Construction and Characterization of Indium Liquid Ion Selective Electrodes Based on Crown Ethers in a PVC Matrix Membrane

Sayfa
687–696
Yayın Tarihi
2005-06-01
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

Each membrane showed a rapid decrease in absorbance when it reached the end of its life due to leaching of the membrane contents to the external solution.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:Several indium ion selective electrodes were constructed based on 15-crown-5 (15C5), dicyclohexano-18-crown-6 (DCH18C6) and dibenzo-24-crown-8 (DB24C8) as sensors and di-octyl phenyl phosphonate (DOPP), di-butyl phthalate (DBPH), tri-butyl phosphate (DBP), tri-n-amyl phosphate (TAP) and bis (2-ethyl hexyl) phosphate (BEHP) as plasticizers in a PVC matrix membrane.
📊
Temel Bulgular & Veri:The indium electrode based on 15C5 using DOPP as a plasticizer gave the best response for low levels of In3+ ions.
💡
Sonuç & Katkı:Each membrane showed a rapid decrease in absorbance when it reached the end of its life due to leaching of the membrane contents to the external solution.
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Indium electrodes · crown compounds · selectivity measurements

Özet

Several indium ion selective electrodes were constructed based on 15-crown-5 (15C5), dicyclohexano-18-crown-6 (DCH18C6) and dibenzo-24-crown-8 (DB24C8) as sensors and di-octyl phenyl phosphonate (DOPP), di-butyl phthalate (DBPH), tri-butyl phosphate (DBP), tri-n-amyl phosphate (TAP) and bis (2-ethyl hexyl) phosphate (BEHP) as plasticizers in a PVC matrix membrane. Electrode parameters, pH effect and interferences were also studied. The indium electrode based on 15C5 using DOPP as a plasticizer gave the best response for low levels of In3+ ions. The linear range for this electrode ranged from 3.8 x 10-5 to 5.0 x 10-2 M (with correlation coefficient r = 0.9997), the slope was 20.1 mV/decade with a detection limit of 1.2 x 10-5 M, and the working pH range was 3.8--5.9. Interferences of Na+ , K+ , NH4+ , Ca2+ , Ba2+ , Cd2+ , La3+ , Nd3+ and Eu3+ on the electrode response were also investigated using a mixed solution method. Mono- and divalent cations were found to interfere less than trivalent cations. The concentrations of indium ions in the prepared standard solutions were determined by direct, standard addition (SA), multiple standard addition (MSA) and titration methods. The results were compared with those obtained by atomic absorption spectrophotometry. The precision and accuracy of each method were calculated and expressed in terms of the relative standard deviation and the relative error. Optical measurements were also studied by measuring the absorbance for each membrane after steady intervals of immersion in a standard solution of its ions using UV/Visible spectrophotometry. Each membrane showed a rapid decrease in absorbance when it reached the end of its life due to leaching of the membrane contents to the external solution.

Abstract

Several indium ion selective electrodes were constructed based on 15-crown-5 (15C5), dicyclohexano-18-crown-6 (DCH18C6) and dibenzo-24-crown-8 (DB24C8) as sensors and di-octyl phenyl phosphonate (DOPP), di-butyl phthalate (DBPH), tri-butyl phosphate (DBP), tri-n-amyl phosphate (TAP) and bis (2-ethyl hexyl) phosphate (BEHP) as plasticizers in a PVC matrix membrane. Electrode parameters, pH effect and interferences were also studied. The indium electrode based on 15C5 using DOPP as a plasticizer gave the best response for low levels of In3+ ions. The linear range for this electrode ranged from 3.8 x 10-5 to 5.0 x 10-2 M (with correlation coefficient r = 0.9997), the slope was 20.1 mV/decade with a detection limit of 1.2 x 10-5 M, and the working pH range was 3.8--5.9. Interferences of Na+ , K+ , NH4+ , Ca2+ , Ba2+ , Cd2+ , La3+ , Nd3+ and Eu3+ on the electrode response were also investigated using a mixed solution method. Mono- and divalent cations were found to interfere less than trivalent cations. The concentrations of indium ions in the prepared standard solutions were determined by direct, standard addition (SA), multiple standard addition (MSA) and titration methods. The results were compared with those obtained by atomic absorption spectrophotometry. The precision and accuracy of each method were calculated and expressed in terms of the relative standard deviation and the relative error. Optical measurements were also studied by measuring the absorbance for each membrane after steady intervals of immersion in a standard solution of its ions using UV/Visible spectrophotometry. Each membrane showed a rapid decrease in absorbance when it reached the end of its life due to leaching of the membrane contents to the external solution.

🔖

Bu Makaleye Atıf Yap

ABDULLA, N. I., Al--HAIDER, A. (2005). Construction and Characterization of Indium Liquid Ion Selective Electrodes Based on Crown Ethers in a PVC Matrix Membrane. Turkish Journal of Chemistry, Cilt: 29 - Sayı: 6, 687–696. https://ajindex.com/articles/227807

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (13)

  1. A.R. Fakhari, M.R. Ganjali and M. Shamsipur, Anal. Chem. 69, 3693 (1997).
  2. R. Garcia-Valls, M. Munoz and M. Valiente, Anal. Chim. Acta 387, 77 (1999).
  3. A.C. Didarul, O. Takashi, K. Satsuo and O. Kousaburo, Anal. Chem. 68, 366 (1996).
  4. N.S. Nassory, Talanta 36, 672 (1989).
  5. N.S. Nassory and N.H. Karim, Turk. J. Chem. 18, 126 (1994).
  6. J.C. Bailar, J.R. Urbana and H.J. Emelius, “Comprehensive Inorganic Chemistry” Vol. 1, (1973).
  7. P.C. Stark, M. Huff, E.A. Babiaan, L.M. Barden, D.C. Hrncir, S.G. Bott and J.L. Atwood, J. Inclusion Phenom. 5, 683 (1987).
  8. K. Fujihawa, M. Akiyamu and A.K. Shono, Kagaku Robunshu Jpn. 15, 381 (1989).
  9. K. Kondo, Y. Yamamoto, M. Matsumoto, J. Membr. Sci. 137, 9 (1997).
  10. L. Cunxiong and H. Jing-Guo, Talanta 44, 729 (1997). R.W. Cattrall, “Chemical Sensors” University Press Oxford, 1997.
  11. E. Bakker, P. Buhlmann and E. Pretsch, Chem. Rev. 97, 3083 (1997).
  12. A. Siminara and A. Musumeci, Inorganica Chim. Acta 39, 9 (1980).
  13. A. Craggs, G.J. Moody and J.D.R. Thomas, Chem. Educ. 51, 541 (1974). P.L. Bailey,.“ Analysis with Ion Selective Electrodes” Heyden and Son Ltd., London, 1976. 1970. 616. Orion Research Incorporated, “Gran’ s plots and other schemes”, Nov./Dec. K.S. Suresh, K.G. Vinod and J. Suresh, Anal. Chem. 67, 4234 (1995).