Dergiler / The European Research Journal / 2019 / Cilt: 5 - Sayı: 1

Wrongs known as right in thyroid scintigraphy and uptake study

Sayfa
142–147
Yayın Tarihi
2022-10-18
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

The resolutiondifferences between the images were evaluated.Conclusion: It was determined that imaging using 10 cm object-to-pinhole distanceand 4 mm diameter pinhole collimator was best in terms of image resolution and optimum 99mTc-pertechnetate uptake level.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Objectives: Thyroid scintigraphy using 99mTc-pertechnetate is commonly used to study function and structureof thyroid gland.
📊
Temel Bulgular & Veri:Thepurpose of the present study was to determine actual radiopharmaceutical uptake value in an experimental99mTc-pertechnetate thyroid scintigraphy and uptake model.Methods: Thyroid hyperactive and hypoactive nodule models were created using 4 mCi (148 MBq) 99mTcpertechnetate.
💡
Sonuç & Katkı:The resolutiondifferences between the images were evaluated.Conclusion: It was determined that imaging using 10 cm object-to-pinhole distanceand 4 mm diameter pinhole collimator was best in terms of image resolution and optimum 99mTc-pertechnetate uptake level.
🔬Tür:Modelleme & Sayısal Simülasyon
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Alan Parametreleri

Abstract

Objectives: Thyroid scintigraphy using 99mTc-pertechnetate is commonly used to study function and structureof thyroid gland. Pin-hole collimator is generally preferred in thyroid scintigraphy and uptake studies. Thepurpose of the present study was to determine actual radiopharmaceutical uptake value in an experimental99mTc-pertechnetate thyroid scintigraphy and uptake model.Methods: Thyroid hyperactive and hypoactive nodule models were created using 4 mCi (148 MBq) 99mTcpertechnetate. In the experimental model, 4 mm, 6 mm and 8 mm diameter pin-hole collimators, and 5 cm, 7cm and 10 cm object-to-pinhole distances were investigated.Results: In thyroid hyperactive nodule model, despite the same activity value, uptake at 7 cm object-to-pinholedistance was higher compared to 10 cm distance (122% and 103%, respectively). In the patient with Graves’disease, despite the same activity value, uptake at 5 cm object-to-pinhole distance was higher compared to 10cm distance (8% and 4%, respectively). In thyroid hypoactive nodule model, 4 mm, 6 mm and 8 mm diameterspin-hole collimators were imaged at 5 cm, 10 cm and 15 cm object-to-pinhole distances. The resolutiondifferences between the images were evaluated.Conclusion: It was determined that imaging using 10 cm object-to-pinhole distanceand 4 mm diameter pinhole collimator was best in terms of image resolution and optimum 99mTc-pertechnetate uptake level.

🔖

Bu Makaleye Atıf Yap

Gül, S. S. (2022). Wrongs known as right in thyroid scintigraphy and uptake study. The European Research Journal, Cilt: 5 - Sayı: 1, 142–147. https://ajindex.com/articles/597561

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (16)

  1. Higgins HP, Ball D, Eastham S. 20-Min 99mTc thyroid uptake: a simplified method using the gamma camera. J Nucl Med 1973;14:907-11
  2. Schneider PB. Simple, rapid thyroid function testing with 99mTc-pertechnetate thyroid uptake ratio and neck/thigh ratio. Am J Roentgenol 1979;132:249-353
  3. Becker D, Charkes ND, Dworkin H, Hurley J, McDougall IR, Price D, et al. Procedure guideline for thyroid uptake measurement: 1.0. Society of Nuclear Medicine. J Nucl Med 1996;37:1266-8
  4. Ziessman HA, Fahey FH, Gochoco JM. Impact of radiocontaminants in commercially available iodine-123: dosimetric evaluation. J Nucl Med 1986;27:428-32
  5. Ramos CD, Zantut-Wittmann DE, Tambascia MA, Assumpcao L, Etchebehere EC, Camargo EE. Thyroid suppression test with L-thyroxine and [99mTc] pertechnetate. Clin Endocrinol (Oxf) 2000;52:471-7
  6. Anjos DA, Etchebehere EC, Santos AO, Lima MC, Ramos CD, Paula RB, et al. Normal values of [99mTc] pertechnetate uptake and excretion fraction by major salivary glands. Nucl Med Commun 2006;27:395-403
  7. Smith JJ, Croft BY, Brookeman VA, Teates CD. Estimation of 24-hour thyroid uptake of I-131 sodium iodide using a 5- minute uptake of technetium-99m pertechnetate. Clin Nucl Med 1990;15:80-3
  8. Gullberg GT, Zeng GL, Datz FL, Christian PE, Tung CH, Morgan HT. Review of convergent beam tomography in single photon emission computed tomography. Phys Med Biol 1992;37:507-34
  9. Anger K, Feine U, Kaiser E. Improved quality of thyroid scintigrams by using a gamma camera with pin-hole collimator. Rofo 1981;135:188-92
  10. Tai YC, Wu H, Pal D, O’Sullivan JA. Virtual-Pinhole PET. J Nucl Med 2008;49:471-9
  11. Selby JB, Buse MG, Gooneratne NS, Moore DO. The Anger camera and the pertechnetate ion in the routine evaluation of thyroid uptake and imaging. Clin Nucl Med 1979;4:233-7
  12. Volckaert V, Vandermeulen E, Duchateau L, Saunders JH, Peremans K. Inter- and intraobserver variability of (semi-) quantitative parameters commonly used in feline thyroid scintigraphy. Res Vet Sci 2016;105:87-91
  13. Ramos CD, Zantut Wittmann DE, Etchebehere EC, Tambascia MA, Silva CA, Camargo EE. Thyroid uptake and scintigraphy using 99mTc pertechnetate: standardization in normal individuals. Sao Paulo Med J 2002;120:45-8
  14. Sarkar SD. Benign thyroid disease: what is the role of nuclear medicine? Semin Nucl Med 2006;36:185-93
  15. Clerc J. Imaging the thyroid in children. Best Pract Res Clin Endocrinol Metab 2014;28:203-20
  16. Biassoni L, Easty M. Paedi