Journals / Dicle Tıp Dergisi / 2026 / Cilt: 47 Sayı: 2

A Comparison Of Manual Hemocytometry And Different Modes Of Mindray BC-6800 Hematology Analyzer For Cell Count In Peritoneal And Pleural Fluids

Ünsal Savcı(HİTİT ÜNİVERSİTESİ)
Mustafa Şahin(HİTİT ÜNİVERSİTESİ)
Barış Eser(HİTİT ÜNİVERSİTESİ)
Pages
278–285
Publication Date
2020-06-17
🏛️Bibliographic metadata and full text access for this article are provided via DergiPark.
DergiPark Record ↗
✨

AI Summary & Key Takeaways

AI Insight
⚡ TL;DR(Too Long; Didn't Read)

Results: There was a statistically significant difference between manual count and hemogram mode (for peritoneal fluids p 100 cells/mL.

📌 3 Key Takeaways

🎯
Objective & Hypothesis:Objective: Total white blood cell (WBC) count in body fluids (BF) is important in diagnosis and treatment of inflammatory and infectious diseases.
📊
Key Findings & Data:We aimed to compare hemogram and BF modes of Mindray BC-6800 hematology analyzer with the manual hemocytometry method for determination of total WBC count in peritoneal and pleural fluids.
💡
Conclusion & Impact:Results: There was a statistically significant difference between manual count and hemogram mode (for peritoneal fluids p 100 cells/mL.
🔬Design:Araştırma Makalesi
📍Scope:143 specimens
⚙️Variables:Automated hematology analyzer · cell count

Özet

Objective: Total white blood cell (WBC) count in body fluids (BF) is important in diagnosis and treatment of inflammatory and infectious diseases. We aimed to compare hemogram and BF modes of Mindray BC-6800 hematology analyzer with the manual hemocytometry method for determination of total WBC count in peritoneal and pleural fluids. Method: Our study consisted of a total of 143 specimens, 109 peritoneal fluid and 34 pleural effusion fluid. Each sample was analyzed consecutively twice, first with the manual method and, later with both hemogram mode and the BF mode of Mindray BC-6800 automated hematology analyzer. Results: There was a statistically significant difference between manual count and hemogram mode (for peritoneal fluids p 100 cells/mL.

🔖

Cite This Article

Savcı, Ü., Şahin, M., Eser, B. (2020). A Comparison Of Manual Hemocytometry And Different Modes Of Mindray BC-6800 Hematology Analyzer For Cell Count In Peritoneal And Pleural Fluids. Dicle Tıp Dergisi, Cilt: 47 Sayı: 2, 278–285. https://doi.org/10.5798/dicletip.755702

📄 Full Text PDF Viewer

📄

PDF görüntüleyici hazırlanıyor...

References (13)

  1. De Smet D, Van Moer G, Martens GA, et al. Use of the Cell-Dyn Sapphire hematology analyzer for automated counting of blood cells in body fluids. Am J Clin Pathol. 2010; 133: 291-9.
  2. Cognialli RCR, Comar SR, Souza AMD, Singer GMB. Evaluation of pleural and ascitic fluid analysis on the Sysmex XE-5000 hematology analyzer. J Bras Patol Med Lab. 2017; 53; 150-8.
  3. Fuster O, Andino B, Pardo A, Laiz B. Continuous ambulatory peritoneal dialysis, ascitic and pleural body fluids evaluation with the Mindray BC-6800 hematology analyzer. Journal of clinical laboratory analysis. J Clin Lab Anal. 2018; 32: e22240.
  4. Runyon BA. Management of adult patients with ascites due to cirrhosis: an update. Hepatology. 2009; 49: 2087-107.
  5. Conner BD, Lee YG, Branca P, Rogers JT, Rodriguez RM, Light RW. Variations in pleural fluid WBC count and differential counts with different sample containers and different methods. Chest. 2003; 123: 1181-7.
  6. Bottini PV, Garlipp CR, Nogueira BF, Pompeo DB, Souza MI (2018). Performance evaluation of a new generation of automated analyzer for pleural and peritoneal fluids cytology. Clin Lab. 2018; 64: 1783-6.
  7. Schonenberg H, Fassl H. Comparison of the coulter-counter-method and the counting-chamber to determine the cell count of the cerebrospinal fluid. Arztliche Forschung. 1970; 24: 118.
  8. Walker TJ, Nelson LD, Dunphy BW, Anderson DM, Kickler TS. Comparative Evaluation of the Iris iQ200 Body Fluid Module With Manual Hemacytometer Count. Am J Clin Pathol. 2009; 131: 333-8.
  9. Talstad I. Electronic counting of spinal fluid cells. Am J Clin Pathol. 1984; 81: 506-11.
  10. Aune MW, Becker JL, Brugnara C, et al. Automated flow cytometric analysis of blood cells in cerebrospinal fluid. Am J Clin Pathol. 2004; 121: 690-700.
  11. Takemura H, Ai T, Kimura K, et al. Evaluation of cell count and classification capabilities in body fluids using a fully automated Sysmex XN equipped with high-sensitive Analysis (hsA) mode and DI-60 hematology analyzer system. PloS one. 2018; 13: e0195923.
  12. Li PKT, Szeto CC, Piraino B, et al. Peritoneal dialysis-related infections recommendations: 2010 update. Perit Dial Int. 2010; 30: 393-423.
  13. Fleming C, Russcher H, Lindemans J, de Jonge R. Clinical relevance and contemporary methods for counting blood cells in body fluids suspected of inflammatory disease. Clin Chem Lab Med. 2015; 53: 1689-706.