Can We Use Anterior Segment Parameters of an Optical Biometer and a Combined Topography System Interchangeably?

Pages
38–44
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Studies with larger sample sizes are warranted to evaluate whether the magnitude of these differences is clinically important or to see whether two devices can be used interchangeably for these parameters.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Purpose: To compare and analyze the agreement between the anterior segment parameters and pupillometric measurements obtained by Aladdin optical biometer, and Sirius corneal topographer.
📊
Key Findings & Data:Subjects with a history of ocular surgery or disease, refractive errors greater than 3 D and using ophthalmic medication or contact lenses were excluded.
💡
Conclusion & Impact:Studies with larger sample sizes are warranted to evaluate whether the magnitude of these differences is clinically important or to see whether two devices can be used interchangeably for these parameters.
🔬Design:Review & Systematic Survey
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Özet

Purpose: To compare and analyze the agreement between the anterior segment parameters and pupillometric measurements obtained by Aladdin optical biometer, and Sirius corneal topographer. Material and Methods: Medical records of healthy patients applying for refractive surgery were reviewed. Anterior chamber depth (ACD), whiteto-white distance (WTW) and keratometry (K), as well as pupillometric measurements, were compared. Subjects with a history of ocular surgery or disease, refractive errors greater than 3 D and using ophthalmic medication or contact lenses were excluded. Results: Eighty eyes of eighty patients were included. Mean ACD and WTW measured by Aladdin were 3.65±0.25mm and 12.15±0.24mm and Sirius were 3.73±0.22mm and 12.48±0.32mm, respectively. The difference in ACD and WTW was statistically signifi cant between two devices (p=0.026, p=0.001 respectively). Mean K1 and K2 were 42.53±0.90 D and 43.98±0.86D with Aladdin, whereas they were 42.49±0.85D and 44.20±0.97D by Sirius, respectively. The difference was statistically nonsignifi cant (p=0.776, p=0.310 respectively). Mean photopic and mesopic pupil diameters were 3.21±0.33mm and 4.99±0.44mm with Aladdin while they were 3.65±0.48mm and 4.43±0.63mm with Sirius, which was different between two devices (p<0.01,p<0.01 respectively). Conclusion: Keratometric measurements obtained by Aladdin seemed to be in agreement statistically with Sirius topographer, however, poor agreement was observed for mean ACD, WTW, and pupillometric measurements. Studies with larger sample sizes are warranted to evaluate whether the magnitude of these differences is clinically important or to see whether two devices can be used interchangeably for these parameters.

🔖

Cite This Article

CERAN, B. B., KARATEPE, A. S. H., TAŞINDI, E. (2020). Can We Use Anterior Segment Parameters of an Optical Biometer and a Combined Topography System Interchangeably?. Glokom Katarakt, Cilt: 15 - Sayı: 1, 38–44. https://ajindex.com/articles/226807

References (14)

  1. Olsen T. Calculation of intraocular lens power: a review. Acta Ophthalmol Scand. 2007;85:472-85
  2. Lee AC, Qazi MA, Pepose JS. Biometry and intraocular lens power calculation. Curr Opin Ophthalmol. 2008;19:13-7
  3. Haigis W. IOL Power Calculations. Ophthalmology 2010;117:400-1
  4. Gokce SE, De Oca IM, Cooke DL, et al. Accuracy of intraocular lens calculation formulas in relation to anterior chamber depth in patients with normal axial lengths. J Cataract Refract Surg, 2018; 44, 362-8
  5. Ghoreishi M, Abdi-Shahshahani M, Peyman A, et al. A model for predicting sulcus-to-sulcus diameter in the posterior chamber phakic intraocular lens candidates: correlation between ocular biometric parameters. Int J Ophthalmol. 2018; 39, 661-6
  6. Salah-Mabed I, Saad A, Gatinel D. Assessing Repeatability of Pupillometric Measurements in the Eyes of Refractive Surgery Candidates Using Infrared Pupillometer. J Refract Surg. 2017;33:552-7
  7. Rozema JJ, Wouters K, Mathysen DG, et al. Overview of the repeatability, reproducibility, and agreement of the biometry values provided by various ophthalmic devices. Am J Ophthalmol. 2014;158:1111-20
  8. Domínguez-Vicent A, Monsálvez-Romín D, Águila-Carrasco AJD, et al. Measurements of anterior chamber depth, whiteto-white distance, anterior chamber angle, and pupil diameter using two Scheimpfl ug imaging devices. Arq Bras Oftalmol. 2014;77:233-7
  9. Nemeth G, Szalai E, Hassan Z, et al. Corneal biomechanical data and biometric parameters measured with Scheimpfl ugbased devices on normal corneas. Int J Ophthalmol. 2017;10:217
  10. Domínguez-Vicent A, Pérez-Vives C, Ferrer-Blasco T, et al. Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review. Int J Ophthalmol. 2016;9:1057
  11. Aramberri J, Araiz L, Garcia A, et al. Dual versus single Scheimpfl ug camera for anterior segment analysis: precision and agreement. J Cataract Refract Surg. 2012;38:1934-49
  12. Polat O, Baysal Z, Ozcan S, et al. Comparison of anterior segment measurements obtained by Aladdin optical biometer and Sirius corneal topography. Turk J Ophthalmol. 2016;46:259
  13. Kwon J, Sung KR, Han S, et al. Subclassifi cation of primary angle closure using anterior segment optical coherence tomography and ultrasound biomicroscopic parameters. Ophthalmology. 2017;124:1039-47
  14. Liu L, Liu X, Huang C, et al. Associated factors of acute primary angle closure glaucoma in a sub-group of Chinese people: comparison between at