Are measurement readings of hospital sphygmomanometers accurate?

Hastahane sfigmomanometrelerinin ölçüm değerleri doğru mu?

Pages
108–112
Publication Date
2022-11-05
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

There was a statistical relationship between accuracy and usage duration of the devices (p

📌 3 Key Takeaways

🎯
Objective & Hypothesis:OBJECTIVE: An accurate sphygmomanometer should be used in the diagnosis and follow up of hypertension.
📊
Key Findings & Data:The aims of this study were to evaluate accuracy of hospital blood pressure measuring devices and to inform health care staff measuring blood pressure.MATERIAL and METHODS: The study which included 119 aneroid and seven mercury sphygmomanometers was conducted in 2009.
💡
Conclusion & Impact:There was a statistical relationship between accuracy and usage duration of the devices (p
🔬Design:Survey & Empirical Study
📍Scope:2009 Verileri
⚙️Variables:Domain Parameters

Abstract

OBJECTIVE: An accurate sphygmomanometer should be used in the diagnosis and follow up of hypertension. The aims of this study were to evaluate accuracy of hospital blood pressure measuring devices and to inform health care staff measuring blood pressure.MATERIAL and METHODS: The study which included 119 aneroid and seven mercury sphygmomanometers was conducted in 2009. The knowledge of the staff and information about devices was determined with a questionnaire. The accuracy of the devices was checked by a Y-connection system. In addition, the mechanical features of the devices were also evaluated.RESULTS: A total of 56 staff responded to the questionnaire about 126 devices and 53 staff (95%) knew that sphygmomanometers should be checked for accuracy. Six (11%) staff knew how the devices should be checked. All cuff sizes were standard. 49 (41%) of the aneroid devices were inaccurate. Inaccuracy was greater than 10 mm Hg in only five (4%) of the devices. There was a statistical relationship between accuracy and usage duration of the devices (p

Özet

AMAÇ: Hipertansiyonun tanı ve takibinde doğru ölçen bir sfigmomanometre kullanılmalıdır. Bu çalışmanın amaçları hastahanede bulunan kan basıncı ölçüm aletlerinin doğruluğunu değerlendirmek ve kan basıncını ölçen sağlık personelini bu konuda bilgilendirmektir.GEREÇ ve YÖNTEMLER: 119 aneroid ve 7 adet cıvalı aletin dahil edildiği çalışma 2009 yılında yapıldı. Bir anket uygulanarak personelin konu ile ilgili bilgisi ve aletler hakkında bilgi sorgulandı. Aletlerin doğruluğu bir Y bağlantı sistemi kullanılarak araştırıldı. Ayrıca aletlerin mekanik özellikleri gözden geçirildi.BULGULAR: Toplam 56 çalışan 126 aletle ilgili ankete yanıt verdi, 53 çalışan (%95) sfigmomanometrelerin doğru ölçüm yapıp yapmadığının kontrol edilmesi gerektiğini biliyordu. Aletin doğruluğunun nasıl kontrol edileceğini bilen kişi sayısı ise 6 (%11) idi. Aletlerin hepsinin manşet boyutu standarttı. 119 aneroid aletten 49'u (%41) hatalı idi. Hata, aneroid aletlerin sadece 5 tanesinde (%4) 10 mmHg'nın üzerinde idi. Aneroid aletlerin doğruluğu ile kullanım süresi arasında istatistiksel olarak anlamlı bir fark vardı. (p

🔖

Cite This Article

ERDEM, E., AKPOLAT, T., AYDOĞDU, T., SELİM, N., SARI, A., KADI, R., BİÇEN, C. (2022). Are measurement readings of hospital sphygmomanometers accurate?. Türk Nefroloji Diyaliz ve Transplantasyon Dergisi, Cilt: 19 - Sayı: 2, 108–112. https://ajindex.com/articles/636467

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (13)

  1. Kearney PM, Whelton M, Reynolds K, Muntner P, Whelton PK, He J: Globen burden of hypertension: Analysis of worldwide data. Lancet. 2005
  2. 223 2. Dilek M, Adıbelli Z, Aydoğdu T, Köksal AR, Çakar B, Akpolat T: Self measurement of blood pressure at home: Is it reliable? Blood Pressure 2008
  3. 41 3. Beevers G, Lip GY, O’Brien E: ABC of hypertension. Blood pressure measurement. Part II-conventional sphygmomanometery: Technique of auscultatory blood pressure measurement. BMJ 2001
  4. 322(7293):1043-1047 4. Rouse A, Marshall T: The extent and implications of sphygmomanometer calibration error in primary care. J Hum Hypertens 2001
  5. 591 5. Ashworth M, Gordon K, Baker G, Deshmukh A: Sphygmomanometer calibration: A survey of one inner-city primary care group. J Hum Hypertens 2001
  6. 262 6. Coleman AJ, Steel SD, Ashworth M, Vowler SL, Shennan A: Accuracy of the pressure scale of sphygmomanometers in clinical use within primary care. Blood Press Monit 2005
  7. 188 7. Ali S, Rouse A: Practice audits: reliability of sphygmomanometers and blood pressure recording bias. J Hum Hypertens 2002
  8. 361 8. Burke MJ, Towers HM, O’Malley K, Fitzgerald DJ, O’Brien ET: Sphygmomanometers in hospital and family practice: Problems and recommendations. Br Med J (Clin Res Ed) 1982
  9. 471 9. Basak O, Karazeybek S: Accuracy of sphygmomanometers. Tr J Medical Sciences 1999
  10. 491 10. Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN, Jones DW, Kurtz T, Sheps SG, Rocella EJ: Recommendations for blood pressure measurement in humans and experimental animals. Part I. Blood pressure measurement in humans: A statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Hypertension 2005
  11. 161 11. O’Brien E, Asmar R, Beilin L, Imai Y, Mallion JM, Mancia G, Mengden T, Myers M, Padfi eld P, Palatini P, Parati G, Pickering T, Redon J, Staessen J, Stergiou G, Verdecchia P: European Society of Hypertension Working Group on Blood Pressure Monitoring. European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurement. J Hypertens 2003
  12. 21(5):821-848 12. Campbell NR, Chockalingam A, Fodor JG, McKay DW: Accurate, reproducible measurement of blood pressure. CMAJ 1990
  13. 24 13. Waugh JJ, Gupta M, Rushbrook J, Halligan A, Shennan AH. Hidden errors of aneroid sphygmomanometers. Blood Press Monit 2