AI Insight
⚡ TL;DR(Too Long; Didn't Read)

The prescription of eyeglasses should not be made without cycloplegic retinoscopy.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objectives: The purpose of the study was to compare the refractive error measurements of pediatric patients per-formed with a Plusoptix A09 photorefractor (PO; Plusoptix AG, Nuremberg, Germany), a Retinomax K-plus 3 (RTX; Right Group, Tokyo, Japan), and the new handheld auto refracto-keratometer, the Nidek HandyRef-K (HDY; Nidek SA, Créteil, France), and to evaluate the intermethods agreement.
📊
Key Findings & Data:Methods: A total of 194 eyes of 194 children were included in the study.
💡
Conclusion & Impact:The prescription of eyeglasses should not be made without cycloplegic retinoscopy.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

Objectives: The purpose of the study was to compare the refractive error measurements of pediatric patients per-formed with a Plusoptix A09 photorefractor (PO; Plusoptix AG, Nuremberg, Germany), a Retinomax K-plus 3 (RTX; Right Group, Tokyo, Japan), and the new handheld auto refracto-keratometer, the Nidek HandyRef-K (HDY; Nidek SA, Créteil, France), and to evaluate the intermethods agreement. Methods: A total of 194 eyes of 194 children were included in the study. All of the children underwent refraction mea-surement with the PO before cycloplegia and 2 autorefractors were used after cycloplegia: the RTX and the HDY. Results: The mean age of the patients was 16.65±10.04 months (range: 3-34 months). There were no statistically signifi-cant differences between the spherical values (SV) or cylindrical axis values (CAV) measured with the PO (SV: 1.61±1.79 diopters [D]; CAV: 94.25±72.47 D), the RTX (SV: 1.91±2.06 D; CAV: 94.3±73.44 D), and the HDY (SV: 1.89±2.04 D; CAV: 93.55±73.71 D) (p>0.05).There was a statistically significant difference in the cylindrical values (CV) assessed with the RTX (CV: -0.97±0.75 D) and the HDY (CV: -1.11±0.76 D) (p=0.003) and the HDY and the PO (CV: -0.92±0.68 D) (p=0.002), while there was no statistically significant difference between the values determined with the RTX and the PO (p>0.05). Statistically significant differences were demonstrated for spherical equivalent values (SEV) obtained with the RTX (SEV: 1.43±1.97 D) and the PO (SEV: 1.15±1.74 D) (p=0.02), and the HDY (SEV: 1.34±1.95 D) and the PO (p=0.03), but no significant difference was found between the RTX and the HDY values (p>0.05). Conclusion: No significant difference was found between the SEV measured by the RTX and the HDY, but the PO was significantly less hyperopic than the RTX and the HDY. The CV measured by HDY was higher than that of other devices. These devices can be used for screening in crowded pediatric ophthalmology clinics and may be an easier way of measuring refractive errors in children younger than 3 years of age, but high SEV and CV results should serve as an alert to physicians. It should also be kept in mind that cycloplegic retinoscopy is still the gold standard and these alternative methods can only be used for screening. The prescription of eyeglasses should not be made without cycloplegic retinoscopy.

🔖

Cite This Article

SEYMEN, Z., VURAL, E., ERİS, E., VURAL, A., ÖĞREDEN, T., ASLAN, Ö., ÇELEBİ, A. R. C., PERENTE, İ. (2019). HandyRef-K: Comparison of the Latest Handheld Auto Refracto-keratometer with Retinomax and Plusoptix in Patients Younger than Three Years of Age. BEYOGLU EYE JOURNAL, Cilt: 4 - Sayı: 1, 32–37. https://ajindex.com/articles/143739

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (10)

  1. Rotsos T, Grigoriou D, Kokkolaki A, Manios N. A comparison of manifest refractions, cycloplegic refractions and retinoscopy on the RMA-3000 autorefractometer in children aged 3 to 15 years. Clin Ophthalmol 2009;3:429–31.
  2. Williams C, Lumb R, Harvey I, Sparrow JM. Screening for re-fractive errors with the Topcon PR2000 Pediatric Refractome-ter. Invest Ophthalmol Vis Sci 2000;41:1031–7.
  3. Akil H, Keskin S, Çavdarli C. Comparison of the refractive mea-surements with hand-held autorefractometer, table-mounted autorefractometer and cycloplegic retinoscopy in children. Korean J Ophthalmol 2015;29:178–84.
  4. Moghaddam AA, Kargozar A, Zarei-Ghanavati M, Najjaran M, Nozari V, Shakeri MT. Screening for amblyopia risk factors in pre-verbal children using the Plusoptix photoscreener: a cross-sec-tional population-based study. Br J Ophthalmol 2012;96:83–6. Ruben J. Reimbursements and resources for pediatric vision screening. Am Orthop J 2006;56:54–61.
  5. Liang CL, Hung KS, Park N, Chan P, Juo SH. Comparison of measurements of refractive errors between the hand-held Retinomax and on-table autorefractors in cyclopleged and noncyclopleged children. Am J Ophthalmol 2003;136:1120–8.
  6. Oral Y, Gunaydin N, Ozgur O, Arsan AK, Oskan S. A compar-ison of different autorefractors with retinoscopy in children. J Pediatr Ophthalmol Strabismus 2012;49:370–7.
  7. Ying GS, Maguire M, Quinn G, Kulp MT, Cyert L; Vision In Preschoolers (VIP) Study Group. ROC analysis of the accuracy of Noncycloplegic retinoscopy, Retinomax Autorefractor, and SureSight Vision Screener for preschool vision screening. Invest Ophthalmol Vis Sci 2011;52:9658–64.
  8. Rajavi Z, Parsafar H, Ramezani A, Yaseri M. Is noncycloplegic photorefraction applicable for screening refractive amblyopia risk factors? J Ophthalmic Vis Res 2012;7:3–9.
  9. Huang D, Zhang X, Wang Y, Zhu H, Ding H, Bai J, et al. Pupil-lary measurements and anisocoria in Chinese preschoolers 3-4 years of age screened using the plusoptiX A12C. J AAPOS 2017;21:262.e1–262.e5. Payerols A, Eliaou C, Trezeguet V, Villain M, Daien V. Accuracy of PlusOptix A09 distance refraction in pediatric myopia and hyperopia. BMC Ophthalmol 2016;16:72. Harvey EM, Miller JM, Wagner LK, Dobson V. Reproducibility and accuracy of measurements with a hand held autorefractor in children. Br J Ophthalmol 1997;81:941–8.
  10. Farook M, Venkatramani J, Gazzard G, Cheng A, Tan D, Saw SM. Comparisons of the handheld autorefractor, table-mounted au-torefractor, and subjective refraction in Singapore adults.