Journals / Eurasian Journal of Medical Investigation / 2021 / Cilt: 5 - Sayı: 3

Comparison of Phacoemulsification Parameters Using Active Fluidic Pressure Control System versus Gravity-Fluidic Syste

Pages
309–312
Publication Date
2022-10-12
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Conclusion: AFPCS was found to be more successful than GFS, with low-energy consumption, shorter surgical timeand better postoperative outcomes.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objectives: This study aimed to compare the mean surgical time, cumulative dissipated energy (CDE), aspiration timeand estimated aspiration fluid parameters in phacoemulsification surgery using an active-fluidic pressure control sys tem (AFPCS) in comparison with gravity-fluidic system (GFS).
📊
Key Findings & Data:Methods: The study analysed 202 eyes of 202 patients diagnosed with senile nuclear cataract and admitted to ourclinic.
💡
Conclusion & Impact:Conclusion: AFPCS was found to be more successful than GFS, with low-energy consumption, shorter surgical timeand better postoperative outcomes.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Objectives: This study aimed to compare the mean surgical time, cumulative dissipated energy (CDE), aspiration timeand estimated aspiration fluid parameters in phacoemulsification surgery using an active-fluidic pressure control sys tem (AFPCS) in comparison with gravity-fluidic system (GFS). Methods: The study analysed 202 eyes of 202 patients diagnosed with senile nuclear cataract and admitted to ourclinic. Phacoemulsification surgery was performed by an experienced surgeon using the Centurion Vision System de vice and the Infiniti Vision System device. After each case, the mean surgical time, CDE, aspiration time and estimatedaspiration fluid parameters were recorded. Complicated cases were excluded from the study. Results: In the AFPCS group, the mean age of the patients was 64.15±3.83 years, with the following results: mean surgi cal time, 11.88±5.44 min; CDE, 11.63±9.37; aspiration time, 3.91±1.49 min; estimated aspiration fluid, 71.89±27.56 ml. Inthe GFS group, the mean age of the patients was 64.99±3.80 years, and the following results were obtained: mean sur gical time, 14.35±3.93 min; CDE, 13.60±7.61; aspiration time, 5.56±1.8 min; estimated aspiration fluid, 101.23±15.62 ml. Conclusion: AFPCS was found to be more successful than GFS, with low-energy consumption, shorter surgical timeand better postoperative outcomes.

🔖

Cite This Article

BULUT, A. (2022). Comparison of Phacoemulsification Parameters Using Active Fluidic Pressure Control System versus Gravity-Fluidic Syste. Eurasian Journal of Medical Investigation, Cilt: 5 - Sayı: 3, 309–312. https://ajindex.com/articles/508572

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (3)

  1. Çetinkaya S, Uçar F. Matür kataraktı olan hastalarda femtusaniye lazer fakoemülsifikasyon cerrahisi. Bozok Tıp Dergisi 2020;10:126–31
  2. Leon P, Umari I, Mangogna A, Zanei A, Tognetto D. An evaluation of intraoperative and postoperative outcomes of torsional mode versus longitudinal ultrasound mode phacoemulsification: a Meta-analysis. Int J Ophthalmol 2016;9:890–7
  3. Hida WT, Tzelikis PF, Vilar C, Chaves MAPD, Motta AFP, Carricondo PC, et al. Outcomes study between femtosecond laserassisted cataract surgery and conventional phacoemulsification surgery using an active fluidics system. Clin Ophthalmol 2017;11:1735–9. [CrossRef] 4. Blehm C, Potvin R. Evaluating refractive and visual outcomes after bilateral implantation of an apodized diffractive multifocal toric intraocular lens with a moderate add in the dominant eye and a higher add in the fellow eye. Clin Ophthalmol 2020;14:1035–41. [CrossRef] 5. Tekin K, Çıtırık M, Uzel M, İlhan Ç. Fakoemülsifikasyon cihazı, çalışma prensipleri ve parametreler. Glokom Katarakt 2016;11:73–84 6. Christakis PG, Braga-Mele RM. Intraoperative performance and postoperative outcome comparison of longitudinal, torsional, and transversal phacoemulsification machines. J Cataract Refract Surg 2012;38:234–41. [CrossRef] 7. Solomon KD, Lorente R, Fanney D, Cionni RJ. Clinical study using a new phacoemulsification system with surgical intraocular pressure control. J Cataract Refract Surg 2016;42:542– 9. [CrossRef] 8. Khokhar S, Aron N, Sen S, Pillay G, Agarwal E. Effect of balanced phacoemulsification tip on the outcomes of torsional phacoemulsification using an active fluidics system. J Cataract Refract Surg 2017;43:22–8. [CrossRef] 9. Chen M, Anderson E, Hill G, Chen JJ, Patrianakos T. Comparison of cumulative dissipated energy between the Infiniti and Centurion phacoemulsification systems. Clin Ophthalmol 2015;9:1367–72. [CrossRef] 10. Malik PK, Dewan T, Patidar AK, Sain E. Effect of IOP based infusion system with and without balanced phaco tip on cumulative dissipated energy and estimated fluid usage in comparison to gravity fed infusion in torsional phacoemulsification. Eye Vis (Lond) 2017;4:22. [CrossRef] 11. Dasgupta S, Mehra R. Comparative studies between longitudinal and torsional modes in phacoemulsification, using active fluidics technology along with the intrepid balanced tip. Indian J Ophthalmol 2018;66:1417–22. [CrossRef] 12. Gonzalez-Salinas R, Garza-Leon M, Saenz-de-Viteri M, Solis-S JC, Gulias