The Effects of Collagen Cross-Linking on Intraocular Pressure and Central Corneal Thickness

Kollajen Çapraz Bağlamanın Göz İçi Basıncı ve Merkezi Kornea Kalınlığı Üzerine Etkisi

Pages
100–102
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

This may be due to the increased stiffness of the treated cornea

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Purpose: To investigate the effect of corneal collagen cross-linking (CCL) on central corneal thickness (CCT) and intraocular pressure (IOP).
📊
Key Findings & Data:Materials and Methods: CCL with 0.1% riboflavin solution and 30 min of ultraviolet A radiation was performed on the right eyes of 20 New Zealand albino adult rabbits (2.4-3.2 kg), (20 eyes).
💡
Conclusion & Impact:This may be due to the increased stiffness of the treated cornea
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Purpose: To investigate the effect of corneal collagen cross-linking (CCL) on central corneal thickness (CCT) and intraocular pressure (IOP). Materials and Methods: CCL with 0.1% riboflavin solution and 30 min of ultraviolet A radiation was performed on the right eyes of 20 New Zealand albino adult rabbits (2.4-3.2 kg), (20 eyes). The left eyes served as control. IOP was measured using Tonopen hand-held device (corneal applanation tonometer) and CCT by the pachymeter (Accutome, AccuPachV pachymeter, USA) before treatment and weekly for one month following CCL. Results: Before CCL, IOP readings and CCT measurements were similar between the two eyes (p>0.05). Following CCL treatment, the mean IOP was found significantly higher in the treatment group, however, mean CCT was not significantly different between the groups. (0.005<p<0.03). Conclusions: Measuring IOP by tonopen yielded overestimated IOP results following CCL. This may be due to the increased stiffness of the treated cornea

Özet

Amaç: Kollajen çapraz bağlamanın (KÇB) göz içi basıncı(GİB) ve merkezi kornea kalınlığı (MKK) üzerine etkisinin araştırılması. Gereç ve Yöntem: Yeni Zelanda cinsi toplam 20 albino tavşanın (2.4-3.2 kg) sağ gözlerine (20 göz) %0.1 riboflavin solüsyon ve 30 dakika ultraviyole A ışını uygulanarak KÇB işlemi yapıldı. Sol gözler kontrol grubu olarak alındı. GİB değerleri Tonopen XL (Medtronic, Jacksonville, Florida, USA) kulanılarak, MKK pakimetri (Accutome, AccuPachV pachymeter, USA) kullanılarak tedavi öncesi ve tedavi sonrası haftalık olmak üzere 1 ay boyunca ölçüldü. Bulgular: KÇB öncesi, IOP ve MKK ölçümleri iki göz arasında benzer saptandı (p>0.05). KÇB’yi takiben, ortalama GİB ölçümleri tedavi edilen gözlerde anlamlı oranda yüksek saptanırken, MKK ölçümleri gruplar arasında anlamlı farklılık göstermedi (p>0.05). Sonuç: KÇB sonrası tonopen ile ölçülen GİB değerleri normalden yüksek bulunmuştur. Tedavi edilen korneada artmış rijiditenin buna sebep olabileceği düşünülmektedir.

🔖

Cite This Article

ZENGİN, M. Ö., ÇINAR, E., ÇELİK, A., KÜÇÜKERDÖNMEZ, C. (2015). The Effects of Collagen Cross-Linking on Intraocular Pressure and Central Corneal Thickness. Glokom Katarakt, Cilt: 10 - Sayı: 2, 100–102. https://ajindex.com/articles/201619

References (14)

  1. Ku JY, Niederer RL, Patel DV, et al. Laser scanning in vivo confocal analysis of keratocyte density in keratoconus. Ophthalmology 2008;115:845-50
  2. Rabinowitz YS. Keratoconus. Surv Ophthalmol 1998; 42:297-319
  3. Dahl BJ, Spotts E, Truong JQ. Corneal collagen crosslinking: An introduction and literature review. Optometry 2012;83:33-42
  4. Kling S, Ginis H, Marcos S. Corneal biomechanical properties from twodimensional corneal flap extensiometry: application to uv-riboflavin cross-linking. Invest Ophthalmol Vis Sci 2012;53:5010-5
  5. Goldich Y, Barkana Y, Morad Y, et al. Can we measure corneal biomechanical changes after collagen cross-linking in eyes with keratoconus? pilot study. Cornea 2009;28:498-502
  6. Kymionis GD, Grentzelos MA, Kounis GA, et al. Intraocular pressure measurements after corneal collagen crosslinking with riboflavin and ultraviolet A in eyes with keratoconus. J Cataract Refract Surg 2010;36:1724-7
  7. Spoerl E, Huhle M, Seiler T. Induction of crosslinks in corneal tissue. Exp Eye Res 1998;66:97-103
  8. Spoerl E, Seiler T. Techniques for stiffening the cornea. J Refract Surg 1999;15:711-3
  9. Wollensak G, Spoerl E, Seiler T. Stress-strain measurements of human and porcine corneas after riboflavin-ultraviolet-A-induced crosslinking. J Cataract Refract Surg 2003;29:1780-5
  10. Kohlhaas M, Spoerl E, Schilde T, et al. Biomechanical evidence of the distribution of cross-links in corneas treated with riboflavin and ultraviolet A light. J Cataract Refract Surg 2006;32:279-83
  11. Seiler T, Hafezi F. Corneal cross-linking-induced stromal demarcation line. Cornea 2006;25:1057-9
  12. Lim JI, Blair NP, Higginbotham EJ, et al. Assessment of intraocular pressure in vitrectomized gas-containing eyes. A clinical and manometric comparison of the Tono-Pen to the pneumotonometer. Arch Ophthalmol 1990;108:684-8
  13. Lim KS, Wickremasinghe SS, Cordeiro MF, et al. Accuracy of intraocular pressure measurements in New Zealand white rabbits. Invest Ophthalmol Vis Sci 2005;46:2419-23
  14. Abrams LS, Vitale S, Jampel HD. Comparison of three tonometers for measuring intraocular pressure in rabbits. Invest Ophthalmol Vis Sci 1996;37:940-4.