Journals / Acta Orthopaedica et Traumatologica Turcica / 2015 / Cilt: 49 - Sayı: 5

Reliability of threaded locking screws

Pages
552–557
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

All 5-mm unthreaded screws and 5.5-mm low threaded stainless steel screws can be used safely in full weight-bearing conditions of unreliable patients

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objective: A frequent problem for interlocking nailing that affects the treatment of the fracture is locking screw deformation.
📊
Key Findings & Data:The aim of this study is to determine whether bending resistance is different between high, low, and unthreaded locking screws of interlocking femoral nails.Methods: Ninety screws were used in this experimental study, with 10 screws used in each of 9 groups.
💡
Conclusion & Impact:All 5-mm unthreaded screws and 5.5-mm low threaded stainless steel screws can be used safely in full weight-bearing conditions of unreliable patients
🔬Design:Research Article
📍Scope:2071 Verileri
⚙️Variables:Domain Parameters

Abstract

Objective: A frequent problem for interlocking nailing that affects the treatment of the fracture is locking screw deformation. The aim of this study is to determine whether bending resistance is different between high, low, and unthreaded locking screws of interlocking femoral nails.Methods: Ninety screws were used in this experimental study, with 10 screws used in each of 9 groups. Three-point bending tests were performed on 6 groups of 5 mm screws (titanium, stainless steel, crossed with unthreaded, low threaded, and high threaded) and the same 3 thread types of 5.5 mm stainless steel screws in a 30-mm inner diameter steel tube, imitating the level of the lesser trochanter. An axial compressor was used to determine the yield points for permanent deformation in the locking screws by way of 3-point bending tests.Results: The mean yield point value of the 3-point bending tests of 5-mm low threaded stainless steel locking screws was 2071 N, 53% less than that of unthreaded screws (3169 N). The mean yield point value of 5-mm high threaded stainless steel locking screws was 556 N, 272% less than that of low threaded screws (2071 N).Conclusion: To avoid locking screw deformation, high threaded screws must not be used as locking screws. In cases of unreliable patients, 5-mm low threaded screws should not be used in the nailing of comminuted or oblique femur shaft fractures. All 5-mm unthreaded screws and 5.5-mm low threaded stainless steel screws can be used safely in full weight-bearing conditions of unreliable patients

🔖

Cite This Article

KARAARSLAN, A. A., KARAKAŞLI, A., KARCI, T., AYCAN, H., SESLİ, E. (2015). Reliability of threaded locking screws. Acta Orthopaedica et Traumatologica Turcica, Cilt: 49 - Sayı: 5, 552–557. https://ajindex.com/articles/353973

References (12)

  1. Colchero F, Orst G, Reboul C, Villalobos F, Vidal J. In- tramedullary locking nailing. Experimental study. Surgi- cal technic. Results. [Article in French] Rev Chir Orthop Reparatrice Appar Mot 1983;69:547-55. [Abstract] Kempf I, Grosse A, Taglang G, Favreul E. Gamma nail in the treatment of closed trochanteric fractures. Results and indications of 121 cases. Orthop Traumatol Surg Res 2014;100:75-83.
  2. Whittle AP, Wester W, Russell TA. Fatigue failure in small diameter tibial nails. Clin Orthop Relat Res 1995;315:119-28. Boenisch UW, de Boer PG, Journeaux SF. Unreamed in- tramedullary tibial nailing--fatigue of locking bolts. Injury 1996;27:265-70.
  3. Whittle AP, Russell TA, Taylor JC, Lavelle DG. Treat- ment of open fractures of the tibial shaft with the use of interlocking nailing without reaming. J Bone Joint Surg Am 1992;74:1162-71.
  4. Court-Brown CM, Will E, Christie J, McQueen MM. Reamed or unreamed nailing for closed tibial fractures. A prospective study in Tscherne C1 fractures. J Bone Joint Surg Br 1996;78:580-3.
  5. Chao CK, Hsu CC, Wang JL, Lin J. Increasing bending strength of tibial locking screws: mechanical tests and finite element analyses. Clin Biomech (Bristol, Avon) 2007;22:59-66.
  6. Hou SM, Wang JL, Lin J. Mechanical strength, fatigue life, and failure analysis of two prototypes and five conventional tibial locking screws. J Orthop Trauma 2002;16:701-8.
  7. Taylor SJ, Walker PS, Perry JS, Cannon SR, Woledge R. The forces in the distal femur and the knee during walking and other activities measured by telemetry. J Arthroplasty 1998;13:428-37.
  8. Taylor SJ, Walker PS. Forces and moments telemetered from two distal femoral replacements during various activi- ties. J Biomech 2001;34:839-48.
  9. Noble PC, Alexander JW, Lindahl LJ, Yew DT, Granberry WM, Tullos HS. The anatomic basis of femoral compo- nent design. Clin Orthop Relat Res 1988;235:148-65.
  10. Rubin PJ, Leyvraz PF, Aubaniac JM, Argenson JN, Estève P, de Roguin B. The morphology of the proximal femur. A three-dimensional radiographic analysis. J Bone Joint Surg Br 1992;74:28-32.
  11. Umer M, Sepah YJ, Khan A, Wazir A, Ahmed M, Jawad MU. Morphology of the proximal femur in a Pakistani population. J Orthop Surg (Hong Kong) 2010;18:279- 81. Sen RK, Tripathy SK, Kumar R, Kumar A, Dhatt S, Dhil- lon MS, et al. Proximal femoral medullary canal diameters in Indians: correlation between anatomic, radiographic, and computed tomographic measurements. J Orthop Surg (Hong Kong) 2010;18:189-94.
  12. Lin J, Hou SM. Bending strength an