Interference of Variant Hemoglobin in Measurement of HBA1C by Ion-Exchange HPLC

İyon Değiştirici HPLC ile HBA1C Ölçümünde Varyant Hemoglobin İnterferansı

Pages
129–133
Publication Date
2022-11-05
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Therefore, in the presence of clinically incompatible HbA1c levels, hemoglobin variants should be investigated.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Background: It has been established that hemoglobinopathies and chemically modified derivatives of hemoglobin, interfered with some HbA1c assay methods either positively or negatively.
📊
Key Findings & Data:In this study we present a case with a low HbA1c value as measured by ion-exchange HPLC method.
💡
Conclusion & Impact:Therefore, in the presence of clinically incompatible HbA1c levels, hemoglobin variants should be investigated.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Background: It has been established that hemoglobinopathies and chemically modified derivatives of hemoglobin, interfered with some HbA1c assay methods either positively or negatively. In this study we present a case with a low HbA1c value as measured by ion-exchange HPLC method. Case presentation: A 55 year old patient admitted to policlinic of Internal Medicine and the result of HbA1c was 2.1%, measured by ion-exchange HPLC method in our laboratory. Unusual peaks and warning about variant hemoglobin were detected in the result chromatogram. Following on suspicion of hemoglobin variant sent to referral laboratory, was determined result of HbD-Los Angeles 84.4%, HbA 6.9%, HbA2 3.9%, HbF 4.8% (Trinity Biotech variant analyzer). Discussion: Variant hemoglobin may cause negative interference in methods such as ion-exchange HPLC, while immunological methods do not cause significant interference in the measurement. Therefore, in the presence of clinically incompatible HbA1c levels, hemoglobin variants should be investigated.

Özet

Giriş: Hemoglobinopatilerin ve hemoglobinin kimyasal olarak modifiye olmuş türevlerinin bazı HbA1c ölçüm yöntemlerini pozitif ya da negatif interfere ettiği ispatlanmıştır. Bu çalışmada iyon değiştirici kromotografi ile HbA1c değeri alt referans sınırın altında saptanan bir olgu sunulmuştur.Olgu sunumu: Dahiliye polikliniğine başvuran 55 yaşındaki hastadan istenen HbA1c testinin laboratuvarımızda iyon değiştirici HPLC metoduyla ölçülen değeri %2.1 idi. Kromatogramda olağan dışı pikler ve varyant hemoglobin uyarısı görüldü. Varyant hemoglobinin değerlendirilmesi amacıyla Trinity Biotech variant analyzer ile HPLC metoduyla yapılan değerlendirme sonucu HbD Los Angeles %84.4, HbA %6.9, HbA2 %3.9, HbF %4.8 olarak saptandı. Tartışma: Varyant Hb'ler immün yöntemlerle ölçümde ciddi bir interferansa neden olmazken, iyon değiştirici HPLC gibi yük farkına göre ölçüm yapılan yöntemlerde negatif interferansa neden olabilirler. Bu nedenle klinik ile uyumlu olmayan HbA1c düzeylerinde hemoglobin varyantlarının varlığı araştırılmalıdır.

🔖

Cite This Article

UÇAR, F., TEMEL, İ., ÇETİN, E., YALÇINDAĞ, A., HANCI, L. T., YAY, F. (2022). Interference of Variant Hemoglobin in Measurement of HBA1C by Ion-Exchange HPLC. Türk Klinik Biyokimya Dergisi, Cilt: 15 - Sayı: 3, 129–133. https://ajindex.com/articles/634397

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (14)

  1. Rahbar S. An abnormal hemoglobin in red cells of diabetics. Clinica Chimica Acta. 1968;22(2):296-8
  2. Sacks DB. Diabetes Mellitus. Burtis CA, Ashwood ER, Bruns DE, eds.Tietz textbook of clinical chemistry and molecular diagnostics. Fifth edition. St. Louis, missouri:Elsevier saunders; 2012. p. 1441
  3. Jeppsson JO, Kobold U, Barr J, Finke A, Hoelzel W, Hoshino T, et al. Approved IFCC reference method for the measurement of HbA1c in human blood. Clin Chem Lab Med. 2002;40(1):78-89
  4. American diabetes association. Standards of medical care in diabetes-2012. Diabetes Care. 2012;35(Suppl):11-63
  5. John WG, Mosca A, Weykamp C, Goodall I. HbA1c standardisation: history, science and politics. Clin Biochem Rev. 2007;28(4):163-8
  6. Gillery P. A history of HbA1c through Clinical Chemistry and Laboratory Medicine. Clin Chem Lab Med. 2013; 51(1):65-74
  7. Thevarajah TM, Nani N, Chew YY. Performance evaluation of the Arkray Adams HA-8160 HbA1c analyser. Malays J Pathol. 2008;30(2): 81-6
  8. Jeppsson JO, Jerntorp P, Sundkvist G, Englund H, Nylund V. Measurement of hemoglobin A1c by a new liquid-chromatographic assay: methodology, clinical utility, and relation to glucose tolerance evaluated. Clin Chem. 1986;32(10):1867-72
  9. Roberts WL, Chiasera JM, Ward-Cook KM. Glycohemoglobin results in samples with hemoglobin C or S trait: a comparison of four test systems. Clin Chem. 1999;45(6 Pt 1):906-9
  10. Klenk DC, Hermanson GT, Krohn RI, Fujimoto EK, Mallia AK, Smith PK, et al. Determination of glycosylated hemoglobin by affinity chromatography: comparison with colorimetric and ion-exchange methods, and effects of common interferences. Clin Chem.1982;28(10):2088-94
  11. Iwamoto H, Motomiya Y, Miura K, Morisawa M, Yoshimura Y, Maruyama I. Immunochemical assay of hemoglobin with N(epsilon)- (carboxymethyl) lysine at lysine 66 of the beta chain. Clin Chem. 2001;47(7):1249-55
  12. Yun YM, Ji M, Ko DH, Chun S, Kwon GC, Lee K, et al. Hb variants in Korea: effect on HbA 1c using five routine methods. Clin Chem Lab Med. 2017;55(8):1234-42
  13. Little RR, Rohlfing CL, Hanson S, Connolly S, Higgins T, Weykamp CW, et al. Effects of Hemoglobin (Hb) E and HbD Traits on Measurements of Glycated Hb (HbA 1c ) by 23 Methods. Clinical Chemistry. 2008;54(8):1277-82
  14. Higgins T, M Sc, Eckfeldt JH, Barton JC, Doumas BT. Hemoglobin, Iron,and Bilirubin. Burtis CA,Ashwood ER,Bruns DE, eds.Tietz textbook of clinical chemistry and molecular diagnostics. Fifth edition, St. Louis, missouri.E