Journals / European Oral Research / 2018 / Cilt: 52 - Sayı: 2

Evaluation of deformation and fracture rates for nickel-titanium rotary instruments according to the frequency of clinical use

Pages
89–93
Publication Date
2018-05-01
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Results The overall rates of deformation and fracture were 10% and 1.2%, respectively.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:DOI: 10.26650/eor.2018.461Purpose To evaluate the deformation and fracture rates for ProTaper Universal (PTU) nickel-titanium rotary instruments according to the frequency of clinical use.
📊
Key Findings & Data:Materials and Methods A total of 619 PTU instruments (S1, S2, F1, F2, and F3) that have been used in the clinic by a single endodontist were collected over a period of 4 years.
💡
Conclusion & Impact:Results The overall rates of deformation and fracture were 10% and 1.2%, respectively.
🔬Design:Research Article
📍Scope:2018 Verileri
⚙️Variables:NiTi rotary system · ProTaper Universal · dental instruments

Abstract

DOI: 10.26650/eor.2018.461Purpose To evaluate the deformation and fracture rates for ProTaper Universal (PTU) nickel-titanium rotary instruments according to the frequency of clinical use. Materials and Methods A total of 619 PTU instruments (S1, S2, F1, F2, and F3) that have been used in the clinic by a single endodontist were collected over a period of 4 years. These instruments were grouped on the basis of one to three (Group A), four to six (Group B) and seven to nine (Group C) clinical uses (one canal = one use). All instruments were evaluated by a blinded investigator under a stereomicroscope at 15×–45× magnification for the presence of deformation and fracture. Results The overall rates of deformation and fracture were 10% and 1.2%, respectively. The deformation and fracture rates for the S2, F1, and F2 instruments showed no significant differences among groups. However, fracture rate for S1 instruments in Group A was significantly higher than for those in Group B (p=0.025) and Group C (p=0.004). In Group C, the S1 instruments showed a significantly higher deformation rate compared with the S2 (p=0.04), F1 (p=0.008) and F2 (p=0.049) instruments; there were no other significant differences within groups. Conclusion Under the conditions of the current study, frequency of use seemed to influence the deformation rates of PTU rotary instruments. Except S1, these instruments could be used without any fracture or deformation in up to 9 clinical cases by an experienced endodontist.

🔖

Cite This Article

YILMAZ, A., GÖKYAY, S. S., DAĞLAROĞLU, R., KÜÇÜKAY, İ. K. (2018). Evaluation of deformation and fracture rates for nickel-titanium rotary instruments according to the frequency of clinical use. European Oral Research, Cilt: 52 - Sayı: 2, 89–93. https://ajindex.com/articles/242993

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (16)

  1. Pruett JP, Clement DJ, Carnes DL, Jr. Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod 1997; 23: 77-85
  2. Shen Y, Peng B, Cheung GS. Factors associated with the removal of fractured niti instruments from root canal systems. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2004; 98: 605-10
  3. Sattapan B, Nervo GJ, Palamara JE, Messer HH. Defects in rotary nickel-titanium files after clinical use. J Endod 2000; 26: 161-5
  4. Shen Y, Cheung GS, Bian Z, Peng B. Comparison of defects in profile and protaper systems after clinical use. J Endod 2006; 32: 61-5
  5. Shen Y, Haapasalo M, Cheung GS, Peng B. Defects in nickel-titanium instruments after clinical use. Part 1: Relationship between observed imperfections and factors leading to such defects in a cohort study. J Endod 2009; 35: 129-32
  6. Vieira EP, Franca EC, Martins RC, Buono VT, Bahia MG. Influence of multiple clinical use on fatigue resistance of protaper rotary nickel-titanium instruments. Int Endod J 2008; 41: 163-72
  7. Gambarini G. Cyclic fatigue of profile rotary instruments after prolonged clinical use. Int Endod J 2001; 34: 386-9
  8. Peters OA, Barbakow F. Dynamic torque and apical forces of profile.04 rotary instruments during preparation of curved canals. Int Endod J 2002; 35: 379-89
  9. Wolcott S, Wolcott J, Ishley D, Kennedy W, Johnson S, Minnich S, Meyers J. Separation incidence of protaper rotary instruments: A large cohort clinical evaluation. J Endod 2006; 32: 1139-41
  10. Yared GM, Bou Dagher FE, Machtou P. Cyclic fatigue of profile rotary instruments after clinical use. Int Endod J 2000; 33: 204-7
  11. Yared GM, Dagher FE, Machtou P, Kulkarni GK. Influence of rotational speed, torque and operator proficiency on failure of greater taper files. Int Endod J 2002; 35: 7-12
  12. Hargreaves KM, Berman LH. Pathways of the pulp. 11th ed. Missouri, USA: Elsevier 2016, p.209-279
  13. Martin B, Zelada G, Varela P, Bahillo JG, Magan F, Ahn S, Rodriguez C. Factors influencing the fracture of nickel-titanium rotary instruments. Int Endod J 2003; 36: 262-6
  14. Zelada G, Varela P, Martin B, Bahillo JG, Magan F, Ahn S. The effect of rotational speed and the curvature of root canals on the breakage of rotary endodontic instruments. J Endod 2002; 28: 540-2
  15. Parashos P, Gordon I, Messer HH. Factors influencing defects of rotary nickel-titanium endodontic instruments after clinical use. J Endod 2004; 30: 722-5
  16. Ullmann CJ, Peters OA. Effect of cyclic fatigue on static f