Impact of Magnetic Field on Pressures of Programmable Cerebrospinal Fluid Shunts: An Experimental Study

Sayfa
935–941
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

However, further controlled, clinical trials are warranted to elucidate the in-vivo effects of magnetic field exposure

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:AIm: To investigate whether programmable cerebrospinal fluid (CSF) shunts are influenced by exposure to the magnetic field and to compare the effects of magnetic field in 4 different brands of programmable CSF shunts. mATERIAl and mEThODS: This experimental study was performed in the laboratory using a novel design of magnetic field.
📊
Temel Bulgular & Veri:Four types of programmable CSF shunts (Miethke®, Medtronic®, Sophysa® and Codman®Hakim®) were exposed to the magnetic field generated by an apparatus consisting of Helmholtz coil for 5 minutes.
💡
Sonuç & Katkı:However, further controlled, clinical trials are warranted to elucidate the in-vivo effects of magnetic field exposure
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Alan Parametreleri

Özet

AIm: To investigate whether programmable cerebrospinal fluid (CSF) shunts are influenced by exposure to the magnetic field and to compare the effects of magnetic field in 4 different brands of programmable CSF shunts. mATERIAl and mEThODS: This experimental study was performed in the laboratory using a novel design of magnetic field. Four types of programmable CSF shunts (Miethke®, Medtronic®, Sophysa® and Codman®Hakim®) were exposed to the magnetic field generated by an apparatus consisting of Helmholtz coil for 5 minutes. In every CSF shunt, initial pressures were adjusted to 110 mm H2O and pressures after exposure to magnetic field were noted. These measurements were implemented at frequencies of 5 Hz, 20 Hz, 30 Hz, 40 Hz, 60 Hz and 80 Hz. In each type, three shunts were utilized and evaluations were made twice for every shunt. RESUlTS: At 5, 30, 40 and 60 Hz, Groups 1, 2 and 3 had significantly higher average pressures than Group 4. At 20 and 80 Hz, Groups 1 and 2 had notably different pressure values than Groups 3 and 4. Group 3 displayed the highest pressure, while Group 4 demonstrated the lowest pressure. CONClUSION: Exposure to magnetic fields may affect the pressures of programmable CSF shunts. However, further controlled, clinical trials are warranted to elucidate the in-vivo effects of magnetic field exposure

🔖

Bu Makaleye Atıf Yap

ALTUN, İ., YÜKSEL, K. Z., MERT, T. (2017). Impact of Magnetic Field on Pressures of Programmable Cerebrospinal Fluid Shunts: An Experimental Study. Turkish Neurosurgery, Cilt: 27 - Sayı: 6, 935–941. https://ajindex.com/articles/404539

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (1)

  1. Anderson RC, Walker ML, Viner JM, Kestle JR: Adjustment and malfunction of a programmable valve after exposure to toy magnets. Case report. J Neurosurg 101: 222-225, 2004 2. Jandial R, Aryan HE, Hughes SA, Levy ML: Effect of vagus nerve stimulator magnet on programmable shunt settings. Neurosurgery 55: 627-629, 2004 3. Lavinio A, Harding S, Van Der Boogaard F, Czosnyka M, Smielewski P, Richards HK, Pickard JD, Czosnyka ZH: Magnetic field interactions in adjustable hydrocephalus shunts. J Neurosurg Pediatrics 2:222-228, 2008 4. Lindner D, Preul C, Trantakis C, Moeller H, Meixensberger J: Effect of 3T MRI on the function of shunt valves-evaluation of Paedi GAV, Dual Switch and proGAV. Eur J Radiol 56: 56–59, 2005 5. Mert T, Altun I, Celik A, Surer T, Gunay I: Modulation of cytokine levels in ameliorative effects of pulsed magnetic field on an experimental model of Chronic Constriction Injury. Int J Radiat Biol 91: 596-602, 2015 6. Mert T, Ocal I, Gunay I: Pain relieving effects of pulsed magnetic fields in a rat model of carrageenan-induced hindpaw inflammation. Int J Radiat Biol 90:95–103, 2014 7. Miwa K, Kondo H, Sakai N: Pressure changes observed in Codman-Medos programmable valves following magnetic exposure and filliping. Childs Nerv Syst 17:150-153, 2001 8. Nomura S, Fujisawa H, Suzuki M: Effect of cell phone magnetic fields on adjustable cerebrospinal fluid shunt valves. Surg Neurol 63:467-468, 2005 9. Schneider T, Knauff U, Nitsch J, Firsching R: Electromagnetic field hazards involving adjustable shunt valves in hydrocephalus. J Neurosurg 96: 331–334, 2002 10. Sprung C, Schlosser HG, Lemcke J, Meier U, Messing-Jünger M, Trost HA, Weber F, Schul C, Rohde V, Ludwig HC, Höpfner J, Sepehrnia A, Mirzayan MJ, Krauss JK: The adjustable proGAV shunt: A prospective safety and reliability multicenter study. Neurosurgery 66:465-474, 2010 11. Strahle J, Selzer BJ, Muraszko KM, Garton HJL, Maher CO: Programmable shunt valve affected by exposure to a tablet computer. J Neurosurg Pediatrics 10:118-120, 2012