The Effects of Blood Collection Tubes on Folate Measurements

Kan Alma Tüplerinin Folat Ölçümlerine Etkisi

Pages
1–5
Publication Date
2022-11-05
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

The results were compared and p values

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objective: Blood collection tubes with a clot activator are widely used in clinical laboratories.
📊
Key Findings & Data:Material and Methods: We investigated the potential assay interference of three types of plastic tubes with a clot activator; Vacutest REF11030 (Group A); Becton Dickinson (BD) Vacutainer CAT REF367896 (Group B);and BD Vacutainer SSTTM II Advance REF367953 (with gel) (Group C) tubes.
💡
Conclusion & Impact:The results were compared and p values
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Objective: Blood collection tubes with a clot activator are widely used in clinical laboratories. Unexpectedly high results in folate assays obtained on Siemens Advia Centaur XP analyzer triggered a need for an investigation of the effects of blood collection tubes on folate measurements. Material and Methods: We investigated the potential assay interference of three types of plastic tubes with a clot activator; Vacutest REF11030 (Group A); Becton Dickinson (BD) Vacutainer CAT REF367896 (Group B);and BD Vacutainer SSTTM II Advance REF367953 (with gel) (Group C) tubes. Blood samples from apparently healthy 13 volunteers were drawn into these different tube types and serum folate concentrations were analyzed by Siemens Advia Centaur XP, Roche E-170, and Beckmann UniCel DxI800 immunoassay analyzers. The results were compared and p values

Özet

Amaç: Pıhtı aktivatörlü kan alma tüpleri klinik laboratuvarlarda yaygın olarak kullanılmaktadır. Siemens Advia Centaur XP cihazındaki sıra dışı yüksek folat sonuçları kan alma tüplerinin folat ölçümleri üzerine etkisini araştırmamıza sebep oldu. Gereç ve Yöntem: Pıhtı aktivatörlü üç plastik tüp tipi, potansiyel ölçüm interferansı açısından değerlendirildi; Vacutest REF11030 (Grup A), Becton Dickinson (BD) Vacutainer CAT REF367896 (Grup B) ve BD Vacutainer SSTTM II Advance REF367953 (jelli) (Grup C). Görünürde sağlıklı 13 bireyden farklı tüplere alınan kanlarda serum folat konsantrasyonları Siemens Advia Centaur XP, Roche E-170 ve Beckmann Unicell DXI800 cihazlarında ölçüldü ve sonuçlar karşılaştırıldı. İstatistiksel olarak p

🔖

Cite This Article

ERDEMLİ, H. K., KOCABAŞ, R., AKYOL, S. (2022). The Effects of Blood Collection Tubes on Folate Measurements. Türk Klinik Biyokimya Dergisi, Cilt: 14 - Sayı: 1, 1–5. https://ajindex.com/articles/634413

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (1)

  1. Bowen RA, Remaley AT. Interferences from blood collection tube components on clinical chemistry assays. Biochem Med (Zagreb). 2014;24(1):31-44 2. Stankovic AK, Parmar G. Assay interferences from blood collection tubes: a cautionary note. Clin Chem. 2006;52(8):1627-8. 3. Dimeski G. Interference Testing. ClinBiochem Rev. 2008; 29 Suppl (i): S43-8. 4. Bowen RA, Vu C, Remaley AT, Hortin GL, Csako G. Differential effect of blood collection tubes on total free fatty acids (FFA) and total triiodothyronine (TT3) concentration: a model for studying interference from tube constituents. ClinChimActa. 2007;378(1-2):181-93. 5. Yucel A, Karakus R, Cemalettin A. Effect of blood collection tube types on the measurement of human epidermal growth factor.J Immunoassay Immunochem. 2007;28(1):47-60. 6. Bowen RA, Chan Y, Cohen J, Rehak NN, Hortin GL, Csako Get al. Effect ofblood collection tubes on total triiodothyronine and other laboratory assays. Clin Chem. 2005;51(2):424-33. 7. Kroll MH, Elin RJ. Interference with clinical laboratory analyses. Clin Chem. 1994;40 (11 PT 1):1996-2005. 8. Boeynaems JM, De Leener A, Dessars B, Villa-Lobos HR, Aubry JC, Cotton Fet al. Evaluation of a new generation of plastic evacuated blood-collection tubes in clinical chemistry, therapeutic drug monitoring, hormone and trace metal analysis. Clin Chem Lab Med. 2004;42(1):67-71. 9. Bowen RA, Chan Y, Ruddel ME, Hortin GL, Csako G, Demosky SJ Jr et al. Immunoassay interference by a commonly used blood collection tube additive, the organosolicone surfactant silwet L-720. Clin Chem 2005;51(10): 1874–82. 10. Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36 (PT 6):704–21. 11. Wood WG, Gadow A. Immobolization of antibodies and antigens on macro solid phases—a comparison between adsorptive and covalent binding. A critical review of macro solid phases for use in immunoassay systems. Part I. J Clin Chem Clin Biochem 1983;21(12):789–97.