Journals / Kocatepe Tıp Dergisi / 2019 / Cilt: 20 - Sayı: 3

BIOMECHANICAL COMPARISON OF DOUBLE AND TRIPLE TITANIUM ELASTIC NAIL FIXATION IN FEMUR FRACTURE MODELS

FEMUR KIRIK MODELLERİNDE İKİLİ VE ÜÇLÜ TİTANYUM ELASTİK ÇİVİ UYGULAMALARININ BİYOMEKANİK OLARAK KARŞILAŞTIRILMASI

Pages
109–114
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

In communited fracture models, the vertical forces results in 10 degree angulation were found to be 83.8N and 86.1N, respectively, in the double and triple TEN models (p=0.87).CONCLUSIONS: Although it was found that triple TEN models provided more stable fixation in transvers and communited fracture in terms of rotational forces, there was no significant difference between them in terms of the amount of displacement and angulation occurred against vertical forces.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:OBJECTIVE: The aim of this study is to investigate biomechanical stability of vertical and rotational forces on two retrograde titanium elastic nail (TEN) and three (two retrograde and one anterograde) TEN applications in pediatric femur diaphysis transverse and communited fracture models.MATERIAL AND METHODS: The transverse and fragmented fracture patterns formed in 24 synthetic femur models were fixed with double and triple TEN.
📊
Key Findings & Data:The fixation models were subjected to rotational and vertical forces and their stability was examined biomechanically.RESULTS: While the mean rotational forces that leads to 10 degree angulation of the transverse fracture pattern fixed with triple TEN models was 4.92N, it was 1.86N in the double TEN models (p=0.002).
💡
Conclusion & Impact:In communited fracture models, the vertical forces results in 10 degree angulation were found to be 83.8N and 86.1N, respectively, in the double and triple TEN models (p=0.87).CONCLUSIONS: Although it was found that triple TEN models provided more stable fixation in transvers and communited fracture in terms of rotational forces, there was no significant difference between them in terms of the amount of displacement and angulation occurred against vertical forces.
🔬Design:Modeling & Simulation
📍Scope:Comprehensive Sample
⚙️Variables:Biyomekanik çalışma · Titanyum elastik çivi · Femur

Abstract

OBJECTIVE: The aim of this study is to investigate biomechanical stability of vertical and rotational forces on two retrograde titanium elastic nail (TEN) and three (two retrograde and one anterograde) TEN applications in pediatric femur diaphysis transverse and communited fracture models.MATERIAL AND METHODS: The transverse and fragmented fracture patterns formed in 24 synthetic femur models were fixed with double and triple TEN. The fixation models were subjected to rotational and vertical forces and their stability was examined biomechanically.RESULTS: While the mean rotational forces that leads to 10 degree angulation of the transverse fracture pattern fixed with triple TEN models was 4.92N, it was 1.86N in the double TEN models (p=0.002). The rotational forces required for the 10-degree angulation in the communited fracture models fixed with triple and double TEN were found to be 3.43N and 1.77N, respectively (p = 0.002). The mean rotational forces that leads 5-mm displacement at the fracture site in transverse fracture patterns fixed with double TEN was 103.6N, on the other hand these mean forces was 135.7N in the triple TEN models. In communited fracture models, the vertical forces results in 10 degree angulation were found to be 83.8N and 86.1N, respectively, in the double and triple TEN models (p=0.87).CONCLUSIONS: Although it was found that triple TEN models provided more stable fixation in transvers and communited fracture in terms of rotational forces, there was no significant difference between them in terms of the amount of displacement and angulation occurred against vertical forces. It was concluded that a more stable fixation could be achieved in pediatric femur fractures by applying smaller in size but multiple TEN combined anterograde and retrograde technique.

Özet

AMAÇ: Bu çalışmanın amacı, pediatrik femur cisim transvers ve parçalı kırık modellerinde iki adet retrograt titanyum elastik çivi (TEÇ) uygulaması ile iki adet retrograt bir adet anterograt TEÇ uygulamasının, sentetik kemik modeller üzerinde vertikal ve rotasyonel kuvvetlere karşı biyomekanik stabilitelerinin araştırılmasıdır.

🔖

Cite This Article

DOĞAN, R. Ö., ÇALIŞKAN, E., ÇULCU, A., MİNARECİ, E. (2019). BIOMECHANICAL COMPARISON OF DOUBLE AND TRIPLE TITANIUM ELASTIC NAIL FIXATION IN FEMUR FRACTURE MODELS. Kocatepe Tıp Dergisi, Cilt: 20 - Sayı: 3, 109–114. https://ajindex.com/articles/205316

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (14)

  1. Biçimoğlu A, Çelebi L. Çocuk Femur Cisim Kırıkları. TOTBİD Dergisi 2006;5(2):34-43
  2. Flynn JM, Schwend RM. Management of pediatric femoral shaft fractures. J Am Acad Orthop Surg 2004;12:347-359
  3. Imam MA, Negida AS, Elgebaly A, et al. Titanium Elastic Nails Versus Spica Cast in Pediatric Femoral Shaft Fractures: A Systematic Review and Meta-analysis of 1012 Patients. Arch Bone Jt Surg. 2018;6(3):176-188
  4. Mohamed A, Rajeev AS. Clinical outcomes and complications of titanium versus stainless steel elastic nail in management of paediatric femoral fractures-a systematic review. Eur J Orthop Surg Traumatol. 2017;27(2):157-167
  5. Buckley SL. Current trends in the treatment of femoral shaft fractures in children and adolescents. Clin Orthop Relat Res. 1997;(338):60-73
  6. Gwyn DT, Olney BW, Dart BR, et al. Rotational control of various pediatric femur fractures stabilized with titanium elastic intramedullary nails. J Pediatr Orthop. 2004;24(2):172-7
  7. Heinrich SD, Drvaric DM, Darr K, et al. The operative stabilization of pediatric diaphyseal femur fractures with flexible intramedullary nails: a prospective analysis. J Pediatr Orthop. 1994;14(4):501-7. 114 8. Hinton RY, Lincoln A, Crockett MM, et al. Fractures of the femoral shaft in children. Incidence, mechanisms, and sociodemographic risk factors. J Bone Joint Surg Am. 1999;81(4):500-9
  8. Mazda K, Khairouni A, Penneçot GF, et al Closed flexible intramedullary nailing of the femoral shaft fractures in children. J Pediatr Orthop B. 1997;6(3):198-202
  9. Hughes BF, Sponseller PD, Thompson JD. Pediatric femur fractures: effects of spica cast treatment on family and community. J Pediatr Orthop. 1995;15(4):457-60
  10. Ligier JN, Metaizeau JP, Prévot J, et al. Elastic stable intramedullary nailing of femoral shaft fractures in children. J Bone Joint Surg Br. 1988;70(1):74-7
  11. Vrsansky P, Bourdelat D, Al Faour A. Flexible stable intramedullary pinning technique in the treatment of pediatric fractures. J Pediatr Orthop. 2000;20(1):23-7
  12. Huber RI, Keller HW, Huber PM, et al. Flexible intramedullary nailing as fracture treatment in children. J Pediatr Orthop. 1996;16(5):602-5
  13. Flynn JM, Luedtke LM, Ganley TJ, et al. Comparison of titanium elastic nails with traction and a spica cast to treat femoral fractures in children. J Bone Joint Surg Am. 2004;86-A(4):770-7
  14. Mann DC, Weddington J, Davenport K. Closed Ender nailing of femoral shaft fractures in adolescents. J Pediatr Orthop. 1986;6(6):651