Accuracy of Corneal Power Measurement Using Scheimpfl ug Camera After Myopic Refractive Surgery

Pages
5–11
Publication Date
2022-10-18
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Using the Pearson correlation, 4.5 mm MPP, 5.0 mm MPP and 5.5 mm MPP values were signifi cantly correlated with the CHM (r: 0.907, 0.906 and 0.905, respectively; p

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Purpose: To investigate the accuracy of ray tracing (RT) using the Sirius rotating Scheimpfl ug camera (Sirius), combined with Placido disc corneal topography (CSO, Florence, Italy), to measure corneal power (CP) in eyes which had undergone excimer laser surgery for myopia.
📊
Key Findings & Data:Materials and Methods: We retrospectively reviewed the medical records of 59 eyes of 31 patients who had undergone refractive laser surgery for myopia.
💡
Conclusion & Impact:Using the Pearson correlation, 4.5 mm MPP, 5.0 mm MPP and 5.5 mm MPP values were signifi cantly correlated with the CHM (r: 0.907, 0.906 and 0.905, respectively; p
🔬Design:Review & Systematic Survey
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Purpose: To investigate the accuracy of ray tracing (RT) using the Sirius rotating Scheimpfl ug camera (Sirius), combined with Placido disc corneal topography (CSO, Florence, Italy), to measure corneal power (CP) in eyes which had undergone excimer laser surgery for myopia. Materials and Methods: We retrospectively reviewed the medical records of 59 eyes of 31 patients who had undergone refractive laser surgery for myopia. We collected different CP values calculated by RT according to entrance pupil diameter (ranging from 2.5 to 7.0 mm), which was called mean pupil power (MPP) by the manufacturer, and simulated keratometry (simK) value using the Sirius. We then compared these values with those obtained with the clinical history method (CHM). Results: The mean refractive change after laser surgery was 3.14 ± 1.70 diopters (D). Using one-way analysis of variance with Dunnett’s multiple comparisons test; 4.5 mm MPP, 5.0 mm MPP and 5.5 mm MPP values were similar to those obtained with the CHM (40.31 ± 1.99 D) [p> 0.05]. Using the Pearson correlation, 4.5 mm MPP, 5.0 mm MPP and 5.5 mm MPP values were signifi cantly correlated with the CHM (r: 0.907, 0.906 and 0.905, respectively; p

🔖

Cite This Article

YASSA, E. T., UNLU, C. (2022). Accuracy of Corneal Power Measurement Using Scheimpfl ug Camera After Myopic Refractive Surgery. Glokom Katarakt, Cilt: 16 - Sayı: 1, 5–11. https://ajindex.com/articles/525745

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (16)

  1. Haigis W. Intraocular lens calculation after refractive surgery for myopia: Haigis-L formula. J Cataract Refract Surg 2008;34:1658- 63
  2. Seitz B, Langenbucher A, Nguyen NX, et al. Underestimation of intraocular lens power for cataract surgery after myopic photorefractive keratectomy. Ophthalmology.1999;106:693-702
  3. Can Ç, İlhan B, Polat S, ve ark. Refraktif Cerrahi Sonrası İntraoküler Lens Gücü Hesaplanması ve Teknolojik Gelişmeler. Turk J Ophthalmol; 2006; 36:384-8
  4. Utine CA, Çakır H, Çakır M, ve ark. Keratorefraktif Cerrahi Geçirmiş Gözlerde Göz İçi Lens Gücü Hesaplanması. Turk J Ophthalmol; 2007; 37:427-32
  5. Altınkurt E, Bayhan HA. Katarakt Cerrahisinde Güncel Biyometri Ölçüm Teknikleri. Şencan S editör. Multifokal Göz İçi Lensleri
  6. Baskı. Ankara: Türkiye Klinikleri; 2019:12-21
  7. Holladay JT. In discussion of: determining the power of an intraocular lens to achieve a postoperative correction of −1
  8. Refract Corneal Surg.1989;5:202-3
  9. Savini G, Barboni P, Zanini M. Intraocular lens power calculation after myopic refractive surgery: theoretical comparison of different methods. Ophthalmology 2006;113:1271-82
  10. Hoffer KJ. Intraocular lens power calculation for eyes after refractive keratotomy. J Refract Surg. 1995;11:490-3
  11. Koch D, Wang L. Calculating IOL power in eyes that have had refractive surgery [editorial]. J Cataract Refract Surg 2003;29:2039-42
  12. Shammas HJ, Shammas MC, Garabet A, et al. Correcting the corneal power measurements for intraocular lens power calculations after myopic laser in situ keratomileusis. Am J Ophthalmol 2003;136:426-32
  13. Srivannaboon S, Reinstein DZ, Sutton Hf, et al. Accuracy of Orbscan total optical power maps in detecting refractive change after myopic laser in situ keratomileusis. J Cataract Refract Surg 1999;25:1596-9. 12 Savini G, Bedei A, Barboni P, et al. KJ.Intraocular lens power calculation by ray-tracing after myopic excimer laser surgery. Am J Ophthalmol 2014;157:150-3
  14. Schuster AK, Schanzlin DJ, Thomas KE, et al. Intraocular lens calculation adjustment after laser refractive surgery using Scheimpfl ug imaging. J Cataract Refract Surg 2016;42:226-31
  15. Baradaran-Rafi i A, Fekri S, Rezaie M, et al. Accuracy of different topographic instruments in calculating corneal power after myopic photorefractive keratectomy. J Ophthalmic Vis Res 2017;12;254-9
  16. Pan C, Hua Y, Huang J, et al. Corneal Power Measurement With the Dual Scheimpfl ug-Placido Topographer After Myopic Excimer