Evaluation of Bio-Rad D-10® and arkray adams HA-8160® HPLC analyzers in HbA1c measurement

Sayfa
953–956
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Both HPLC analyzers are shown to be reproducible and accurate for routine biochemistry laboratory use.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Glycosylated hemoglobin A1c (HbA1c) is an important parameter used for the assessment of time-dependent glycemic status and the diagnosis and follow-up of diabetes.
📊
Temel Bulgular & Veri:This study aimed to evaluate the analytical performance of two HPLC analyzers, the Bio-Rad D-10®, and the Arkray Adams HA-8160®.
💡
Sonuç & Katkı:Both HPLC analyzers are shown to be reproducible and accurate for routine biochemistry laboratory use.
🔬Tür:Araştırma Makalesi
📍Kapsam:105 samples
⚙️Değişkenler:Alan Parametreleri

Özet

Glycosylated hemoglobin A1c (HbA1c) is an important parameter used for the assessment of time-dependent glycemic status and the diagnosis and follow-up of diabetes. This study aimed to evaluate the analytical performance of two HPLC analyzers, the Bio-Rad D-10®, and the Arkray Adams HA-8160®. Accuracy and imprecision studies were conducted, and a method comparison study was performed with 105 samples. Samples were collected on five consecutive days and measured on both analyzers within two hours. Bland-Altman and regression analyses were used for statistical evaluation of the data. Low and high-level within-day CV values were calculated as 1.22% and 0.60% for the Arkray analyzer and 1.2% and 0.30% for the Bio-Rad analyzer, respectively. They were calculated as 1.27% and 2.52% for the Arkray analyzer and 2.32% and 3.44% for the Bio-Rad analyzer, respectively, between days. The within-day CV values for both analyzers were below the limit of 2.5% specified by the International Federation of Clinical Chemistry (IFCC). The bias average for three months was 3.2% for the Arkray analyzer and 1.2% for the Bio-Rad analyzer. The Spearman correlation coefficient (r) was 0.973 (p<0.0001) between the two measurements, and the slope and intercept values calculated in the regression analyses showed the linear relationship and harmony between the two analyzers [R2 = 0.967, (y = 0.860x + 0.604)]. Both HPLC analyzers are shown to be reproducible and accurate for routine biochemistry laboratory use.

🔖

Bu Makaleye Atıf Yap

YİS, Ö. M., PEHLİVAN, M. B., BUGDAYCİ, G., YUCE, N. (2019). Evaluation of Bio-Rad D-10® and arkray adams HA-8160® HPLC analyzers in HbA1c measurement. Medicine Science, Cilt: 8 - Sayı: 4, 953–956. https://ajindex.com/articles/276613

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (6)

  1. Misra A, Garg S. HbA1c and blood glucose for the diagnosis of diabetes. Lancet. 2011;378:104-6.
  2. Rifai N, Horvath AR, Wittwer CT. Tietz textbook of clinical chemistry and molecular diagnostics. In: Diabetes Mellitus. 6th edition. Elsevier, 2018;1160-94. Gillett MJ. International expert committee report on the role of the A1C Assay in the diagnosis of diabetes: diabetes care 2009;32:1327-34. American Diabetes Association. Diagnosis and classification of diabetes mellitus. Diabetes Care. 2014;37:81-90.
  3. Cohen H, Zohar J, Gidron Y, et al. Blunted HPA axis response to stress ınfluences susceptibility to posttraumatic stress response in rats. Biol Psychiatry. 2006;59:1208-18.
  4. Marshall SM, Barth JH. Standardization of HbA(1c) measurements: A consensus statement. Ann Clin Biochem. 2000;37:45-6. Miedema K. Laboratory tests in diagnosis and management of diabetes mellitus. Practical considerations. Clin Chem Lab Med. 2003;41:1259–65.
  5. Badiou S, Guillot J, Kuster N, et al. Comparison of Arkray/ELITech ADAMS HA-8180V® with Bio-Rad Variant,TMII Turbo2.0® and Tosoh Bioscience HLC®-723G8 for HbA1c Determination. J Clin Lab Anal. 2014;28:428-34.
  6. Goodall I, Colman PG, Schneider HG, et al. Desirable performance standards for HbA1c analysis - Precision, accuracy and standardisation: Consensus statement of the Australasian Association of Clinical Biochemists (AACB), the Australian Diabetes Society (ADS), the Royal College of Pathologists of A. Clin Chem Lab Med. 2007;45:1083-97. 10. Jones TG, Warber KD, Roberts BD. Analysis of Hemoglobin A1c from Dried Blood Spot Samples with the Tina-quant. J Diabetes Sci Technol. 2010;4:1-6. 11. Carey RN. CLSI EP15-A3: verification of precision and estimation of bias. https://www.westgard.com/clsi-ep15a3.htm 12. Madenci ÖÇ, Orçun A, Yildiz Z, et al. Evaluation of new Beckman Coulter 25(OH) Vitamin D assay and potential improvement of clinical interpretation. Biochem Medica. 2017;27:332-41. 13. Weykamp C, John WG, Mosca A. A review of the challenge in measuring hemoglobin A1c. J Diabetes Sci Technol. 2009;3:439-45. 14. Ejilemele A, Unabia J, Ju H, et al. A1c Gear: Laboratory quality HbA1c measurement at the point of care. Clin Chim Acta. 2015;445:139-42. 15. Shubrook J, Butts A, Chamberlain JJ, et al. Standards of medical care in diabetes—2017 abridged for primary care providers. Clin Diabetes. 2017;35:5–26. 16. World Health Organization. Use of Glycated Haemoglobin (HbA1c) in the Diagnosis of Diabetes Mellitus: Abbreviated Report of a WHO Con