Dört immünolojik ölçüm cihazının tiroid stimüle edici hormon (TSH) ve serbest tiroksin (FT4) arasındaki ilişki yönünden karşılaştırılması

Comparison of four immunoassay analyzers for relationship between thyroid stimulating hormone (TSH) and free thyroxine (FT4)

Pages
88–91
Publication Date
2022-11-05
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

In data sets obtained from four different immunoassay analyzers, poor and statistically different correlation was observed between analyzers at TSH values ranged from 1.0 to 10.0 uIU/mL compared with FT4.Conclusion: These variations between analyzers may affect the clinical decisions especially in the evaluation of subclinical hypothyroidism, clinicians and laboratory specialists should be aware of these situation

📌 3 Key Takeaways

🎯
Objective & Hypothesis:There is an inverse log/linear relationship between TSH and FT4 due to the negative feedback of these hormones on the pituitary.
📊
Key Findings & Data:The objective of our study was to compare this relationship of TSH and FT4 between four different immunoassay analyzers.Methods: In our study, four data sets obtained from the database of four different hospital laboratories each using only a single method of measurement was used to evaluate this relationship between TSH and FT4.These data sets with their assay methods include: 21.102 test results measured by Modular E170 Analyzer (Roche Diagnostics, Germany), 20.241 test results measured by Access DxI 800 Unicel (Beckman Coulter, USA), 22.444 test results measured by Architect i2000sr (Abbott Laboratories, Abbott Park, Illinois, U.S.A) and 20.200 test results measured by ADVIA Centaur XP (Siemens Diagnostics, Tarrytown, NY).
💡
Conclusion & Impact:In data sets obtained from four different immunoassay analyzers, poor and statistically different correlation was observed between analyzers at TSH values ranged from 1.0 to 10.0 uIU/mL compared with FT4.Conclusion: These variations between analyzers may affect the clinical decisions especially in the evaluation of subclinical hypothyroidism, clinicians and laboratory specialists should be aware of these situation
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

There is an inverse log/linear relationship between TSH and FT4 due to the negative feedback of these hormones on the pituitary. The objective of our study was to compare this relationship of TSH and FT4 between four different immunoassay analyzers.Methods: In our study, four data sets obtained from the database of four different hospital laboratories each using only a single method of measurement was used to evaluate this relationship between TSH and FT4.These data sets with their assay methods include: 21.102 test results measured by Modular E170 Analyzer (Roche Diagnostics, Germany), 20.241 test results measured by Access DxI 800 Unicel (Beckman Coulter, USA), 22.444 test results measured by Architect i2000sr (Abbott Laboratories, Abbott Park, Illinois, U.S.A) and 20.200 test results measured by ADVIA Centaur XP (Siemens Diagnostics, Tarrytown, NY). Inverse logarithmic relationship were determined from each data sets and compared between analyzers.Results: The correlation coefficients were -0.439 [95% CI, (-0.450)- (-0.428)], -0.488 [95% CI, (-0.498)- (-0.478)], -0.353 [95% CI, (-0.364)- (-0.342)], -0.430 [95% CI, (-0.441)- (-0.419)] for DxI 800 Unicel, Modular E 170, Architect i2000sr and ADVIA Centaur XP, respectively. In our study, all immunoassay analyzers showed poor correlation in a concentration range of TSH between 1.0 to 10.0 uIU/mL and the inverse log/linear relationship was not observed. Inverse relationship between TSH and FT4 is the result of negative thyroid pituitary hypothalamic feedback mechanism and this relationship is important especially in the diagnosis and treatment of hypo and hyperthyroidism. In data sets obtained from four different immunoassay analyzers, poor and statistically different correlation was observed between analyzers at TSH values ranged from 1.0 to 10.0 uIU/mL compared with FT4.Conclusion: These variations between analyzers may affect the clinical decisions especially in the evaluation of subclinical hypothyroidism, clinicians and laboratory specialists should be aware of these situation

Özet

TSH ve FT4 arasında, bu hormonların hipofiz bezi üzerindeki negatif feedback etkisine bağlı olarak, ters logaritmik lineer bir ilişki mevcuttur. Çalışmamızın amacı TSH ve FT4 arasındaki bu ilişkinin dört farklı immünolojik ölçüm cihazı arasında karşılaştırılmasıdır.Metod: Çalışmamızda, TSH ve FT4 arasındaki bu ilişkinin değerlendirilmesi amacıyla, her biri yalnızca tek bir ölçüm metodunu kullanan hastane laboratuvarlarının veritabanından elde edilmiş veri serileri kullanılmıştır. Bu veri serileri ve ölçüm metotları, Modular E170 Analyzer (Roche Diagnostics, Germany) ile ölçülen 21.102, Access DxI 800 Unicel (Beckman Coulter, USA) ile ölçülen 20.241, Architect i2000sr (Abbott Laboratories, Abbott Park, Illinois, U.S.A) ile ölçülen 22.444, ADVIA Centaur XP (Siemens Diagnostics, Tarrytown, NY) ile ölçülen 20.200 test sonucundan oluşmaktadır. Ters logaritmik ilişki herbir veri serisinde belirlenmiş ve cihazlar arasında karşılaştırılmıştır.Bulgular: Korelasyon katsayıları DxI 800 Unicel için -0.439 [95% CI, (-0.450)- (-0.428)], Modular E 170 için -0.488 [95% CI, (-0.498)- (-0.478)], Architect i2000sr için -0.353 [95% CI, (-0.364)- (-0.342)] ve ADVIA Centaur XP için -0.430 [95% CI, (-0.441)- (-0.419)] saptanmıştır. Çalışmamızda immunolojik ölçüm cihazlarının hepsinde 1.0-10 uIU/mL TSH değerleri için zayıf korelasyon bulunmuş ve ters logaritmik lineer ilişki gözlenmemiştir. TSH ve FT4 arasındaki ters ilişki negatif hipotalamo-hipofizo-tiroidal feedback mekanizmasının sonucu olup, bu ilişki özellikle hipo ve hipertirodizmin tanı ve tedavisinde önemlidir. Dört farklı immunolojik ölçüm cihazından elde edilen veri serilerinde özellikle 1-10 uIU/mL aralığındaki TSH değerleri FT4 ile karşılaştırıldığında cihazlar arasında farklı ve zayıf korelasyon görülmüştür.Sonuç: Cihazlar arasındaki bu farklılıklar özellikle subklinik hipotoroidinin değerlendirilmesinde klinik kararları etkileyebileceğinden klinisyenler ve laboratuvar uzmanları bu durumun farkında olmalıdırlar

🔖

Cite This Article

OZGURTAS, T., SERDAR, M. A., İSPİR, E., CİRACİ, Z., PASAOGLU, H., KURT, I., GULBAHAR, O. (2022). Dört immünolojik ölçüm cihazının tiroid stimüle edici hormon (TSH) ve serbest tiroksin (FT4) arasındaki ilişki yönünden karşılaştırılması. Türk Biyokimya Dergisi, Cilt: 40 - Sayı: 1, 88–91. https://ajindex.com/articles/621161

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (3)

  1. Ladenson PW, Singer PA, Ain KB, Bagchi N, Bigos ST, et al. American Thyroid Association guidelines for detection of thyroid dysfunction. Arch Intern Med 2000; 160(11):1573-5.
  2. Demers LM, Spencer CA. Laboratory medicine practice guide- lines: laboratory support for the diagnosis and monitoring of thy- roid disease. Clin Endocrinol (Oxf) 2003; 58(2):138-40. van Deventer HE, Mendu DR, Remaley AT, Soldin SJ. Inverse log-linear relationship between thyroid-stimulating hormone and free thyroxine measured by direct analog immunoassay and tan- dem mass spectrometry. Clin Chem 2011; 57(1):122-7.
  3. Soldin SJ, Soukhova N, Janicic N, Jonklaas J, Soldin OP. The