Central Corneal Thickness Measurements with Different Imaging Devices: Ultrasound Pachymetry, Noncontact Specular Microscopy, and Tono-Pachymetry

Pages
54–58
Publication Date
2022-11-06
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

The mean CCT measured using NCT (559.3±39 μm) was significantly higher than those measured using SM (534.8±41 μm) and USP (542.6±43 μm, p

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objectives: Investigation of the compatibility between central corneal thickness (CCT) measurements in healthy eyes by comparing standard ultrasound pachymetry (USP) with noncontact tono-pachymetry (NCT) and specular microscopy (SM) devices was aimed.
📊
Key Findings & Data:Methods: Forty-five eyes of 45 healthy volunteers aged between 18 and 60 years were included in this study.
💡
Conclusion & Impact:The mean CCT measured using NCT (559.3±39 μm) was significantly higher than those measured using SM (534.8±41 μm) and USP (542.6±43 μm, p
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Objectives: Investigation of the compatibility between central corneal thickness (CCT) measurements in healthy eyes by comparing standard ultrasound pachymetry (USP) with noncontact tono-pachymetry (NCT) and specular microscopy (SM) devices was aimed. Methods: Forty-five eyes of 45 healthy volunteers aged between 18 and 60 years were included in this study. CCT of all cases was evaluated with USP, NCT, and SM devices. The same examiner performed all examinations. Bland–Altman plots and intraclass correlation coefficients were used to evaluate the agreement between instruments. Results: The mean age of the patients was 31±10.2 years. Fifteen (33.3%) cases were male and 30 (66.7%) were female. The mean CCT measured using NCT (559.3±39 μm) was significantly higher than those measured using SM (534.8±41 μm) and USP (542.6±43 μm, p

🔖

Cite This Article

CEYLAN, A., ÖNAL, I., MERGEN, B., YILDIRIM, Y. (2022). Central Corneal Thickness Measurements with Different Imaging Devices: Ultrasound Pachymetry, Noncontact Specular Microscopy, and Tono-Pachymetry. BEYOGLU EYE JOURNAL, Cilt: 7 - Sayı: 1, 54–58. https://ajindex.com/articles/649044

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (12)

  1. Brandt JD. Corneal thickness in glaucoma screening, diagnosis, and management. Curr Opin Ophthalmol 2004;15:85–9
  2. Chihara E. Assessment of true intraocular pressure: the gap between theory and practical data. Surv Ophthalmol 2008;53:203–18
  3. Grewal DS, Brar GS, Grewal SPS. Assessment of central corneal thickness in normal, keratoconus, and post-laser in situ keratomileusis eyes using Scheimpflug imaging, spectral domain optical coherence tomography, and ultrasound pachymetry. J Cataract Refract Surg 2010;36:954–64
  4. González-Pérez J, Queiruga Piñeiro J, Sánchez García Á, González Méijome JM. Comparison of central corneal thickness measured by standard ultrasound pachymetry, corneal topography, tono-pachymetry and anterior segment optical coherence tomography. Curr Eye Res 2018;43:866–72
  5. Fujioka M, Nakamura M, Tatsumi Y, Kusuhara A, Maeda H, Negi A. Comparison of Pentacam Scheimpflug camera with ultrasound pachymetry and noncontact specular microscopy in measuring central corneal thickness. Curr Eye Res 2007;32:89– 94
  6. Williams R, Fink BA, King-Smith PE, Mitchell GL. Central corneal thickness measurements: Using an ultrasonic instrument and 4 optical instruments. Cornea 2011;30:1238–43
  7. Demir ST, Odabaşi M, Oba ME, Dirim AB, Can E, Kara O. Ultrasonik pakimetre ve orbscan korneal topografi santral kornea kalinlik ölçümlerinin karşilaştirilmasi ve ultrasonik pakimetrenin tekrarlanabilirliğinin değerlendirilmesi. Turk Oftalmoloiji Derg 2014;44:263–7
  8. Fujimura F, Kamiya K, Fujiwara K, Shoji N, Shimizu K. Repeatability and reproducibility of measurements using a NT-530P noncontact tono/pachymeter and correlation of central corneal thickness with intraocular pressure. Biomed Res Int 2013;2013:370592
  9. Kawana K, Tokunaga T, Miyata K, Okamoto F, Kiuchi T, Oshika T. Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis. Br J Ophthalmol 2004;88:466–8
  10. Suzuki S, Oshika T, Oki K, Sakabe I, Iwase A, Amano S, et al. Corneal thickness measurements: Scanning-slit corneal topography and noncontact specular microscopy versus ultrasonic pachymetry. J Cataract Refract Surg 2003;29:1313–8
  11. Sedaghat MR, Daneshvar R, Kargozar A, Derakhshan A, Daraei M. Comparison of central corneal thickness measurement using ultrasonic pachymetry, rotating scheimpflug camera, and scanning-slit topography. Am J Ophthalmol 2010;150:780–9
  12. McLaren JW, N