Detection of Brucella melitensis with BACTEC 9120 automated blood culture system

Otomatik kan kültür sistemi BACTEC 9120 ile Brucella melitensis'in saptanması

Pages
183–185
Publication Date
2022-10-17
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Mean time for detection was 59.30+3.96 hours at 5 cfu/ml and 47.37±15.85 hours at 50 cfu/ml concentration (p

📌 3 Key Takeaways

🎯
Objective & Hypothesis:The purpose of the study was to determine the optimum detection time for Brucella melitensis by BACTEC 9120 (Becton Dickinson Diagnostic Instrument Systems) blood culture system.
📊
Key Findings & Data:The study was performed with 12 B. melitensis strains isolated from blood cultures.
💡
Conclusion & Impact:Mean time for detection was 59.30+3.96 hours at 5 cfu/ml and 47.37±15.85 hours at 50 cfu/ml concentration (p
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

The purpose of the study was to determine the optimum detection time for Brucella melitensis by BACTEC 9120 (Becton Dickinson Diagnostic Instrument Systems) blood culture system. The study was performed with 12 B. melitensis strains isolated from blood cultures. A suspension of each strain was made in 5 ml of trypticase soy broth (TSB) and adjusted to approximately McFarland No. 1 standard. These suspensions consisting of 3x$10^8$ cfu/ml were then diluted serially with TSB to 3x$10^2$ cfu/ml. Each BACTEC 9120 aerobic bottle was inoculated with 5 ml of fresh blood collected from healthy volunteers. Dublicate bottles were then inoculated with 0.05 ml and 0.5 ml (15 cfu/ml and 150 cfu/ml) of 3x$10^2$ cfu/ml dilution. All bottles were transiently vented, inserted into the instrument, and monitored until they became positive. Mean time for detection was 59.30+3.96 hours at 5 cfu/ml and 47.37±15.85 hours at 50 cfu/ml concentration (p

Özet

Bu çalışmanın amacı, BACTEC 9120 (Becton Dickinson Diagnostic Instrument Systems) otomatik kan kültür sisteminin Brucella melitensis'i saptama süresinin bakteri konsantrasyonu ile ilişkisini saptamak idi. Bu amaçla, kan kültürlerinden izole edilen 12 B. melitensis suşu çalışma kapsamına alındı. Her bir suşun 5 ml triptikaz soy buyonda McFarland 1'e göre 3x$10^8$ koloni içeren süspansiyonu hazırlandı. Seri sulandırımlarla 3x$10^2$'lik süspansiyon elde edilerek buradan alınan 0.05 ve 0.5 ml(15 cfu/ml ve 150 cfu/ml) örnekler, sağlıklı gönüllülerden alman 5'er ml taze insan kanı ilave edilmiş olan BACTEC 9120 aerop kan kültürü şişelerine ekildi. Kan kültürü şişeleri pozitif sinyal verinceye kadar aygıtta inkübe edildi. Ortalama üreme zamanı 15 cfu/ml konsantrasyonda 59.30±3.96 saat, 150 cfu/ml konsantrasyonda ise 47.37±15.85 saat olarak saptandı (p

🔖

Cite This Article

ALTOPARLAK, Ü., ÖZKURT, Z., EROL, S. (2022). Detection of Brucella melitensis with BACTEC 9120 automated blood culture system. İnfeksiyon Dergisi, Cilt: 18 - Sayı: 2, 183–185. https://ajindex.com/articles/546542

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (13)

  1. Young EJ. Brucella species. In: Mandell GL, Douglas R, Bennett J, eds. Principles and Practice of Infectious Diseases. New York: Churchill Livingstone, 2000:2386-93
  2. Young EJ. An overview of human brucellosis. Clin Infect Dis 1995; 21: 283-90
  3. Hall WH, Khan MY. Brucellosis. In: Hoeprich PD, Jordan MC, eds. Infectious Disease. 4th ed. Philadelphia: JB Lippincott Co, 1989:1282-8
  4. Nolte FS, Williams JM, Jerris RC, et al. Multicenter clinic evaluation of a continuous monitoring blood culture system using a flourescent-sensor technology. J Clin Microbiol 1993; 29:287-90
  5. Hardy DJ, Hulbert BB, Migneault PC. Time of detection of Bact Alert blood cultures and lack of need for routine subculture of 5- to 7-day negative cultures. J Clin Microbiol 1992; 30:2743-5
  6. Gedikoğlu S, Helvacı S, Özakın F, Gokırmak F, Kılıçturgay K. Detection of Brucella melitensis by BACTEC NR 730 and BACTEC 9120 systems. Eur J Epidemiol 1996; 12: 649-50
  7. Durmaz G, Us T, Aydinli A, Kiremitçi A, Kiraz N, Akgun Y. Optimum detection times for bacteria and yeast species with the BACTEC 9120 aerobic blood culture system: evaluation for a 5-year period in a Turkish university hospital. J Clin Microbiol 2003; 41:819-21
  8. Ruiz J, Lorente I, Perez J, Simarro E, Martinez-Campos L. Diagnosis of brucellosis by using blood cultures. J Clin Microbiol 1997; 35: 2417-8
  9. Bannatyne MR, Jackson MC, Memish Z. Rapid diagnosis of Brucella bacteremia by using the BACTEC 9240 system. J Clin Microbiol 1997; 35: 2673-4
  10. Yagupsky P, Peled N, Press J, Abramson O, Abu-Rashid M. Comparison of BACTEC 9240 Peds Plus medium and isolator 1.5 microbial tube for detection of Brucella melitensis from blood cultures. J Clin Microbiol 1997; 35:1382-4
  11. Öztürk R, Mert A, Koçak F, et al. BACTEC 9240 kan kültür sistemiyle bruselloz tanısı. Flora 2002; 7:129-31
  12. Gamazo C, Vitas Al, Lopez-Goni I, Diaz R, Moriyon I. Factors affecting detection of Brucella melitensis by BACTEC NR730, a nonradiometric system for hemocultures. J Clin Microbiol 1993; 31:3200-3
  13. Zimmerman SJ, Gillikin S, Sofat N, Bartholomew WR, Amsterdam D. Case report and seeded blood culture study of Brucella bacteremia. J Clin Microbiol 1990; 28: 2139-41.