Bu Makaleye Atıf Yap

HU, Y., DONG, W., HANN, S., YUAN, Z., SUN, X., XIE, H., ZHANG, M. (2016). Construction of Finite Element Model for an Artificial Atlanto-Odontoid Joint Replacement and Analysis of Its Biomechanical Properties. Turkish Neurosurgery, Cilt: 26 - Sayı: 3, 430–436. https://ajindex.com/articles/404692

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (1)

  1. Ai FZ, Yin QS, Xu DC, Xia H, Wu ZH, Mai XH: Transoral atlantoaxial reduction plate internal fixation with transoral transpedicular or articular mass screw of C2 for the treatment of irreducible atlantoaxial dislocation: Two case reports. Spine 36: E556-562, 2011 2. Ashraf J, Crockard HA, Ransford AO, Stevens JM: Transoral decompression and posterior stabilisation in Morquio’s disease. Arch Dis Child 66:1318-1321, 1991 3. Bozkus H, Karakas A, Hancı M, Uzan M, Bozdag E, Sarioglu AC: Finite element model of the Jefferson fracture: Comparison with a cadaver model. Eur Spine J 10: 257-263, 2001 4. Brolin K, Halldin P: Development of a finite element model of the upper cervical spine and a parameter study of ligament characteristics. Spine: 29: 376-385, 2004 5. Crockard HA, Pozo JJ, Ransford AO, Stevens JM, Kendall BE, Essigman WK: Transoral decompression and posterior fusion for rheumatoid atlanto-axial subluxation. J Bone Joint Surgery BR 68: 350-356, 1986 6. del Palomar AP, Calvo B, Doblaré M: An accurate finite element model of the cervical spine under quasi-static loading. J Biomech 41: 523-531, 2008 7. Dickinson AS, Taylor AC, Ozturk H, Browne M: Experimental validation of a finite element model of the proximal femur using digital image correlation and a composite bone model. J Biomech Eng 133: 014504, 2011 8. Dickman CA, Sonntag VK: Posterior C1-C2 transarticular screw fixation for atlantoaxial arthrodesis. Neurosurgery 43: 275-280, 1998 9. Gebauer M, Goetzen N, Barvencik F, Beil FT, Rupprecht M, Rueger JM, Puschel K, Morlock M, Amling M: Biomechanical analysis of atlas fractures: A study on 40 human atlas specimens. Spine 33: 766-770, 2008 10. Guo X, Ni B, Zhao W, Wang M, Zhou F, Li S, Ren Z: Biomechanical assessment of bilateral C1 laminar hook and C1-2 transarticular screws and bone graft for atlantoaxial instability. J Spinal Disord Tech 22: 578-585, 2009 11. Hu Y, Yang SH, Xie H, Xu RM, Ma WH: Quantitative anatomic study of atlanto-odontoid joint and design of an artificial atlanto. Odontoid joint for the orthopedic clinic. Chin J Traumatol 10: 138-144, 2007 12. Hu Y, Yang S, Xie H, He X, Xu R, Ma W, Feng J, Chen Q: The anatomic study on replacement of artificial atlanto-odontoid joint through transoral approach. J Huazhong Univ Sci Technology Med Sci 28: 327-332, 2008 13. Hu Y, Gu Yj, He Xf, Xu RM, Ma WH, Zhao WD, Liang DZ, Zhong SZ: Biomechanical evaluation of stability and threedimensional movements of the atlantoaxial joint after artificial at