Verification studies in glucometers: Should we use capillary blood or venous blood for comparison?

Sayfa
218–221
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

In glucometer verification studies, capillary blood obtained with capillary blood sampling systems and used instead of venous blood should be the preferred sample.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Glucometers are widely used in the diagnosis of blood glucose levels in patients with diabetes mellitus.
📊
Temel Bulgular & Veri:EN ISO 15197 suggests that glucometer comparison studies should have 100 capillary blood samples be worked on at least twice.
💡
Sonuç & Katkı:In glucometer verification studies, capillary blood obtained with capillary blood sampling systems and used instead of venous blood should be the preferred sample.
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Alan Parametreleri

Özet

Glucometers are widely used in the diagnosis of blood glucose levels in patients with diabetes mellitus. EN ISO 15197 suggests that glucometer comparison studies should have 100 capillary blood samples be worked on at least twice. In this study, we planned on comparing the glucose results measured in a routine biochemistry analyzer from two different glucometers, capillary and venous blood samples, and aimed to discuss the effects of blood taking systems on the glucometer validation studies. Capillary and venous blood samples were taken from 101 individuals and their glucose concentrations measured simultaneously using two different glucometers (Accu-chek and GlucoMax). Capillary and venous blood samples were centrifuged after clotting and analyzed in the Roche P modular system. In the fasting condition, the equations for regression analysis that were found y=0,873x+24,32 (r=0,857) in between Accu-chek and venous blood glucose, y=0,9x+16,15 (r=0,920) in between Accu-chek and capillary blood glucose, y=0,811x+20,94 (r=0,776) in between GlucoMax and venous blood glucose, and y=0,851x+12,28 (r=0,863) in between GlucoMax and capillary blood glucose.In the postprandial state, the equations were y=0,713x+48,46 (r=0,258) in between Accu-chek and venous blood glucose, y=0,981x+11,77 (r=0,718) in between Accu-chek and capillary blood glucose, y=0,706x+39,12 (r=0,453) in between GlucoMax and venous blood glucose, and y=0,790+22,35 (r=0,787) in between GlucoMax and capillary blood glucose. In the fasting and postprandial state, the capillary glucose levels showed better correlation with glucometer measurements than venous blood glucose levels. In glucometer verification studies, capillary blood obtained with capillary blood sampling systems and used instead of venous blood should be the preferred sample.

🔖

Bu Makaleye Atıf Yap

TOPÇUOĞLU, C., YILMAZ, F. M. (2018). Verification studies in glucometers: Should we use capillary blood or venous blood for comparison?. Medicine Science, Cilt: 7 - Sayı: 1, 218–221. https://ajindex.com/articles/277293

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (11)

  1. Looie RF, Tang Z, Sutton DV, Lee JH, Kost GJ. Point-of-care testing effects of critical variables, influence of reference instruments, modular glucose meter design. Arch Pathol Lab Med. 2000;124(2):257-66
  2. Khan AI, Vasquez Y, Gray J, Wians FH Jr, Kroll MH. The variability of results between point-of-care testing glucose meters and the central laboratory analyzer. Arch Pathol Lab Med. 2006;130(10):1527-32.
  3. Marley JV, Davis S, Coleman, Hayhow BD, Brennan G, Mein JK, Nelson C, Atkinson D, Maguire GP. Point-of-care testing of capillary glucose in the exclusion and diagnosis of diabetes in remote Australia. Med J Aust. 2007;186(10):500-3.
  4. Püntmann I, Wosniok W, Haeckel R. Comparison of several point-of-care testing (POCT) glucometers with an established laboratory procedure fort the diagnosis of type 2 diabetes using the discordance rate. A new statistical approach. Clin Chem Lab Med. 2003;41(6):809-20.
  5. Robinson CS, Sharp P. Tighter accuracy standards within point-of-care blood glucose monitoring: how six commonly used systems compare. Journal of Diabetes Science and Technology. 2012;6(3):547-54. Hawkins RC. Evaluation of Roche Accu-Chek go and Medisense Optium blood glucose meters. Clin Chim Acta. 2005;353(1-2):127-31.
  6. Mraovic B, Schwenk ES, Ebstein RH. Intraoperative accuracy of a pointof-care glucose meter compared with simultaneous central laboratory measurements. J Diabetes Sci Technol. 2012;6(3):541-6. International Organization for Standardization. In vitro diagnostic test systems-requirement for blood glucose monitoring systems for self-testing in managing diabetes mellitus. ISO 15197:2003.
  7. Funk DI, Chan L, Lutz N, Verdile VP. Comparison of capillary and venous glucose measurements in healthy volunteer. Prehop Emerg Care. 2001;5(3):275-7
  8. Boyd R, Leigh B, Stuart P. Capillary versus venous bedside blood glucose estimations. Emerg Me J. 2005;22(3):177-9.
  9. Critchell CD, Savarese V, Callahan A, Aboud C, Jabbour S, Marik P. Accuracy of bedside capillary blood glucose measurements in critically ill patients. Intensive Care Med. 2007;33(12):2079-84.
  10. Petersen JR, Graves DF, Tacker DH, Okorodudu AO, Mohammad AA, Cardenas VJ Jr. Comparison of POCT and central laboratory blood glucose results using arterial, capillary, and venous samples from MICU patients on a tigt glycemic protocol. Clin Chim Acta. 2008;396(1-2):10-13.
  11. Kuwa K, Nakayama T, Hoshino T, Tominaga M. Relationships of glucose concentrations in capillary whole blood, venous whole blood and venous plasma. C