Journals / BEYOGLU EYE JOURNAL / 2020 / Cilt: 5 - Sayı: 3

Measurement of Intraocular Pressure with Applanation, Dynamic Contour, and Air-Puff Tonometers: A Comparative Study in Primary Open-Angle Glaucoma and Healthy Cases

Pages
178–187
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

This new-generation digital tonometer can be used safely in glaucoma diagnosis and follow-up.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objectives: This study aimed to investigate and compare the reliability of Goldmann applanation tonometer (GAT), dynamic contour (DCT), and noncontact (NCT) tonometers in intraocular pressure (IOP) measurement and the affecting parameters in healthy subjects and cases with primary open-angle glaucoma (POAG).Methods: Left eyes of 64 cases (32 males and 32 females) were selected for this prospective, controlled study.
📊
Key Findings & Data:Of these cases, 33 had POAG, and 31 were selected as control.
💡
Conclusion & Impact:This new-generation digital tonometer can be used safely in glaucoma diagnosis and follow-up.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

Objectives: This study aimed to investigate and compare the reliability of Goldmann applanation tonometer (GAT), dynamic contour (DCT), and noncontact (NCT) tonometers in intraocular pressure (IOP) measurement and the affecting parameters in healthy subjects and cases with primary open-angle glaucoma (POAG).Methods: Left eyes of 64 cases (32 males and 32 females) were selected for this prospective, controlled study. Of these cases, 33 had POAG, and 31 were selected as control. IOP measurement was performed using NCT, DCT, and GAT consecutively for each patient, and then central corneal thickness (CCT) was measured. Ocular pulse amplitude (OPA) and all values were recorded.Results: The mean age was 53.36±10 years (31–80 years), and CCT was 561±45 μ. IOP was found as 16.39±3.75 mmHg with GAT, 17.89±3.55 mmHg with DCT, and 15.76±3.49 mmHg with NCT. A significant difference was found between DCT with NCT and GAT. Whereas, a positive correlation was found between CCT with all the three methods used, with DCT as the weakest. While the correlation between all the three methods was excellent, the strongest was found to be between DCT and GAT. Thick corneas affected all the three methods, but DCT was the least affected. While DCT tends to measure higher than both GAT and NCT, this difference decreased as the corneal thickness increased. OPA was found to be 2.56±1.04 mmHg; no statistical difference was found between the groups. A correlation was found between OPA and IOP, and OPA was found to be significantly higher in women.Conclusion: DCT is minimally affected by corneal factors, especially in thin corneas, and shows excellent correlation with GAT. This new-generation digital tonometer can be used safely in glaucoma diagnosis and follow-up.

🔖

Cite This Article

ÖZBİLEN, K. T., KOCABORA, M. S. (2020). Measurement of Intraocular Pressure with Applanation, Dynamic Contour, and Air-Puff Tonometers: A Comparative Study in Primary Open-Angle Glaucoma and Healthy Cases. BEYOGLU EYE JOURNAL, Cilt: 5 - Sayı: 3, 178–187. https://ajindex.com/articles/353104

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (16)

  1. Ozcetin H. Göz tansiyonu, glokom tanısı ve tipleri. 2nd ed. Göz tansiyonu ve glokomlar. İstanbul: Nobel Tıp Kitabevi; 2009. p. 343
  2. Flammer J, Orgül S, Costa VP, Orzalesi N, Krieglstein GK, Serra LM, et al. The impact of ocular blood flow in glaucoma. Prog Retin Eye Res 2002;21:359–93
  3. Watson PG, Jakeman C, Ozturk M, Barnett MF, Barnett F, Khaw KT. The complications of trabeculectomy (a 20-year follow-up). Eye (Lond) 1990;4:425–38
  4. Yanoff M, Duker J S. Ophtalmology. 2nd ed Türkçe baskı. İstanbul: Hayat Tıp Kitabevi; 2007. p. 1535–610
  5. Kanski J. Kanski's Clinical Ophtalmology. 8th ed. Philadelpiha: Elsevier; 2016. P.306–58
  6. Bengisu Ü: Göz Hastalıkları. 4th ed. Glokom. Ankara: Palme Yayınevi; 1998. p:139–58 7. Budenz DL, Anderson DR, Feuer WJ, Beiser JA, Schiffman J, Parrish RK 2nd, et al; Ocular Hypertension Treatment Study Group. Detection and prognostic significance of optic disc hemorrhages during the Ocular Hypertension Treatment Study. Ophthalmology 2006;113:2137–43
  7. Hoskins Jr HD, Kass M. Primary open angle glaucoma, Becker- Shaffers Diagnosis and Therapy of the Glaucomas. 6th Ed. Maryland Heights: Mosby 1989. p. 277–301
  8. Kaufman PL, Mittag TW. Epidemiology of open-angle glaucoma. In: Podos SM, Yanoff M, editors. Textbook of Ophthalmology. 7th ed. London: Mosby; 1994. p. 8.29–8.33
  9. Schottenstein EM. Intraocular pressure and tonometry. In: Ritch R, Shields MB, Krupin T, editors. The Glaucomas – Basic Sciences. 2nd ed. St. Louis: Mosby; 1996. p. 407–28
  10. Krupin T. Methods of measuring intraocular pressure. Manual of Glaucoma. New York: Churchill Livingstone; 1988. p. 7–18
  11. Ozcetin H. Klinik göz hastalıkları; göziçi basıncı ve glokomlar: B1. 6th Ed. İstanbul: Nobel Tıp kitabevi; 2003. p. 137–82
  12. Ozcetin H. Göziçi basıncı ölçümü ve tonometreler. Göz tansiyonu ve glokomlar. 2nd ed. İstanbul: Nobel Tıp Kitabevi; 2009. p. 55–99
  13. Kniestedt C, Nee M, Stamper RL. Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states. Graefes Arch Clin Exp Ophthalmol 2005;243:359–66
  14. Kanngiesser HE, Nee M, Kniestedt C, Inversini C, Stamper RL. The theoretical foundations of dynamic contour tonometry. Poster presented at: The Annual Meeting of the Association for Research in Vision and Ophthalmology. May 6, 2003; Fort Lauderdale, FL
  15. SMT Swiss Microtechnology AG. Pascal Dynamic Contour To nometer Operating Manuel, version 1.4 2004:30–2
  16. Kanngiesser HE, Kn