Simulation of Water Equivalency of Polymer Gel Dosimeters with GAMOS

Pages
51–58
Publication Date
2022-10-19
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

The mixture sample showing radiological behavior nearest to water over all energies was observed to be 31a (Water, PEG, HEMA, THPC), while the largest difference is observed in 11a (Water, SODIUM CMC, HEMA, THPC).Conclusion: As a result, simulated gel formulations were found to be radiologically water equivalent.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Purpose: Polymer gel materials are used as dosimeters to provide three dimensional radiation dose distribution.
📊
Key Findings & Data:This study aims to determine and compare the water equivalencies of various new polymer gel mixtures using Geant4-based Architecture for Medicine-Oriented Simulations (GAMOS) software program.
💡
Conclusion & Impact:The mixture sample showing radiological behavior nearest to water over all energies was observed to be 31a (Water, PEG, HEMA, THPC), while the largest difference is observed in 11a (Water, SODIUM CMC, HEMA, THPC).Conclusion: As a result, simulated gel formulations were found to be radiologically water equivalent.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Purpose: Polymer gel materials are used as dosimeters to provide three dimensional radiation dose distribution. An ideal gel dosimeter should offer radiological properties equivalent to water. This study aims to determine and compare the water equivalencies of various new polymer gel mixtures using Geant4-based Architecture for Medicine-Oriented Simulations (GAMOS) software program. Methods: To do this, percentage depth dose curves of the simulated polymer gel mixtures were compared with water curves. Different combinations of materials from 2 main classes were used in the mixtures: (1) Sodium Carboxymethyl Cellulose (SODIUM CMC), Polyvinyl Alcohol (PVA) and Polyethylene Glycol (PEG) as gel matrix, (2) 2-Hydroxyethyl Methacrylate (HEMA), 1-Vinyl-2-Pyrrolidinone (VP) and Diethylene Glycol Dimethacrylate (DEGDMA) as monomer. Each mixture also contained ultra-pure water and an antioxidant material called Tetrakis Hydroxymethyl Phosphonium Chloride (THPC). Results: Percentage depth doses of gel mixtures and water was calculated for 6, 18 MV and 50, 70, 100, 150 kVp X rays. The mixture sample showing radiological behavior nearest to water over all energies was observed to be 31a (Water, PEG, HEMA, THPC), while the largest difference is observed in 11a (Water, SODIUM CMC, HEMA, THPC).Conclusion: As a result, simulated gel formulations were found to be radiologically water equivalent.

🔖

Cite This Article

YURT, A., ÖZBAY, T., ÖZSOYKAL, İ. (2022). Simulation of Water Equivalency of Polymer Gel Dosimeters with GAMOS. Journal of Basic and Clinical Health Sciences, Cilt: 4 - Sayı: 1, 51–58. https://ajindex.com/articles/560551

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (16)

  1. Baldock C, De Deene Y, Doran S, et al. Polymer gel dosimetry. Phys Med Biol 2010
  2. R1–R63. [CrossRef ] 2. McJury M, Oldham M, Cosgrove VP, et al. Radiation dosimetry using polymer gels: methods and applications. Br J Radiol 2000
  3. 919–929. [CrossRef ] 3. De Deene Y, De Wagter C, Van Duyse B, et al. Validation of MR-based polymer gel dosimetry as a preclinical three dimensional verification tool in conformal radiotherapy. Magn Reson Med 2000
  4. 116–125. [CrossRef ] 4. Maryanski MJ, Schulz RJ, Ibbott GS, et al. Magnetic resonance imaging of radiation dose distributions using a polymer-gel dosimeter. Phys Med Biol 1994
  5. 1437–1455. [CrossRef ] 5. Brindha S, Venning AJ, Hill B, Baldock C. Experimental study of attenuation properties of normoxic polymer gel dosimeters. Phys Med Biol 2004
  6. N353–N361. [CrossRef ] 6. Hill B, Venning A, Baldock C. The dose response of normoxic polymer gel dosimeters measured using X-ray CT. Br J Radiol 2005
  7. 623–630. [CrossRef ] 7. Zahran RR, Kandeil AY, Higazy AA, Kassem ME. Ultrasonic and thermal properties of gamma irradiated low-density polyethylene. J Appl Poly Sci 1993
  8. 1291–1297. [CrossRef ] 8. Zahran RR. Effects of gamma irradiation on the ultrasonic and structural properties of polyoxymethylene. Mater Lett 1998
  9. [CrossRef ] 9. Maryanski MJ, Zastavker YZ, Gore JC. Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels: II. Optical properties of the BANG polymer gel. Phys Med Biol 1996
  10. [CrossRef ] 10. Oldham M, Siewerdsen JH, Shetty A, Jaffray DA. High resolution gel-dosimetry by optical-CT and MR scanning. Med Phys 2001
  11. 1436–1445. [CrossRef ] 11. Maryanski MJ, Ranade MK. Laser micro-beam scanning of dosimetry gels. Proc SPIE 4320 2001
  12. 764–774. [CrossRef ] 12. Lepage M, Whittaker AK, Rintoul L, Back SA, Baldock C. Modelling of post-irradiation events in polymer gel dosimeters. Phys Med Biol 2001
  13. 2827–2839. [CrossRef ] 13. Rintoul L, Lepage M, Baldock C. Radiation dose distribution in polymer gels by Raman spectroscopy. Appl Spectrosc 2003
  14. 51–57. [CrossRef ] 14. Taylor ML, Kron T, Franich RD. A contemporary review of stereotactic radiotherapy: Inherent dosimetric complexities and the potential for detriment. Acta Oncol 2011
  15. 483–508. [CrossRef ] 15. Sigma Aldrich. Acrylamide properties in product catalogue
  16. http://www.sigmaaldrich.com/catalog/product/sigma/a9099?lang=en®ion=TR 16. Rogers DWO. Fifty years of Monte Carlo simulatio