Journals / Türk Oftalmoloji Dergisi / 2020 / Cilt: 50 - Sayı: 5

Effect of Body Position on Intraocular Pressure Measured by Rebound Tonometer in Healthy Children

Sağlıklı Çocuklarda Vücut Pozisyonunun ReboundTonometre ile Ölçülen Göz İçi Basıncı Üzerine Etkisi

Pages
271–274
Publication Date
2022-11-06
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Conclusion: IOP values measured with the ICARE rebound tonometer in healthy children are not affected by body position

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objectives: To evaluate the effect of body position on intraocular pressure (IOP) measurement in the pediatric age group.
📊
Key Findings & Data:Forty-nine eyes of 49 pediatric patients were included in the study.
💡
Conclusion & Impact:Conclusion: IOP values measured with the ICARE rebound tonometer in healthy children are not affected by body position
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Objectives: To evaluate the effect of body position on intraocular pressure (IOP) measurement in the pediatric age group. Materials and Methods: Children whose general condition was healthy and ophthalmic examination was within normal limits wereincluded. Forty-nine eyes of 49 pediatric patients were included in the study. IOP was measured with an ICARE rebound tonometer(ICARE PRO; ICARE, Helsinki, Finland) while patients were in standing, sitting, and supine positions. Differences between theconsecutive measurements were compared statistically. Results: Twenty-two of the 49 patients were female, 27 were male. The mean age was 9.61±2.66 (5-15) years. Mean IOP values in thestanding, sitting, and supine positions were 18.81±2.97 (11.6-26.2) mmHg, 18.88±3.44, (12-28.2) mmHg, and 19.01±2.8 (13.5-25.9)mmHg, respectively. There were no statistically significant differences in pairwise comparisons of the measurements taken in the differentpositions (p=0.846, p=0.751, p=0.606). There was a statistically significant correlation between corneal thickness and intraocularpressure values in all measurements (p=0.001, r=0.516). Conclusion: IOP values measured with the ICARE rebound tonometer in healthy children are not affected by body position

Özet

Amaç: Pediatrik yaş grubunda vücut pozisyonunun göz içi basınç (GİB) ölçümü üzerine olan etkisini incelemek.Gereç ve Yöntem: Çalışmaya genel durumu sağlıklı ve oftalmik muayenesi normal sınırlarda olan pediatrik olgular dahil edildi. Çalışmaya 49 pediyatrik hastanın 49 gözü dahil edildi. GİB değerleri ayakta, otururken ve yatar pozisyonda ICARE rebound tonometre (ICARE PRO; ICARE, Helsinki, Finlandiya) ile ölçüldü. Müteakip ölçümler arasındaki farklılıklar istatistiksel olarak kıyas edildi.Bulgular: Çalışmaya 22’si kız, 27’si erkek 49 hasta dahil edildi. Ortalama yaş 9,61±2,66 (5-15) idi. Ortalama GİB ayakta iken 18,81±2,97 (11,6-26,2) mmHg, otururken 18,88±3,44 (12-28,2) mmHg ve yatar pozisyonda ise 19,01±2,8 (13,5-25,9) mmHg olarak ölçüldü. Ayakta, otururken ve yatar pozisyonda ölçülen müteakip GİB değerleri kıyaslandığında ölçümler arasında istatistiksel olarak anlamlı bir farklılık bulunmadı (p=0,846, p=0,751, p=0,606). Korneal kalınlık ve GİB değerleri arasında tüm ölçümlerde istatistiksel anlamlı korelasyon mevcuttu (p=0,001, r=0,516).Sonuç: Sağlıklı çocuklarda ICARE rebound tonometre ile ölçülen GİB değerleri vücut pozisyonundan etkilenmemektedir.Anahtar Kelimeler: Çocuk, göz içi basıncı, vücut pozisyonu, oküler tonometre

🔖

Cite This Article

ORUÇ, Y., UZLU, D., AKYOL, N., SOMUNCU, M., TÜRK, A., GÜRSOY, N. (2022). Effect of Body Position on Intraocular Pressure Measured by Rebound Tonometer in Healthy Children. Türk Oftalmoloji Dergisi, Cilt: 50 - Sayı: 5, 271–274. https://ajindex.com/articles/637646

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (14)

  1. Heijl A, Leske MC, Bengtsson B, Hyman L, Bengtsson B, Hussein M; Early Manifest Glaucoma Trial Group. Reduction of intraocular pressure and glaucoma progression: results from the Early Manifest Glaucoma Trial. Arch Ophthalmol. 2002;120:1268-1279
  2. Leske MC, Heijl A, Hussein M, Bengtsson B, Hyman L, Komaroff E; Early Manifest Glaucoma Trial Group. Factors for glaucoma progression and the effect of treatment: the early manifest glaucoma trial. Arch Ophthalmol. 2003;121:48-56
  3. Pakrou N, Gray T, Mills R, Landers J, Craig J. Clinical comparison of the ICare tonometer and Goldmann applanation tonometry. J Glaucoma. 2008;17:43- 47
  4. Brusini P, Salvetat ML, Zeppieri M, Tosoni C, Parisi L. Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients. J Glaucoma. 2006;15:213-217
  5. Iliev ME, Goldblum D, Katsoulis K, Amstutz C, Frueh B. Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness. Br J Ophthalmol. 2006;90:833-835
  6. Malihi M, Sit AJ. Effect of Head and Body Position on Intraocular Pressure. Ophthalmology. 2012;119:987-991
  7. Barkana Y. Postural change in intraocular pressure: comparison of measurement with a Goldmann tonometer, Tonopen XL, Pneumatonometer, and HA-2. J Glaucoma. 2014;23:23-28
  8. Krieglstein G, Langham ME. Influence of body position on the intraocular pressure of normal and glaucomatous eyes. Ophthalmologica. 1975;171:132- 145
  9. Lee JY, Yoo C, Jung JH, Hwang YH, Kim YY. The Effect of Lateral Decubitus Position on Intraocular Pressure in Healthy Young Subjects. Acta Ophthalmol. 2012;90:68-72
  10. Mansouri K, Weinreb RN, Liu JHK. Effects of aging on 24-hour intraocular pressure measurements in sitting and supine body positions. Invest Ophthalmol Vis Sci. 2012;53:112-116
  11. Barkana Y, Gutfreund S. Measurement of the difference in intraocular pressure between the sitting and lying body positions in healthy subjects: direct comparison of the Icare Pro with the Goldmann applanation tonometer, Pneumatonometer and Tonopen XL. Clin Exp Ophthalmol. 2014;42:608- 614
  12. De Bernardo M, Borrelli M, Cembalo G, Rosa N. Intraocular Pressure Measurements in Standing Position with a Rebound Tonometer. Medicina (Kaunas). 2019;55:701
  13. Mayalı H, Tekin B, Kayıkçıoğlu ÖR, Kurt E, İlker SS. Evaluation of the Effect of Body Position on Intraocular Pressure Measured with Rebound Tonometer. Turk J Ophthalmol. 2019;28:49:6-9
  14. Sultan M, Blondeau P. Episcleral