Comparison of the sodium, potassium, chloride and glucose levels measured by a blood gas analyzer and an autoanalyzer

Pages
2639–2644
Publication Date
2022-10-12
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Conclusion: We concluded that the Na+ and Cl- results obtained from BGA can be used instead of those obtained from an AA; however, the blood gas K+ and glucose results of the former cannot replace those of the latter.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Aim: A blood gas analyzers (BGA) are vital equipment frequently used in emergency departments and intensive care units.
📊
Key Findings & Data:This study aimed to compare the sodium (Na+), potassium (K+), chloride (Cl-) and glucose values in venous blood samples measured with a BGA and a standard AA with external quality control values.Material and Methods: The results of patients that presented to our emergency department between April 1, 2019 and July 1, 2019 and underwent the measurements of Na+ (n = 5,908), K+ (n = 5,755), Cl- (n = 5,101) and glucose (n = 5,871) simultaneously by BGA and AA were retrospectively compared.
💡
Conclusion & Impact:Conclusion: We concluded that the Na+ and Cl- results obtained from BGA can be used instead of those obtained from an AA; however, the blood gas K+ and glucose results of the former cannot replace those of the latter.
🔬Design:Research Article
📍Scope:2019 Verileri
⚙️Variables:Domain Parameters

Abstract

Aim: A blood gas analyzers (BGA) are vital equipment frequently used in emergency departments and intensive care units. It is clinically important that the measurements of a BGA and an autoanalyzer (AA) provide equivalent results, which is confirmed by their proximitiy to the absolute value. This study aimed to compare the sodium (Na+), potassium (K+), chloride (Cl-) and glucose values in venous blood samples measured with a BGA and a standard AA with external quality control values.Material and Methods: The results of patients that presented to our emergency department between April 1, 2019 and July 1, 2019 and underwent the measurements of Na+ (n = 5,908), K+ (n = 5,755), Cl- (n = 5,101) and glucose (n = 5,871) simultaneously by BGA and AA were retrospectively compared. Results: In the Spearman correlation analysis between the two measurements, the correlation coefficient (r) was found as 0.78, 0.88, 0.89 and 0.97 for Na+, K+, Cl- and glucose, respectively. According to the Bland-Altman analysis, in the comparison of Na+, K+, Cl- and glucose values, the average bias percentages at the 95% confidence interval were -0.8 (4.8 to -6.4), -9 (8.6 to -26.5), -0.2 (6.2 to -6.5), and 0.3 (21.8 to -21.3), respectively. Conclusion: We concluded that the Na+ and Cl- results obtained from BGA can be used instead of those obtained from an AA; however, the blood gas K+ and glucose results of the former cannot replace those of the latter.

🔖

Cite This Article

YÜCEL, K., KAYA, Ö. (2022). Comparison of the sodium, potassium, chloride and glucose levels measured by a blood gas analyzer and an autoanalyzer. Annals of Medical Research, Cilt: 27 - Sayı: 10, 2639–2644. https://ajindex.com/articles/446124

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (15)

  1. Aygencel G. Arter kan gazlarının yorumlanması. Arch Turk Soc Cardiol 2014;42:194-202
  2. Ayhan H, Erdogan MO, Yigit Y, et al. The Reliability of Blood Gas Electrolytes. JAEM 2014;13:49-52
  3. Oliveira B, Prasanna M, Lemyze M, et al. A comparison berween measured and calculated central venous oxygen saturation in critically ill patients. Plos One. 2018;8:13
  4. Lee JW. Fluid and electrolyte disturbances in critically ill patients. Electrolyte Blood Press 2010;8:72-81
  5. Scott MG, LeGrys VA, Klutts JS. Electrolytes and blood gases. In: Burtis DE, ashwood ER, bruns DE, editors. Tietz Textbook of Clinical Chemistry and Molecular Diagnostics. 4th edition. St Louis, MO; Elsevier.2006
  6. Rang LC, Murray HE, Wells GA, et al. Can peripheral venous blood gases replace arterial blood gases in emergency department patients? CJEM 2002;4:7-15
  7. Bakoglu E, Kebapcioğlu AS, Ak A, et al. Acil Serviste Periferik Venöz Kan Gazının Arter Kan Gazı Yerine Kullanılabilirliğinin Araştırılması. Eur J Basic Med Sci 2013;3:29-33
  8. Budak YU, Huysal K, Polat M, Use of a blood gas analyzer and a laboratory autoanalyzer in routine practice to measure electrolytes in intensive care unit patients. BMC Anesthesiol 2012:3;12-7
  9. Memisogullari R, Ozcan ME, Celbek G, et al. Kan gazı analiz cihazı ve direkt enzimatik ölçüm metodu ile elde edilen bikarbonat değerlerinin korelasyonu. Turk J Biochem 2011;36;270-2
  10. Kozancı N, Gungor F, Ay MO, et al. Can venous blood gas values be used instead of arterial blood gas values in respiratory alkalosis? Turk J Biochem 2014;39;113- 8
  11. Kelly AM. Review article: Can venous blood gas analysis replace arterial in emergency medical care. Emerg Med Australas 2010;22:493-8
  12. Bradenburg MA, Dire DJ. Comparison of arterial and venous blood gas volues in the emergency department evaluation of patients with diabetic ketoasidosis. Ann Emerg Med 1998;31:459-65
  13. Zhang JB, Lin J, Zhao XD. Analysis of bias in measurements of potassium, sodium and hemoglobin by an emergency department-based blood gas analyzer relative to hospital laboratory autoanalyzer results. PLoS One 2015;7;10:0122383
  14. Ustundag BY, Huysal K. Application of sigma metrics and performance comparison between two biochemistry analyser and a blood gas analyser for the determination of electrolytes. J Clin Diagn Res 2017;11:6-9
  15. Kozaci N, Ay MO, Guven R, et al. Comparison of Na, K, Cl, Hb and Hct values measured by blood gas analyzer and laboratory auto-analyzer, Turkish J of