While mass spectra were sometimes identical in electron impact mode, particularly for the ion m/z 58, using chemical ionization mode, characteristic ions were obtained.
📌 3 Key Takeaways
🎯
Objective & Hypothesis:Positive-ion electron impact and positive-ion chemical ionization mass spectra of 30 antihistamines H1 are presented.
📊
Key Findings & Data:Drugs were extracted under alkaline conditions, and separated on a 25 m BP5 capillary column.
💡
Conclusion & Impact:While mass spectra were sometimes identical in electron impact mode, particularly for the ion m/z 58, using chemical ionization mode, characteristic ions were obtained.
🔬Design:Araştırma Makalesi
📍Scope:Comprehensive Sample
⚙️Variables:Antihistamines H1 · Forensic analysis · Mass spectrometry
Abstract
Positive-ion electron impact and positive-ion chemical ionization mass spectra of 30 antihistamines H1 are presented. Drugs were extracted under alkaline conditions, and separated on a 25 m BP5 capillary column. GC retention times are reported on both a retention indices basis as well as relative to SKF 525 A internal standard. While mass spectra were sometimes identical in electron impact mode, particularly for the ion m/z 58, using chemical ionization mode, characteristic ions were obtained.
Kintz, P. (1991). Screening Procedure for 30 Antihistamines H1 Using Capillary Gas Chromatography - Mass Spectrometry. Adli Tıp Dergisi, Cilt: 7 Sayı: 3-4, 80–85. https://ajindex.com/articles/1352959