Journals / ACADEMIC PLATFORM-JOURNAL OF ENGINEERING AND SCIENCE / 2020 / Cilt: 8 - Sayı: 2

Calibration of Standard Platinum Resistance Thermometers in Triple Point of Water Cell and Evaluation of Measurement Uncertainty

Standart Platin Direnç Termometrelerin Suyun Üçlü Noktası Hücresinde Kalibrasyonu ve Ölçüm Belirsizliği Hesabı

Pages
403–406
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

The measurement uncertainty budget was evaluated by analyzing the solid, liquid and gas phases in the cell at the same time with the solid CO2 method within 15 minutes.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:In this study, detailed information about the preparation of triple point of water cell used in the calibration and control of resistance thermometers and calculation of measurement uncertainty is given.
📊
Key Findings & Data:In the preparation of the cell, ice mantle was obtained using solid CO2.
💡
Conclusion & Impact:The measurement uncertainty budget was evaluated by analyzing the solid, liquid and gas phases in the cell at the same time with the solid CO2 method within 15 minutes.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

In this study, detailed information about the preparation of triple point of water cell used in the calibration and control of resistance thermometers and calculation of measurement uncertainty is given. The triple point of water cell was designed and produced in Yeditepe University Physics Department Metrology and Calibration Laboratory (YUKAL). In the preparation of the cell, ice mantle was obtained using solid CO2. The measurement uncertainty budget was evaluated by analyzing the solid, liquid and gas phases in the cell at the same time with the solid CO2 method within 15 minutes.

Özet

Bu çalışmada direnç termometrelerin kalibrasyonların da ve kontrollerinde kullanılan suyun üçlü noktası hücresinin hazırlanması ve ölçüm belirsizliğinin hesaplanmasına ilişkin detaylı bilgi verilmiştir. Suyun üçlü noktası hücresi Yeditepe Üniversitesi Fizik Bölümü Metroloji ve Kalibrasyon Laboratuvarında (YUKAL) özel olarak tasarlanarak üretilmiştir. Hücrenin hazırlanması aşamasında katı CO2 kullanılarak buz tabakası elde edilmiştir. Katı CO2 yöntemiyle 15 dakikalık sürede hücre içerisinde katı, sıvı ve gaz fazlarının aynı anda elde edildiği bu yöntemle yapılan ölçümler analiz edilerek ölçüm belirsizliği bütçesi oluşturulmuştur.

🔖

Cite This Article

ALPER, F. M. P. (2020). Calibration of Standard Platinum Resistance Thermometers in Triple Point of Water Cell and Evaluation of Measurement Uncertainty. ACADEMIC PLATFORM-JOURNAL OF ENGINEERING AND SCIENCE, Cilt: 8 - Sayı: 2, 403–406. https://ajindex.com/articles/354917

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (11)

  1. H. P. Thomas, The International Temperature Scale of 1990 (ITS-90), Metrologia, cilt 27, no. 1, 1990
  2. T. J. Quinn, Temperature, Academic Press, 2013
  3. T. D. McGee, Principles and Methods of Temperature Measurement, New York, United States: John Wiley & Sons Inc, 1988
  4. R. E. Bentley, Handbook of Temperature Measurement.: Volume 1, Temperature and Humidity Measurement, Springer Science & Business Media, 1998
  5. A. Peruzzi, E. Mendez-Lango, J. Zhang ve M. Kalemci, Guide to the Realization of the ITS-90 Triple Point of Water, Bureau International des Poids et Mesures, 2017 [6] A. Peruzzi, M. Dobre, J. v. Geel, A. Uytun, Kalemci, M. Kalemci, E. Uysal, G. Strouse ve Y. Nue, Effect of Impurities on Water Triple-Point Cells, International Journal of Thermophysics, no. 35, 1084–1096
  6. G. T. Furukawa, B. W. Mangum ve G. F. Strouse, Effects of different methods of preparation of ice mantles of triple point of water cells on the temporal behaviour of the triplepoint temperatures, Metrologia, cilt 34, p. 215, 1997
  7. J. V. Nicholas, T. D. Dransfield ve D. R. White, Isotopic composition of water used for triple point of water cells, Metrologia, cilt 33, no. 3, 1996
  8. D. R. White ve W. L. Tew, Improved Estimates of the Isotopic Correction Constants for the Triple Point of Water, International Journal of Thermophysics, no. 31, p. 1644– 1653, 2010
  9. G. T. Furukawa, B. W. Mangum ve G. F. Strouse, Effects of different methods of preparation of ice mantles of triple point of water cells on the temporal behaviour of the triple-point temperatures, Metrologia, cilt 34, p. 215, 1997
  10. A. Peruzzi, M. Dobre, G. Strouse, J. Van Geel, Doping Experiments with Water Triple-Point Cells, Int J Thermophys (2011) 32:1553–1562 [12] K. N. Quelhas, · R. N. Teixeira, An Alternative Method for Manufacturing Water Triple-Point Cells, Int J Thermophys (2014) 35:1097–1108 [13] A. İnce, A. Kartal, Uluslararası Sıcaklık Ölçeği 1990 (ITS-90)'nın Ulusal Metroloji Enstitüsünde (Ume) Gerçekleştirilmesi Ve Muhafazası, TMMOB II. Ulusal Ölçümbilim Kongresi Bildiriler Kitabı, 23 - 24 EKİM 1997, Eskişehir, [14] EA-4/02 Evaluation of the Uncertainty of Measurement in Calibration, EA Laboratory Committee, 2013 [15] S. Armstrong, The Development of a Primary Laboratory for the Calibration of Standard Platinum Resistance Thermometers, AIP Conference Proceedings 684, 351 (2003
  11. J.J. Connolly, R.E. Bentley, Platinum Resistance Thermometry, First Edition, NMI, 2004 [17] J.