Journals / BEYOGLU EYE JOURNAL / 2017 / Cilt: 2 - Sayı: 1

Comparison of Goldmann Applanation Tonometry and Icare Rebound Tonometry in Eyes of Varied Corneal Thickness

Pages
9–12
DOI
—
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

There was no correlation between CCT and difference in IOP obtained by RBT and GAT.Conclusion: Icare RBT demonstrated agreement with Goldmann applanation tonometry, except in thin corneas.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objectives: The aim of the present study was to compare intraocular pressure (IOP) level as measured with Goldmann applanation tonometry (GAT) and Icare rebound tonometry (RBT; Icare Finland Oy, Vantaa, Finland) in patients with varied central corneal thickness (CCT).Methods: Total of 138 eyes of 138 consecutive patients from glaucoma department were studied.
📊
Key Findings & Data:Ultrasound pachymeter was used to measure CCT in all cases and patients were divided into 3 groups according to CCT.
💡
Conclusion & Impact:There was no correlation between CCT and difference in IOP obtained by RBT and GAT.Conclusion: Icare RBT demonstrated agreement with Goldmann applanation tonometry, except in thin corneas.
🔬Design:Research Article
📍Scope:Comprehensive Sample
⚙️Variables:Domain Parameters

Abstract

Objectives: The aim of the present study was to compare intraocular pressure (IOP) level as measured with Goldmann applanation tonometry (GAT) and Icare rebound tonometry (RBT; Icare Finland Oy, Vantaa, Finland) in patients with varied central corneal thickness (CCT).Methods: Total of 138 eyes of 138 consecutive patients from glaucoma department were studied. Ultrasound pachymeter was used to measure CCT in all cases and patients were divided into 3 groups according to CCT. Group 1 comprised patients with thin CCT (<520 ?), Group 2 was made up of patients with normal CCT (520-570 ?), and patients with thick CCT (>570 ?) composed Group 3. IOP was measured with RBT, which uses average of 6 readings, followed by GAT. Measurements of groups were compared.Results: Mean age of patients was 59.4±10.9 years. Mean CCT value was 494.9±15.9 ? in Group 1 (n=40), 549.1±16.3 ? in Group 2 (n=52), and 604.0±22.2 ? in Group 3 (n=46). Mean IOP measurement obtained by GAT and RBT was 13.2±4.0 and 13.8±3.9 mmHg, respectively, in Group 1 (p=0.013), 16.9±3.2 and 16.8±3.1 mmHg in Group 2 (p=0.745), and 18.1±3.4 and 18.4±3.8 mmHg in Group 3 (p=0.359). Mean difference in IOP for entire study group between GAT and RBT was -0.21±1.7 mmHg (p=0.15). This difference was statistically significant only in Group 1 (p=0.013). Positive correlation was found between CCT and measurement via both GAT and RBT (p<0.01). There was no correlation between CCT and difference in IOP obtained by RBT and GAT.Conclusion: Icare RBT demonstrated agreement with Goldmann applanation tonometry, except in thin corneas.

🔖

Cite This Article

PAŞAOĞLU, I., ALTAN, Ç., ÖLÇÜCÜ, O., ALAGÖZ, N., ŞATANA, B., BAŞARIR, B., ONMEZ, F., TAŞKAPILI, M. (2017). Comparison of Goldmann Applanation Tonometry and Icare Rebound Tonometry in Eyes of Varied Corneal Thickness. BEYOGLU EYE JOURNAL, Cilt: 2 - Sayı: 1, 9–12. https://ajindex.com/articles/147834

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (13)

  1. Whitacre MM, Stein RA, Hassanein K. The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 1993;115:592-6
  2. Whitacre MM, Stein R. Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol 1993;38:1-30
  3. Pakrou N, Gray T, Mills R, Landers J, Craig J. Clinical comparison of the Icare tonometer and Goldmann applanation tonometry. J Glaucoma 2008;17:43-7
  4. Poostchi A, Mitchell R, Nicholas S, Purdie G, Wells A. The iCare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness. Clin Exp Ophthalmol 2009;37:687-91
  5. Rao A, Kumar M, Prakash B, Varshney G. Relationship of central corneal thickness and intraocular pressure by iCare rebound tonometer. J Glaucoma 2014;23:380-4
  6. Martinez-de-la-Casa JM, Jimenez-Santos M, Saenz-Frances F, Matilla-Rodero M, Mendez-Hernandez C, Herrero-Vanrell R, Garcia-Feijoo J. Performance of the rebound, noncontact and Goldmann applanation tonometers in routine clinical practice. Acta Ophthalmol 2011;89:676-80
  7. Iliev ME, Goldblum D, Katsoulis K, Amstutz C, Frueh B. Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness. Br J Ophthalmol 2006;90:833-5
  8. Munkwitz S, Elkarmouty A, Hoffmann EM, Pfeiffer N, Thieme H. Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range. Graefes Arch Clin Exp Ophthalmol 2008;246:875-9
  9. Kontiola A, Puska P. Measuring intraocular pressure with the Pulsair 3000 and Rebound tonometers in elderly patients without an anesthetic. Graefes Arch Clin Exp Ophthalmol 2004;242:3-7. 10. van der Jagt LH, Jansonius NM. Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry. Ophthalmic Physiol Opt 2005;25:429-35
  10. Rehnman JB, Martin L. Comparison of rebound and applanation tonometry in the management of patients treated for glaucoma or ocular hypertension. Ophthalmic Physiol Opt 2008;28:382-6
  11. Salim S, Du H, Wan J. Comparison of intraocular pressure measurements and assessment of intraobserver and interobserver reproducibility with the portable ICare rebound tonometer and Goldmann applanation tonometer in glaucoma patients. J Glaucoma 2013;22:325-9
  12. Martinez-de-la-Casa JM, Garcia-Feijoo J, Castillo A, Garcia-San- chez J. Reproducibility and clinical evaluation of rebound tonometry. Invest Ophthalmol Vis Sci 2005;46:4578-80
  13. Fernan