Journals / Ankara Üniversitesi Tıp Fakültesi Mecmuası / 1997 / Cilt: 50 - Sayı: 3

Systolic blood pressure measurement with pulse oximetry

Pulse oksimetre kullanılarak ölçülen sistolik kan basıncı değerlerinin diğer tekniklerle karşılaştırılması

Pages
153–157
Publication Date
2022-10-10
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

As a result we suggested that this technique is safe and easy adaptable for patients with Takayasu's syndrome where conventional techniques often fail to monitor SBP.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:We planned this study to determine whether there was a correlation between the SBP values ge¬nerated from movements of pointer of manometer and Korotokoff sound and systolic arterial blood pressures obtained by pulse oximetry.
📊
Key Findings & Data:All patients were monitored with ECG, pulse oximetry (DA TAS-COPE 3000A) and noninvasive blood pressure cuff.
💡
Conclusion & Impact:As a result we suggested that this technique is safe and easy adaptable for patients with Takayasu's syndrome where conventional techniques often fail to monitor SBP.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

We planned this study to determine whether there was a correlation between the SBP values ge¬nerated from movements of pointer of manometer and Korotokoff sound and systolic arterial blood pressures obtained by pulse oximetry. All patients were monitored with ECG, pulse oximetry (DA TAS-COPE 3000A) and noninvasive blood pressure cuff. The SBP was measured by three techniques: Using the plethismographic waveform of the pulse oxi¬metry, by Korotokoff sound and with oscillometric noninvasive blood pressure equipment. The SBP was measured in the same arm of each patient whi¬le the subject was at rest and in the sitting position with three different techniques. Statistical analysis was done with Pearson-Correlation analysis met¬hod. We determined a good correlation between the values of SBP-Korotokoff and SBP-manometer and SBP-disappearance either sitting or resting position. As a result we suggested that this technique is safe and easy adaptable for patients with Takayasu's syndrome where conventional techniques often fail to monitor SBP.

Özet

Biz bu çalışmayı sistolik kan basıncı ölçülme¬sinde pulse oksimetrenin pletismografik dalgaların¬dan yararlanmak ve geleneksel yöntemlerle arasında korelasyon olup olmadığını saptamak amacı ile planladık. Hastalar EKG, pulse oksimetre ( DATASCOPE 3000A ) ve non-invaziv tansiyon aleti ile monitorize edildiler. SKB ölçülmesinde 3 ayrı teknik kullanıldı: Pulse oksimetre'nin pletismografik dalgalan kullanı¬larak, Korotokoff sesleri ile ve tansiyon aletinin ma¬nometre ibre hareketine göre. Her hastada ölçümler yatar ve oturur pozisyonlarda tekrarlandı. Pulse oksimetrenin pletismografik dalgaları kullanılarak ölçülen SKB-kaybolan ve SKB-ortalama değerleri ile SKB-Korotokoff ve SKB-manometre ara¬sında iyi korelasyon tespit edildi. Sonuç olarak bu tekniğin Takayasu's sendromu gibi periferik nabız alınamayan hastaların takibinde pratik ve güvenilir bir yöntem olduğunu söyleyebili¬riz.

🔖

Cite This Article

ADSAN, H., ERDEM, D., BÜYÜKKOÇAK, Ü., TULUNAY, M., LEBLEBİCİ, F. (2022). Systolic blood pressure measurement with pulse oximetry. Ankara Üniversitesi Tıp Fakültesi Mecmuası, Cilt: 50 - Sayı: 3, 153–157. https://ajindex.com/articles/444299

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...