Dergiler / Journal of Contemporary Medicine / 2021 / Cilt: 11 - Sayı: 1

Klinik Laboratuvarımızın Toplam Kalite Performansının Altı-Sigma Yöntemiyle Değerlendirilmesi

Evaluation of The Total Quality Performance of Our Clinical Laboratory With Six-Sigma Method

Sayfa
34–40
Yayın Tarihi
2021-01-30
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

An appropriate quality control scheduling should be carried out for each test by using Six-Sigma Methodology in all clinical chemistry laboratories.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Aim: In order to evaluate the performance of the clinical laboratories, a quality measurement method, The Six-Sigma Methodology, is used.
📊
Temel Bulgular & Veri:We aimed to evaluate the analytical performance of our clinical laboratory by using the internal and external quality control data of 28 analytes, and by calculating their process sigma values.
💡
Sonuç & Katkı:An appropriate quality control scheduling should be carried out for each test by using Six-Sigma Methodology in all clinical chemistry laboratories.
🔬Tür:Araştırma Makalesi
📍Kapsam:2018 Verileri
⚙️Değişkenler:Internal Quality Control · External Quality Control · Six Sigma

Abstract

Aim: In order to evaluate the performance of the clinical laboratories, a quality measurement method, The Six-Sigma Methodology, is used. We aimed to evaluate the analytical performance of our clinical laboratory by using the internal and external quality control data of 28 analytes, and by calculating their process sigma values. Methods: Sigma values of 28 analytes of our laboratory were determined from coefficient of variation (CV %) and bias for 3 subsequent months from December 2018 to January 2019. If the sigma values are >6, between 3 and 6, and <3, they are classified as »world-class«, »good« or »un-acceptable«, respectively. Results: 6 (21%) analytes; T Protein, Creatinin, Ca, Na, Cl, HbA1c; provided <3 sigma value, 1 (04%) analyte; lipase; provided between 3-6 value and 11 (39%) analytes; amylase, GGT, ALT, LDH, CRP, Lactate, BUN, Direct Bil., T Bil., CK and CK-MB; provided > 6 sigma value for both IQC1 and IQC2 levels for three consecutive months, respectively. Conclusions: In terms of Six-Sigma Metrics; our results were generally found as world-class or good. An appropriate quality control scheduling should be carried out for each test by using Six-Sigma Methodology in all clinical chemistry laboratories.

🔖

Bu Makaleye Atıf Yap

KATAR, M., DEVECİ, K., ÖZMEN, Z. C. (2021). Klinik Laboratuvarımızın Toplam Kalite Performansının Altı-Sigma Yöntemiyle Değerlendirilmesi. Journal of Contemporary Medicine, Cilt: 11 - Sayı: 1, 34–40. https://ajindex.com/articles/775062

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (16)

  1. Levey S, Jennings E. The use of control charts in the clinical laboratory. American Journal of Clinical Pathology. 1950;20(11_ts):1059-66
  2. Derleme Ç. Laboratuvar tıbbında altı-sigma kalite yönetimi. Türk Biyokimya Dergisi [Turkish Journal of Biochemistry-Turk J Biochem]. 2005;30(4):27
  3. Westgard JO. Six sigma quality design and control: Westgard QC, Incorporated; 2006
  4. Pande N. Cavanagh, The Six Sigma Way. McGraw Hill; 2000
  5. Verma M, Dahiya K, Ghalaut VS, Dhupper V. Assessment of quality control system by sigma metrics and quality goal index ratio: A roadmap towards preparation for NABL. World journal of methodology. 2018;8(3):44
  6. Litten J, Lynne N, Johnson K, Shih J. Evaluation of technopath controls on the architect family of instruments. Clin Chem Lab Med. 2015;53:S1-S1450
  7. Ercan M, BogdaycIoğlu N, Akbulut E, Oğuz E, Topçuoğlu C, YIlmaz F. ASSESSMENT OF THE ANALYTIC PERFORMANCE FOR IMMUNASSAY TESTS WITH SIX SIGMA METHODOLOGY: T425. Clinical Chemistry and Laboratory Medicine. 2015;53
  8. Nanda SK, Ray L. Quantitative application of sigma metrics in medical biochemistry. Journal of clinical and diagnostic research: JCDR. 2013;7(12):2689
  9. Westgard JO, Westgard SA. An assessment of σ metrics for analytic quality using performance data from proficiency testing surveys and the CLIA criteria for acceptable performance. Am J Clin Pathol. 2006;125(3):343-54
  10. Singh B, Goswami B, Gupta VK, Chawla R, Mallika V. Application of sigma metrics for the assessment of quality assurance in clinical biochemistry laboratory in India: a pilot study. Indian Journal of Clinical Biochemistry. 2011;26(2):131-5
  11. Garber C. Six Sigma: its role in the clinical laboratory. Clin Lab News. 2004;4:10-4
  12. Nar R, Emekli DI. The evaluation of analytical performance of immunoassay tests by using six-sigma method. Journal of medical biochemistry. 2017;36(4):301-8
  13. Gulbahar O, Arslan B, Sen B, Cingirt M, Yılmaz S. Evaluation of thyroid function tests in our laboratory by using six sigma analysis. Turkish Journal of Biochemistry. 2015;40(S1)
  14. Adiga US, Preethika A, Swathi K. Sigma metrics in clinical chemistry laboratory-A guide to quality control. Al Am een J Med Sci. 2015;8(4):281-7
  15. Nanda SK, Ray L. Quantitative application of sigma metrics in medical biochemistry. Journal of clinical and diagnostic research: JCDR. 2013;7(12):2689-91
  16. Chaudhary NG, Patani SS, Sharma H, Maheshwari A, Jadhav PM, Maniar MA. Application of six sigma for the qual