Optimizations needed for lateral flow assay for rapid detection of pathogenic E. coli

Pages
954–968
Publication Date
2017-12-01
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Selectivity of the system was very high and the target was recognized by developed strips, regardless of its presence singly or in mixed bacterial samples.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Lateral flow assay (LFA), or the immunochromatographic strip test, is popular to use for rapid and sensitive immunoassays.
📊
Key Findings & Data:Since most enterohemorrhagic infections result from water and food contaminations of Escherichia coli O157:H7, selective and rapid detection of this organism in environmental and biological complexes is necessary.
💡
Conclusion & Impact:Selectivity of the system was very high and the target was recognized by developed strips, regardless of its presence singly or in mixed bacterial samples.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Gold nanoparticles · lateral flow assay · Escherichia coli detection

Abstract

Lateral flow assay (LFA), or the immunochromatographic strip test, is popular to use for rapid and sensitive immunoassays. Gold nanoparticles (GNPs), due to tunable optical characteristics and easy manipulation of size or shape, represent an attractive approach for LFA technology. Since most enterohemorrhagic infections result from water and food contaminations of Escherichia coli O157:H7, selective and rapid detection of this organism in environmental and biological complexes is necessary. In this study, optimized parameters of antibody (Ab)-based LFA for rapid detection of pathogenic E. coli O157:H7 are described. GNPs were used as visualizing agents. The measuring parameters include the Ab concentration on the capture lines, the concentration of gold conjugate, and flow rate. M180 and 36 nm were the ideal membrane and GNP size, respectively, for bacterial detection of LFA. The target, E. coli O157:H7, could be detected with a visual limit of detection of 105 cfu/mL in 3-5 min. Selectivity of the system was very high and the target was recognized by developed strips, regardless of its presence singly or in mixed bacterial samples.

🔖