Dergiler / Eklem Hastalıkları ve Cerrahisi Dergisi (Eski Adı: Artroplasti Artroskopik Cerrahi Derg.) / 2015 / Cilt: 26 - Sayı: 3

Biomechanical comparison of three-point bending resistance of titanium and stainless steel locking screws in intramedullary nails

İntramedüller çivilerde titanyum ve paslanmaz çelik kilit vidaların üç nokta eğilme dirençlerinin biyomekanik yönden karşılaştırılması

Sayfa
145–151
Yayın Tarihi
2022-10-18
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Bulgular: Kilit vidalarda en sık kullanılan yiv tipi olan alçak yivli vidalarda ortalama akma noktası titanyum vidalarda 1413 N idi ve bu değer 70 kg bir kişide düz zemin yürümede femura gelen 1922 N (2.8 BW) yüklenmenin altında idi.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Amaç: Bu çalışmada kilitli femur çivilerinde titanyum ve paslanmaz çelik kilit vidaların eğilme dirençleri arasında anlamlı fark olup olmadığı araştırıldı ve kilitli çivilemelerde sık görülen bir sorun olan kilit vida deformasyonu incelendi.
📊
Temel Bulgular & Veri:Gereç ve yöntemler: Çalışmada toplam 60 adet 5 mm majör çaplı titanyum ve paslanmaz çelik kilit vida üç farklı yiv yapısında (yüksek yivli, alçak yivli ve yivsiz) altı grup halinde kullanıldı.
💡
Sonuç & Katkı:Bulgular: Kilit vidalarda en sık kullanılan yiv tipi olan alçak yivli vidalarda ortalama akma noktası titanyum vidalarda 1413 N idi ve bu değer 70 kg bir kişide düz zemin yürümede femura gelen 1922 N (2.8 BW) yüklenmenin altında idi.
🔬Tür:Araştırma Makalesi
📍Kapsam:2071 · 60 adet
⚙️Değişkenler:Alan Parametreleri

Özet

Amaç: Bu çalışmada kilitli femur çivilerinde titanyum ve paslanmaz çelik kilit vidaların eğilme dirençleri arasında anlamlı fark olup olmadığı araştırıldı ve kilitli çivilemelerde sık görülen bir sorun olan kilit vida deformasyonu incelendi. Gereç ve yöntemler: Çalışmada toplam 60 adet 5 mm majör çaplı titanyum ve paslanmaz çelik kilit vida üç farklı yiv yapısında (yüksek yivli, alçak yivli ve yivsiz) altı grup halinde kullanıldı. Küçük trokanter seviyesini temsil eden 30 mm iç çapı olan paslanmaz çelik tüp içine yerleştirilen çelik vidalar üzerinde üç nokta eğilme testleri yapıldı. Kilit vidalarda kalıcı deformasyonun oluştuğu akma noktalarının tespit edilmesi için aksiyel kompresyon test makinası kullanıldı. Bulgular: Kilit vidalarda en sık kullanılan yiv tipi olan alçak yivli vidalarda ortalama akma noktası titanyum vidalarda 1413 N idi ve bu değer 70 kg bir kişide düz zemin yürümede femura gelen 1922 N (2.8 BW) yüklenmenin altında idi. Eğilme direnci alçak yivli paslanmaz vidalarda 2071 N olarak 1922 N değerinin üzerinde idi. Yüksek yivli vidalarda ortalama eğilme akma noktası titanyum vidalarda 874 N ve paslanmaz çelik vidalarda 556 N idi. Sonuç: Alçak yivli paslanmaz çelik vidalar titanyumlara göre eğilme deformasyonuna %46.5 daha dirençli olduğundan, parçalı femur cisim kırıklarında (tüm vücut ağırlığının yüklenme durumunda) kilit vida deformasyonu olmaması için titanyum vida yerine paslanmaz çelik kilit vida kullanılması daha uygundur. Paslanmaz çelik veya titanyum yüksek yivli kilit vidalar sadece parçalı olmayan kırıklarda çok dikkatli bir şekilde kullanılabilir

Abstract

Objectives: This study aims to investigate whether there is any significant difference in bending resistance between titanium and stainless steel locking screws of femur nails and to review deformation of locking screws which is a common problem in interlocking nailing. Materials and methods: In this study, a total of 60 pieces of 5 mm major diameter titanium and stainless steel locking screws were used as six groups in three different thread depth structures (high threaded, low threaded, and unthreaded). Three-point bending tests were conducted on steel screws placed inside stainless steel tube with 30 mm inner diameter, which imitated the level of lesser trochanter. We used an axial compression testing machine in order to determine the yield points that permanent deformation occurred in the locking screws. Results: For low threaded locking screws, which are the most frequently used thread type for locking screws, the mean bending yield points were 1413 N on the titanium screws and this level was below 1922 N (2.8 BW) of level walking loading on femur for 70 kg person. On low threaded stainless screws, bending resistance was 2071 N, which was above the value of 1922 N. For high threaded locking screws, the mean bending yield points were 874 N on the titanium screws and 556 N on stainless screws. Conclusion: In comminuted femur shaft fractures (in full load bearing conditions), using stainless steel locking screws is better instead of titanium screws to avoid locking screw deformation since low threaded stainless steel screws were 46.5% more resistant to bending deformation than titanium ones. Stainless steel or titanium high threaded locking screws may only be carefully used in non-comminuted fractures.

🔖

Bu Makaleye Atıf Yap

KARAKAŞLI, A., AYCAN, H., KARAARSLAN, A. A., SESLİ, E., ERTEM, F. (2022). Biomechanical comparison of three-point bending resistance of titanium and stainless steel locking screws in intramedullary nails. Eklem Hastalıkları ve Cerrahisi Dergisi (Eski Adı: Artroplasti Artroskopik Cerrahi Derg.), Cilt: 26 - Sayı: 3, 145–151. https://ajindex.com/articles/498539

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (14)

  1. Colchero F, Orst G, Reboul C, Villalobos F, Vidal J. Intramedullary locking nailing. Experimental study. Surgical technic. Results. Rev Chir Orthop Reparatrice Appar Mot 1983
  2. [Abstract] 2. Kempf I, Grosse A, Taglang G, Favreul E. Gamma nail in the treatment of closed trochanteric fractures. Results and indications of 121 cases. Orthop Traumatol Surg Res 2014
  3. Whittle AP, Wester W, Russell TA. Fatigue failure in small diameter tibial nails. Clin Orthop Relat Res 1995
  4. Boenisch UW, de Boer PG, Journeaux SF. Unreamed intramedullary tibial nailing--fatigue of locking bolts. Injury 1996
  5. Whittle AP, Russell TA, Taylor JC, Lavelle DG. Treatment of open fractures of the tibial shaft with the use of interlocking nailing without reaming. J Bone Joint Surg [Am] 1992
  6. Court-Brown CM, Will E, Christie J, McQueen MM. Reamed or unreamed nailing for closed tibial fractures. A prospective study in Tscherne C1 fractures. J Bone Joint Surg [Br] 1996
  7. Aldemir C, Doğan A, İnci F, Sertkaya Ö, Duygun F. Distal locking techniques without fluoroscopy in intramedullar nailing. [Article in Turkish] Eklem Hastalik Cerrahisi 2014
  8. Yilmaz E, Karakurt L, Bulut M, Belhan O, Serin E. Treatment of femoral shaft fractures and pseudoarthrosis with compressive and interlocking intramedullary nailing. [Article in Turkish] Acta Orthop Traumatol Turc 2005
  9. Başcı O, Karakaşlı A, Kumtepe E, Güran O, Havıtçıoğlu H. Combination of anatomical locking plate and retrograde intramedullary nail in distal femoral fractures: comparison of mechanical stability. Eklem Hastalik Cerrahisi 2015
  10. Karakaşli A, Satoğlu İS, Havitçioğlu H. A new intramedullary sustained dynamic compression nail for the treatment of long bone fractures: a biomechanical study. Eklem Hastalik Cerrahisi 2015
  11. Chao CK, Hsu CC, Wang JL, Lin J. Increasing bending strength of tibial locking screws: mechanical tests and finite element analyses. Clin Biomech (Bristol, Avon) 2007
  12. Hou SM, Wang JL, Lin J. Mechanical strength, fatigue life, and failure analysis of two prototypes and five conventional tibial locking screws. J Orthop Trauma 2002
  13. Taylor SJ, Walker PS, Perry JS, Cannon SR, Woledge R. The forces in the distal femur and the knee during walking and other activities measured by telemetry. J Arthroplasty 1998
  14. Taylor SJ, Walker PS. Forces and moments telemetered from two