Dergiler / Turkish Journal of Orthodontics / 2019 / Cilt: 32 - Sayı: 1

Effects of High-Energy Curing Lights on Time-Dependent Temperature Changes of Pulp Space During Orthodontic Bonding

Sayfa
22–27
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

The exposure time is more important than the energy output level of the light-curing system on tempera-ture increments of the pulp chamber.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Objective: The purpose of this study was to assess the temperature changes and cooling times during orthodontic bonding by alight-emitting diode (LED) and plasma arc lights (PAC) in different time and power modes with thermal imaging.Methods: A total of 100 human permanent upper first premolar teeth were included in the study.
📊
Temel Bulgular & Veri:Five groups were conducted, 20teeth each, and different energy outputs of curing lights were used for adhesive polymerization with different exposure times.
💡
Sonuç & Katkı:The exposure time is more important than the energy output level of the light-curing system on tempera-ture increments of the pulp chamber.
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Alan Parametreleri

Özet

Objective: The purpose of this study was to assess the temperature changes and cooling times during orthodontic bonding by alight-emitting diode (LED) and plasma arc lights (PAC) in different time and power modes with thermal imaging.Methods: A total of 100 human permanent upper first premolar teeth were included in the study. Five groups were conducted, 20teeth each, and different energy outputs of curing lights were used for adhesive polymerization with different exposure times. Thetemperature changes in the pulp space and cooldown times were measured by a thermal imaging system. A paired t-test, analysis ofvariance (ANOVA), and Student-Newman-Keuls multiple comparison tests were used for data analysis.Results: A statistically significant temperature rise was detected with all curing lights (p<0.05). The greatest temperature changeswere observed in the LED standard mode with 10 seconds of exposure time (6.66±1.98°C) and LED extra power mode with 6 secondsof exposure time (6.50±1.64°C) among groups, while using PAC for 3 seconds created the smallest temperature increase (1.81±0.99°C).An application of the LED extra power mode for 6 seconds exhibited the longest cooldown time (205.91±47.48 seconds), and theshortest cooldown time was detected as 71.30±43.15 seconds with the PAC 3-second application.Conclusions: LED lights with an increased exposure time induced significant temperature rises, while no PAC light group exceededthe critical threshold value. The exposure time is more important than the energy output level of the light-curing system on tempera-ture increments of the pulp chamber.

🔖

Bu Makaleye Atıf Yap

AMUK, N. G., KURT, G., ER, Ö., ÇAKMAK, G., ASLANTAŞ, V. (2019). Effects of High-Energy Curing Lights on Time-Dependent Temperature Changes of Pulp Space During Orthodontic Bonding. Turkish Journal of Orthodontics, Cilt: 32 - Sayı: 1, 22–27. https://ajindex.com/articles/374999

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (4)

  1. Malkoc S, Uysal T, Usumez S, Isman E, Baysal A. In-vitro assessment of temperature rise in the pulp during orthodontic bonding. Am J Orthod Dentofacial Orthop 2010; 137: 379-83. [CrossRef ] Uzel A, Buyukyilmaz T, Kayalioglu M, Uzel I. Temperature rise during orthodontic bonding with various light-curing units - an in vitro study. Angle Orthod 2006; 76: 330-4.
  2. Lindberg A, Peutzfeldt A, Van Dijken JW. Effect of power density of curing unit, exposure duration, and light guide distance on com- posite depth of cure. Clin Oral Investig 2005; 9: 71-6. [CrossRef ] Goodis HE, White JM, Andrews J, Watanabe LG. Measurement of temperature generated by visible-light cure lamps in an in vitro model. Dent Mater 1989; 5: 230-4. [CrossRef ] Hannig M, Bott B. In-vitro pulp chamber temperature rise during composite resin polymerization with various light-curing sources. Dent Mater 1999; 15: 275-81. [CrossRef ] Tarle Z, Meniga A, Knezevic A, Sutalo J, Ristic M, Pichler G. Compos- ite conversion and temperature rise using a conventional, plasma arc and experimental blue LED curing unit. J Oral Rehabil 2002; 29: 662-7. [CrossRef ] Zach L, Cohen G. Pulp response to externally applied heat. Oral Surg Oral Med Oral Pathol 1965; 19: 515-30. [CrossRef ] Buchalla W, Attin T. External bleaching therapy with activation by heat, light or laser-a systematic review. Dent Mater 2007; 23: 586- 96. [CrossRef ] Ulusoy C, Irmak O, Bagis YH, Ulusoy OI. Temperature rise and shear bond strength of bondable buccal tubes bonded by various light sources. Eur J Orthod 2008; 30: 413-7. [CrossRef ] Cacciafesta V, Sfondrini MF, Barina E, Scribante A, Garino F, Klersy C. Effect of different light sources and guides on shear bond strength of brackets bonded with 2 adhesive systems. Am J Orthod Dentofa- cial Orthop 2005; 128: 99-102. [CrossRef ] Price RB, Ehrnford L, Andreou P, Felix CA. Comparison of quartz-tung- sten-halogen, light-emitting diode, and plasma arc curing lights. J Adhes Dent 2003; 5: 193-207.
  3. Cummings M, Biagioni P, Lamey PJ, Burden DJ. Thermal image anal- ysis of electrothermal debonding of ceramic brackets: an in vitro study. Eur J Orthod 1999; 21: 111-8. [CrossRef ] Kraut J, Radin S, Trowbridge HI, Emling RC, Yankell SL. Clinical eval- uations on thermal versus mechanical debonding of ceramic brack- ets. J Clin Dent 1991; 2: 92-6.
  4. Jost-Brinkmann PG, Stein H, Miethke RR, Nakata M. Histological in- vestigations of human pulp after thermo-debonding of metal and ceramic brackets. Am J Ortho