Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples

Pages
161–165
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Conclusion: A saline replacement technique may be a practical solution to resolve the difficulty of the matrix interfer ence effect seen in chemotherapy patient samples and provide a WBC differential count.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objectives: An automated complete blood cell count (hematology) analyzer cannot measure the white blood cell (WBC) differential of some patient sample, notably those of chemotherapy patients.
📊
Key Findings & Data:A technique to use isotonic sodium chloride solution (0.9% saline) as a replacement for plasma has previously been described.
💡
Conclusion & Impact:Conclusion: A saline replacement technique may be a practical solution to resolve the difficulty of the matrix interfer ence effect seen in chemotherapy patient samples and provide a WBC differential count.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

Objectives: An automated complete blood cell count (hematology) analyzer cannot measure the white blood cell (WBC) differential of some patient sample, notably those of chemotherapy patients. This is often due to the presence of atypical cells in the sample or matrix changes caused by drugs. A technique to use isotonic sodium chloride solution (0.9% saline) as a replacement for plasma has previously been described. The aim of this study was to evaluate to use of saline replacement as a means to resolve the matrix interference effect and achieve a reliable WBC differential count for chemotherapy patients. Methods: Samples of 29 chemotherapy patients whose WBC differential count could not be calculated using a Beck man Coulter LH-780 hematology autoanalyzer (Beckman Coulter, Inc., Brea, CA, USA) were evaluated. A peripheral blood smear was performed and the saline replacement technique was applied and the samples were then re-analyzed using the same autoanalyzer. The WBC count and differential count of WBC of a peripheral smear and a saline-replace ment sample were compared. Results: There was no statistically significant difference between the peripheral blood smear and saline replacement autoanalyzer measurements of the WBC differential count and percent of WBC. Strong relationships in WBC (r=0.99), neutrophil (r=0.98), lymphocyte (r=0.98), monocyte (r=0.91), and eosinophil (r=0.77) counts were observed in a com parison of the peripheral blood smear and saline replacement measurements. Conclusion: A saline replacement technique may be a practical solution to resolve the difficulty of the matrix interfer ence effect seen in chemotherapy patient samples and provide a WBC differential count.

🔖

Cite This Article

YAVUZ, H. B., YAMAN, H., TAT, M., AKIDAN, O., ÖREM, A., KARAHAN, S. C., ALİYAZICIOĞLU, Y. (2021). Saline replacement as a practical solution to matrix interference effect when a leukocyte differential count cannot be measured in chemotherapy patient samples. International Journal of Medical Biochemistry, Cilt: 4 - Sayı: 3, 161–165. https://ajindex.com/articles/413635

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (14)

  1. McPherson RA, Pincus MR, editors. Henry’s Clinical Diagnosis and Management by Laboratory Methods. 22nd Ed. Philadelphia: Elsevier Saunders; 2011
  2. Kim JE, Kim BR, Woo KS, Han JY. Evaluation of the leukocyte differential on a new automated flow cytometry hematology analyzer, Int J Lab Hematol 2012;34(5):547–50
  3. U.S. Department of the Interior, Bureau of Reclamation, Quality Assurance Guidelines for Environmental Measurements, 1998. https://www.usbr.gov/tsc/techreferences/mands/ mands-pdfs/QualityAssuranceGuidelines_2003.pdf (Accessed May 5th 2018
  4. Kealey D, Haines PJ. Analytical Chemistry, its Functions and Applications. In: Hames BD, editor. BIOS Instant Notes in Analytical Chemistry, 1st Ed. London: Taylor & Francis; 2002. p.1–4
  5. Ozer H. The timing of chemotherapy-induced neutropenia and its clinical and economic impact. Oncology (Williston Park) 2006;20(5 Suppl 4):11–5
  6. Dasgupta A, Sepuldeva JL. Accurate Results in the Clinical Laboratory A Guide to Error Detection and Correction. 1st Ed. St. Louis, MO: Elsevier; 2013
  7. Waider KL. Rouleaux and saline replacement. Immunohematology 2018;34(3):91-2
  8. Zandecki M, Genevieve F, Gerard J, Godon A. Spurious counts and spurious results on haematology analysers: a review. Part II: white blood cells, red blood cells, haemoglobin, red cell indices and reticulocytes. Int J Lab Hematol 2007;29(1):21– 41
  9. COULTER LH 780 System, Reference. Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA 92821 U.S.A. https://www.beckmancoulter.com/download/file/wsr-32377/773021AD?type- =pdf. Accessed January 5th 2018.
  10. Substantial Equivalence Determinatıon Decision Summary Assay And Instrument Combinatıon Template. https://www. accessdata.fda.gov/cdrh_docs/reviews/K081930.pdf (Accessed January 15th 2018.
  11. Liu Y, Du X, Chen J, Jin Y, Peng L, Wang HHX, Luo M, Chen L, Zhao Y. Neutrophil-to-lymphocyte ratio as an independent risk factor for mortality in hospitalized patients with COVID-19. J Infect 2020;81(1):e6–12
  12. Ozarda Y, Ichihara K, Bakan E, Polat H, Ozturk N, Baygutalp NK, et al. A nationwide multicentre study in Turkey for establishing reference intervals of haematological parameters with novel use of a panel of whole blood. Biochem Med (Zagreb) 2017;27(2):350–77
  13. Zeng SG, Zeng TT, Jiang H, Wang LL, Tang SQ, Sun YM, et al. A simple, fast correction method of triglyceride interference in blood hemoglobin automated measurement. J Clin Lab Anal 2013;27(5):341–5
  14. Er TK, Ginés MAR, ChenYT, T