Journals / Türk Klinik Biyokimya Dergisi / 2017 / Cilt: 15 - Sayı: 1

Comparison of Parathyroid Hormone Levels Measured with Serum Separator Tubes and Thrombin-Based Rapid Serum Tubes

Serum Separatör Tüp ile Trombin İçeren Hızlı Serum Tüplerinde Analiz Edilen Parathormon Düzeylerinin Karşılaştırılması

Pages
30–35
Publication Date
2022-11-05
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

There was a positive correlation between rapid and serum tubes (p

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objective: Preanalytical variables account for most of laboratory errors and there are many factors that affect and contribute to the results from a patient.
📊
Key Findings & Data:Material and Methods: 20 subjects participated in this study.
💡
Conclusion & Impact:There was a positive correlation between rapid and serum tubes (p
🔬Design:Research Article
📍Scope:2000 Verileri
⚙️Variables:Domain Parameters

Abstract

Objective: Preanalytical variables account for most of laboratory errors and there are many factors that affect and contribute to the results from a patient. In recent years surgeons use intraoperative parathormone (PTH) results to evaluate the efficiency of parathyroidectomy operations. In this study our aim was to compare the PTH results measured with serum separator (SST) and rapid serum tubes (RST). Material and Methods: 20 subjects participated in this study. We waited 30 minutes for serum tubes and 5 minutes for rapid tubes to ensure complete clot formations. After clot formation, all tubes were centrifuged. Parathormone levels were quantified on Siemens Immulite 2000 XPI analyzer. Results: The group consisted of 7 males and 13 females. The mean of parathormone levels for tubes were; 58.9 ± 35.6 pg/mL for serum tubes and 54.4 ± 33.6 pg/mL for rapid tubes. There was statistically significant difference between SST and RST results (p = 0.009). There was a positive correlation between rapid and serum tubes (p

Özet

Amaç: Preanalitik değişkenler laboratuvarlar tarafından verilen sonuçları etkileyen en önemli etkenlerdir. Son yıllarda paratiroidektomi operasyonları sırasında ameliyatın etkinliğini değerlendirmek için intraoperatif Parathormon (PTH) seviyesi kullanılmaktadır. Bu çalışmadaki amacımız, serum separatör tüpleri (SST) ve trombin içeren hızlı serum tüpleri (RST) ile çalışılan PTH sonuçları arasında anlamlı fark olup olmadığını araştırmaktır. Gereç ve Yöntem: Çalışmaya 20 gönüllü dahil edildi. Pıhtı oluşumu için RST 5 dakika, SST 30 dakika bekletildi ardından santrifüj edildi ve Siemens Immulite 2000 XPI analizöründe PTH düzeyleri ölçüldü. Bulgular: Katılımcıların 7'si erkek (%35) 13'ü kadındı (%65). SST için PTH ortalama değeri 58.9 ± 35.6 pg/mL, RST için ise 54.4 ± 33.6 pg/mL olarak bulundu. SST ve RST sonuçları arasında istatistiksel olarak anlamlı fark bulundu (p=0.009). RST ile SST arasında istatistiksel olarak pozitif yönde güçlü bir korelasyon bulundu (p

🔖

Cite This Article

YALÇIN, H., ÇOLAK, A., ARI, E. M., AKŞİT, M. Z., BAYSOY, A., ZENGİN, M. O. (2022). Comparison of Parathyroid Hormone Levels Measured with Serum Separator Tubes and Thrombin-Based Rapid Serum Tubes. Türk Klinik Biyokimya Dergisi, Cilt: 15 - Sayı: 1, 30–35. https://ajindex.com/articles/634403

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (14)

  1. Manor PG. Turnaround times in the laboratory: a review of the literature. Clinical Laboratory Science 1999;12(2):85-89
  2. Chung HJ, Lee W, Chun S, Park HI, Min WK. Analysis of turnaround time by subdividing three phases for outpatient chemistry specimens. Annals of Clinical and Laboratory Science 2009;39(2):144-149
  3. Reichstein E. The importance of preanalytical factors in immunodiagnostic testing. eJIFCC 2003;14:3
  4. Wayne PA. National Committee for Clinical Laboratories Standards(NCCLS) document H18-A2. Procedures of the handling and processing of blood specimens. NCCLS, 1999
  5. Miles RR, Roberts RF, Putnam AR, Roberts WL. Comparison of serum and heparinized plasma samples for measurement of chemistry analytes. Clinical Chemistry 2004;50(9):1704-1706
  6. Er TK, Tsai LY, Jong YJ, Chen BH. Selected analyte values in serum versus heparinized plasma using the SYNCHRON LX PRO assay methods/instrument. Laboratory Medicine 2006;37(12):731-732
  7. Becton Dickinson Vacutainer Blood Collection Tube Product Ordering Information. http://www.bd.com/ vacutainer/products/venous/ordering_info_tubes. asp#specialty_other_plastic. (Accessed October 2016)
  8. Irvin GL, Deriso GT. A new, practical intraoperative parathyroid hormone assay. Am J Surg 1994; 168:466-468
  9. Irvin GL, Solorzano CC, Carneiro DM. Quick Intraoperative Parathyroid Hormone Assay: Surgical Adjunct to Allow Limited Parathyroidectomy, Improve Success Rate, and Predict Outcome World J Surg 2004;28:1287-1292
  10. Yan R, Colantonio D, Wong P-Y, Chen Y. Suitability of Becton Dickinson Vacutainer rapid serum tube for collecting and storing blood samples for antibiotic and anticonvulsant drug monitoring. J Clin Pathol 2014;67:807-810
  11. Yan R, Lou A, Watts G, Tarr H, Smith H, Kinney L, et al. Comparison of Becton Dickinson Vacutainer rapid serum tube with the serum separator tube for routine chemistry and immunoassay tests. J Clin Pathol 2014;67(7):599-604
  12. Ucar F, Erden G, Arzuhal AE, Özdemir S, Öztürk A, Özcan N. Do rapid serum tubes provide comparable test results or improved stability when compared with serum separator tubes? Acta Medica 2015;4:23- 27
  13. Sturgeon CM, Seth J. Why do immunoassays for tumour markers give differing results?-A view from the UK National External Quality Assessment Schemes. Eur J Clin Chem Clin Biochem 1996;34:755-759