Comparison of cyclic fatigue resistance of different and novel heat-treated nickel-titanium rotary file systems at the intracanal temperature

Sayfa
158–164
Yayın Tarihi
2022-10-19
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Novel EndoArt Blue files exhibited significantly greater cyclic fatigue resistance than the other NiTi files.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Aim: The aim of this study was to compare the cyclic fatigue resistance of different heat-treated nickel-titanium rotary systems at intracanal temperature.
📊
Temel Bulgular & Veri:Methodology: A total of 90 OneCurve (Micro-Mega, Besançon, France), VDW.ROTATE (VDW Dental, Munich, Germany), Typhoon (Clinician’s Choice, New Milford, CT, USA), HyFlex EDM (Coltene/Whaledent AG, Altstatten, Switzerland), and EndoArt Gold and Blue (Inci Dental, Istanbul, Turkey) (n = 15) rotary files (#25/0.06) were tested at intracanal temperature (35.5 ℃) using a dynamic model in a stainless-steel artificial canal with an inner diameter of 1.5 mm, 60° angle of curvature, and 2 mm radius of curvature.
💡
Sonuç & Katkı:Novel EndoArt Blue files exhibited significantly greater cyclic fatigue resistance than the other NiTi files.
🔬Tür:Modelleme & Sayısal Simülasyon
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Alan Parametreleri

Abstract

Aim: The aim of this study was to compare the cyclic fatigue resistance of different heat-treated nickel-titanium rotary systems at intracanal temperature. Methodology: A total of 90 OneCurve (Micro-Mega, Besançon, France), VDW.ROTATE (VDW Dental, Munich, Germany), Typhoon (Clinician’s Choice, New Milford, CT, USA), HyFlex EDM (Coltene/Whaledent AG, Altstatten, Switzerland), and EndoArt Gold and Blue (Inci Dental, Istanbul, Turkey) (n = 15) rotary files (#25/0.06) were tested at intracanal temperature (35.5 ℃) using a dynamic model in a stainless-steel artificial canal with an inner diameter of 1.5 mm, 60° angle of curvature, and 2 mm radius of curvature. Testing was conducted until fracturing, at which time the device stopped automatically, and the number of rotations was calculated as seconds. Lengths of fractured parts were measured using a digital caliper. One-way ANOVA test followed by Tukey’s test was used to compare the groups. Scanning electron microscopic evaluation was performed to confirm the types of fracture. Results: EndoArt Blue group had a significantly higher mean time to fracture in all groups, followed by the HyFlex EDM, VDW.ROTATE, OneCurve, EndoArt Gold, and Typhoon. In addition, the HyFlex EDM and VDW.ROTATE groups had no significant differences between each other and were significantly better than the others. No significant differences were found between the OneCurve, EndoArt Gold, and Typhoon groups (p>0.05). Conclusion: This is the first study in the literature for EndoArt NiTi files and the second study for VDW.ROTATE that evaluated cyclic fatigue resistance. Novel EndoArt Blue files exhibited significantly greater cyclic fatigue resistance than the other NiTi files.

🔖

Bu Makaleye Atıf Yap

KESKİN, N. B., HAZNEDAROĞLU, F., GÜNEÇ, H. G. (2022). Comparison of cyclic fatigue resistance of different and novel heat-treated nickel-titanium rotary file systems at the intracanal temperature. International Dental Research, Cilt: 11 - Sayı: 3, 158–164. https://ajindex.com/articles/539487

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (13)

  1. Yilmaz K, Uslu G, Ozyurek T. In vitro comparison of the cyclic fatigue resistance of HyFlex EDM, One G, and ProGlider nickel titanium glide path instruments in single and double curvature canals. Restor Dent Endod 2017;42:282-9. (Crossref
  2. Gavini G, Santos MD, Caldeira CL, Machado MEL, Freire LG, Iglecias EF, et al. Nickel-titanium instruments in endodontics: a concise review of the state of the art. Braz Oral Res 2018;32:e67. (Crossref
  3. Gundogar M, Ozyurek T. Cyclic fatigue resistance of OneShape, HyFlex EDM, WaveOne Gold, and Reciproc Blue nickel-titanium instruments. J Endod 2017;43:1192-96. (Crossref
  4. Pedulla E, Lo Savio F, Boninelli S, Plotino G, Grande NM, La Rosa G, et al. Torsional and cyclic fatigue resistance of a new Nickel-Titanium instrument manufactured by electrical discharge machining. J Endod 2016;42:156-9. (Crossref
  5. Gambarini G, Plotino G, Grande NM, AlSudani D, De Luca M, Testarelli L. Mechanical properties of nickel-titanium rotary instruments produced with a new manufacturing technique. Int Endod J 2011;44:337-41. (Crossref
  6. Modesto TC, Acosta ECP, Resende PD, Pereira ESJ, Peixoto I, Buono VTL, et al. Cyclic flexural fatigue resistance of NiTi controlled memory and blue technology instruments after torsional preloading. J App Oral Sci 2018;26:e20180144. (Crossref
  7. Shen Y, Qian W, Abtin H, Gao Y, Haapasalo M. Fatigue testing of controlled memory wire nickel-titanium rotary instruments. J Endod 2001;37:997-1001. (Crossref
  8. Otsuka K, Wayman CM. Shape memory materials. 1st ed. Cambridge (UK): Cambridge University Press; 1998
  9. Staffoli S, Grande NM, Plotino G, Ozyurek T, Gundogar M, Fortunato L, et al. Influence of environmental temperature, heat-treatment and design on the cyclic fatigue resistance of three generations of a single-file nickel-titanium rotary instrument. Odontology 2019;107:301-7. (Crossref
  10. Elnaghy AM, Elsaka SE. Cyclic fatigue resistance of One Curve, 2Shape, ProFile Vortex, Vortex Blue, and RaCe nickel-titanium rotary instruments in single and double curvature canals. J Endod 2018;44:1725-30. (Crossref
  11. Gambarini G, Piasecki L, Miccoli G, Gaimari G, Di Giorgio R, Di Nardo D, et al. Classification and cyclic fatigue evaluation of new kinematics for endodontic instruments. Aust Endod J 2019;45:154-62. (Crossref
  12. Larsen CM, Watanabe I, Glickman GN, He J. Cyclic fatigue analysis of a new generation of nickel titanium rotary instruments. J Endod 2009;35:401–3. (Crossref
  13. Pirani C, I