Dergiler / Atmospheric Pollution Research / 2020 / Cilt: 11 - Sayı: 4

Traceability of the output concentration of mercury vapor generators

Sayfa
639–645
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

The NIST Prime generator is certified using a primary analytical method for the determination of mercury using sorbent trap analysis with isotope dilution cold-vapor inductively coupled plasma mass spectrometry (ID-CV-ICP-MS).

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:In response to the environmental impact of mercury and the neuro-toxic risk associated with trophic transfer of methyl mercury to humans, the reduction of mercury emissions from coal-fired electric utilities, a significant anthropogenic source, is now receiving global attention.
📊
Temel Bulgular & Veri:To address impending regulations for mercury reduction, there is a need to establish a mercury emissions measurement framework, which has demonstrated accuracy and traceability to national and international standards.
💡
Sonuç & Katkı:The NIST Prime generator is certified using a primary analytical method for the determination of mercury using sorbent trap analysis with isotope dilution cold-vapor inductively coupled plasma mass spectrometry (ID-CV-ICP-MS).
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Alan Parametreleri

Özet

In response to the environmental impact of mercury and the neuro-toxic risk associated with trophic transfer of methyl mercury to humans, the reduction of mercury emissions from coal-fired electric utilities, a significant anthropogenic source, is now receiving global attention. To address impending regulations for mercury reduction, there is a need to establish a mercury emissions measurement framework, which has demonstrated accuracy and traceability to national and international standards. To support this effort, the National Institute of Standards and Technology (NIST) has been working jointly with the United States Environmental Protection Agency (EPA) to develop a traceability system which links the output of elemental mercury calibration instrumentation sited at the power utilities, to the SI (International System of Units). This system is based on a reference (NIST Prime) elemental mercury generator standard, which is used by NIST to certify elemental mercury generators from the instrument vendors. These in turn are used by the instrument vendors to provide NIST traceable calibration for the instrumentation supplied to each electric utility. The NIST Prime generator is certified using a primary analytical method for the determination of mercury using sorbent trap analysis with isotope dilution cold-vapor inductively coupled plasma mass spectrometry (ID-CV-ICP-MS).

🔖

Bu Makaleye Atıf Yap

LONG, S. E., NORRIS, J. E., CARNEY, J., RYAN, J. V., MITCHELL, G. D., DORKO, W. D. (2020). Traceability of the output concentration of mercury vapor generators. Atmospheric Pollution Research, Cilt: 11 - Sayı: 4, 639–645. https://ajindex.com/articles/158961

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (7)

  1. Christopher, S., Long, S., Rearick, M., Fassett, J., 2001. Development of isotope dilution cold vapor inductively coupled plasma mass spectrometry and its application to the certification of mercury in NIST Standard Reference Materials. Anal. Chem. 73, 2190–2199. FEDREG, 2005.
  2. Federal Register, Part II, Environmental Protection Agency, 40 CFR, Parts 60, 72 and 75, Standards for New and Existing Stationary Sources: Electric Utility Steam Generating Units. Final Rule May 18, 2005. FEDREG, 2011.
  3. Federal register, Part II, environmental protection agency, 40 CFR, parts 60 and 63, national emission standards for hazardous air pollutants from coal- and oil-fired electric utility steam generating units and standards of performance for fossil-fuel-fired electric utility, industrial-commercial-institutional, and small industrial-commercial-institutional steam generating units. Proposed Rule, May 3, 2011. JCGM, 2008. 100:2008; Guide to the Expression of Uncertainty in Measurement; (ISO GUM 1995 with Minor Corrections), Joint Committee for Guides in Metrology; BIPM. Sevres Cedex, France. JCGM, 2012. 200:2012; International Vocabulary of Metrology – Basic and General Concepts and Associated Terms (VIM), third ed. BIPM, Sevres, Cedex, France 2012. http://www.bipm.org/en/publications/guides/vim.html.
  4. Levenson, M., Banks, D., Eberhardt, K., Gill, L., Guthrie, W., Liu, H., Vangel, M., Yen, J., Zhang, N., 2000. An approach to combining results from multiple methods motivated by the ISO GUM. J. Res. Natl. Inst. Stand. Technol. 105, 571–579.
  5. Long, S., Kelly, W., 2002. Determination of mercury in coal by isotope dilution cold-vapor generation inductively coupled plasma mass spectrometry. Anal. Chem. 276, 49371–49377.
  6. Long, S., Norris, J., Carney, J., Ryan, J., 2019. Provision of primary NIST traceability to support vapor phase mercury emissions monitoring of combustion sources using isotope dilution inductively coupled plasma mass spectrometry. Atmos. Pollut. Res (in press). Srivastava, A., Hodges, J., 2018. Development of a high-resolution laser absorption spectroscopy method with application to the determination of absolute concentration of gaseous elemental mercury in air. Anal. Chem. 90, 6781–6788. UNEP, 2013. Global Mercury Assessment 2013:
  7. Sources, Emissions, Releases and Environmental Transport. UNEP Chemicals Branch, Geneva, Switzerland. UNEP, 2019. Global Mercury Assessment 2018. UN Environment Programme. Chemicals and Health Branch, Geneva, Switzerland. US EPA, 2009