Journals / Türk Oftalmoloji Dergisi / 2016 / Cilt: 46 - Sayı: 6

Changes in Anterior Chamber Depth after Phacoemulsification in Pseudoexfoliative Eyes and their Effect on Accuracy of Intraocular Lens Power Calculation

Psödoeksfolyatif Gözlerde Fakoemülsifikasyon Sonrası Ön Kamara Derinliğinde Değişiklikler ve Bunların Göz İçi Lens Gücü Hesaplamasının Tutarlılığına Etkisi

Pages
255–258
Publication Date
2022-11-05
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Results: The postoperative mean ACD values were significantly larger than the preoperative ACD values in both groups (p

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objectives: To compare anterior chamber depth (ACD) changes after phacoemulsification surgery in patients with pseudoexfoliation syndrome (PEX) and normal patients using an anterior segment imaging method.
📊
Key Findings & Data:Materials and Methods: Twenty-two eyes of 22 patients with PEX and 30 eyes of 30 normal patients who underwent uneventful phacoemulsification surgery and IOL implantation were included in the study.
💡
Conclusion & Impact:Results: The postoperative mean ACD values were significantly larger than the preoperative ACD values in both groups (p
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Objectives: To compare anterior chamber depth (ACD) changes after phacoemulsification surgery in patients with pseudoexfoliation syndrome (PEX) and normal patients using an anterior segment imaging method. Another aim of this study was to evaluate the effect of these changes on the accuracy of intraocular lens (IOL) power calculation and postoperative refraction. Materials and Methods: Twenty-two eyes of 22 patients with PEX and 30 eyes of 30 normal patients who underwent uneventful phacoemulsification surgery and IOL implantation were included in the study. The ACD of all patients was evaluated preoperatively and at 3 months postoperatively with the ALLEGRO Oculyzer (WaveLight® Oculyzer(TM) II, Alcon, Novartis)-Scheimpflug imaging system. Results: The postoperative mean ACD values were significantly larger than the preoperative ACD values in both groups (p

Özet

Amaç: Ön segment inceleme yöntemleri kullanarak psödoeksfoliyasyon sendromlu (PES) hastaların ve normal hastaların fakoemülsifikasyon cerrahisi sonrası ön kamara derinlik (ÖKD) değişimlerini karşılaştırmaktır. Buna ek olarak bu değişimin göz içi lens (GİL) gücü hesaplamasının tutarlılığına ve postoperatif refraksiyona etkisini incelemektir.Gereç ve Yöntem: Komplikasyonsuz fakoemülsifikasyon cerrahisi ve GİL implantasyonu yapılan 22 PES'li hastanın 22 gözü ve 30 normal hastanın 30 gözü çalışmaya alındı. Tüm hastaların ÖKD'si preoperatif ve postoperatif üçüncü ayda ALLEGRO Oculyzer (WaveLight® Oculyzer(TM) II, Alcon, Novartis) - Scheimpflug görüntüleme yöntemi ile incelendi.Bulgular: Her iki grupta postoperatif ortalama ÖKD değeri preoperatif ÖKD değerinden anlamlı olarak yüksekti (p

🔖

Cite This Article

ASENA, L., SEZENÖZ, A. S., GÜNGÖR, S. G., DEĞİŞİKLİKLER, Ö. K. D., CALCULATİON, P., AKMAN, A., AKSOY, M. (2022). Changes in Anterior Chamber Depth after Phacoemulsification in Pseudoexfoliative Eyes and their Effect on Accuracy of Intraocular Lens Power Calculation. Türk Oftalmoloji Dergisi, Cilt: 46 - Sayı: 6, 255–258. https://ajindex.com/articles/637937

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (10)

  1. Wang JK, Hu CY, Chang SW. Intraocular lens power calculation using the IOLMaster and various formulas in eyes with long axial length. J Cataract Refract Surg. 2008;34:262-267. Olsen T. Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 1992;18:125-129. Findl O. Biometry and intraocular lens power calculation. Curr Opin Ophthalmol. 2005;16:61-64.
  2. Cekic O, Batman C, Totan Y, Emre MI, Zilelioglu O. Changes in anterior chamber depth and intraocular pressure after phacoemulsification and posterior chamber intraocular lens implantation. Ophthalmic Surg Lasers. 1998;29:639-642. Norrby S. Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 2008;34:368-376. Erickson P. Effects of intraocular lens position errors on postoperative refractive error. J Cataract Refract Surg. 1990;16:305-311.
  3. Sanders DR, Higginbotham RW, Opatowsky IE, Confino J. Hyperopic shift in refraction associated with implantation of the single-piece Collamer intraocular lens. J Cataract Refract Surg. 2006;32:2110-2112.
  4. Pereira FA, Cronemberger S. Ultrasound biomicroscopic study of anterior segment changes after phacoemulsification and foldable intraocular lens implantation. Ophthalmology. 2003;110:1799-1806.
  5. Forsius H, Sveinsson K, Als E, Luukka H. Pseudoexfoliation of the lens capsule and depth of anterior chamber in northern Iceland. Acta Ophthalmol (Copenh). 1974;52:421-428. Bartholomew RS. Anterior chamber depth in eyes with pseudoexfoliation. Br J Ophthalmol. 1980;64:322-323.
  6. Jünemann A, Martus P, Handel A, Naumann GOH. Ocular dimensions in pseudoexfoliation syndrome. Ophthalmic Res. 1997;29:88. Gharagozloo NZ, Baker RH, Brubaker RF. Aqueous dynamics in exfoliation syndrome. Am J Ophthalmol. 1992;114:473-478.
  7. Doganay S, Tasar A, Cankaya C, Firat PG, Yologlu S. Evaluation of Pentacam- Scheimpflug imaging of anterior segment parameters in patients with pseudoexfoliation syndrome and pseudoexfoliative glaucoma. Clin Exp Optom. 2012;95:218-222.
  8. Rabsilber TM, Khoramnia R, Auffarth GU. Anterior chamber measurements using Pentacam rotating Scheimpflug camera. J Cataract Refract Surg. 2006;32:456-459.
  9. Ucakhan OO, Ozkan M, Kanpolat A. Anterior chamber parameters measured by the Pentacam CES after uneventful phacoemulsification in normotensive eyes. Acta Ophthalmol. 2009;87:544-548.
  10. Doganay S, Bozgul Firat P, Emre S, Yologlu S. Evaluation of anterior segment parameter changes using the Pentacam after uneventful phacoemulsification. Acta Ophtha