Dergiler / Hittite Journal of Science and Engineering / 2020 / Cilt: 7 - Sayı: 2

Gold Nanoparticles/Graphene Oxide Based Disposable Sensor System for Voltammetric Detection of Ceftizoxime

Sayfa
89–97
Yayın Tarihi
2022-10-18
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

This study results proved that developed disposable sensor is a good alternative for the practical application of CFX analysis.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:In this study, gold nanoparticles AuNPs were deposited onto graphene oxide GO modified pencil graphite electrode PGE in order to construct a disposable sensor platform for the electrochemical detection of ceftizoxime CFX .
📊
Temel Bulgular & Veri:Under optimum experimental conditions, good linearity was obtained for CFX response in the range between 0.02-2.0 μM of CFX concentrations with a low detection limit 0.442 nM by stripping voltammetry.
💡
Sonuç & Katkı:This study results proved that developed disposable sensor is a good alternative for the practical application of CFX analysis.
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Gold nanoparticles · Graphene oxide · Cephalosporins

Abstract

In this study, gold nanoparticles AuNPs were deposited onto graphene oxide GO modified pencil graphite electrode PGE in order to construct a disposable sensor platform for the electrochemical detection of ceftizoxime CFX . Initially, electrode surface was covered with GO by physical adsorption and then AuNPs were deposited on the surface by electro-deposition method. Morphological feature of the developed sensor was investigated by scanning electron microscope. The parameters effecting the experimental conditions such as adsorption time of graphene oxide, deposition time of gold nanoparticles, supporting electrolyte pH, pre-concentrating potential/time were optimized. Under optimum experimental conditions, good linearity was obtained for CFX response in the range between 0.02-2.0 μM of CFX concentrations with a low detection limit 0.442 nM by stripping voltammetry. The AuNPs/GO modified PGE was implemented to pharmaceutical samples with good recovery values. This study results proved that developed disposable sensor is a good alternative for the practical application of CFX analysis.

🔖

Bu Makaleye Atıf Yap

YARDIMCI, C., BOLAT, G., ABACI, S., YAMAN, Y. T. (2022). Gold Nanoparticles/Graphene Oxide Based Disposable Sensor System for Voltammetric Detection of Ceftizoxime. Hittite Journal of Science and Engineering, Cilt: 7 - Sayı: 2, 89–97. https://ajindex.com/articles/537936

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (11)

  1. Abo El-Maali N, Ghandour M a., Kauffmann JM. Cephalosporin antibiotics at carbon paste and modified carbon paste electrodes in both aqueous and biological media. Bioelectrochemistry and Bioenergetics 38 (1995) 91–97. DOI: 10.1016/0302-4598(95)01822-V
  2. Facca B, Frame B, Triesenberg S. Population pharmacokinetics of ceftizoxime administered by continuous infusion in clinically ill adult patients. Antimicrobial Agents and Chemotherapy 42 (1998) 1783–1787
  3. Wiseman LR, Benfield P. Cefprozil: A Review of its Antibacterial Activity, Pharmacokinetic Properties, and Therapeutic Potential. Drugs 45 (1993) 295–317. DOI: 10.2165/00003495-199345020- 00008
  4. Jain R, Gupta VK, Jadon N, Radhapyari K. Voltammetric determination of cefixime in pharmaceuticals and biological fluids. Analytical Biochemistry 407 (2010) 79–88. DOI: 10.1016/j. ab.2010.07.027
  5. Ferreira SMZMD, Domingos GP, Ferreira DDS, Rocha TGR, Serakides R, De Faria Rezende CM, et al. Technetium-99m-labeled ceftizoxime loaded long-circulating and pH-sensitive liposomes used to identify osteomyelitis. Bioorganic and Medicinal Chemistry Letters 22 (2012) 4605–4608. DOI: 10.1016/j.bmcl.2012.05.105
  6. Sanli S, Sanli N, Gumustas M, Ozkan S a., Karadas N, Aboul-Enein HY. Simultaneous estimation of ceftazidime and ceftizoxime in pharmaceutical formulations by HPLC method. Chromatographia 74 (2011) 549–558. DOI: 10.1007/s10337-011-2116-1
  7. Moore CM, Sato K, Hattori H, Katsumata Y. Improved HPLC method for the determination of cephalosporins in human plasma and a new solid-phase extraction procedure for cefazolin and ceftizoxime. Clinica Chimica Acta 190 (1990) 121–123. DOI: 10.1016/0009-8981(90)90290-9
  8. Péhourcq F, Jarry C. Determination of third-generation cephalosporins by high-performance liquid chromatography in connection with pharmacokinetic studies. Journal of Chromatography A 812 (1998) 159–178. DOI: 10.1016/S0021- 9673(98)00265-9
  9. Wang L, Zheng X, Zhong W, Chen J, Jiang J, Hu P. Validation and Application of an LC–MS-MS Method for the Determination of Ceftizoxime in Human Serum and Urine. Journal of Chromatographic Science 54 (2016) 713–719. DOI: 10.1093/ chromsci/bmv243
  10. Al-Momani I. Spectrophotometric determination of selected cephalosporins in drug formulations using flow injection analysis. Journal of Pharmaceutical and Biomedical Analysis 25 (2001) 751– 757. DOI: 10.1016/S0731-7085(01)00368-5
  11. Jain R, Rather JA, Dwivedi A, Vikas. Highly Sensitive and Selective Voltammetric S