Dosimetric Accuracy of an HDR Brachytherapy Treatment Planning System for Different Irradiation Lengths with Monte Carlo Simulation
- Sayfa
- 36–44
- DOI
- —
Abstract
OBJECTIVEThe purpose of the present study was to verify calculated dose rate profiles of Ir-192 for four differentirradiation lengths and various numbers of dwell positions by using treatment planning system (TPS)calculations and Monte Carlo (MC) simulations.METHODSDose rate profiles per air-kerma strength in μGy m2/h (abbreviated as U) were calculated on transverseaxis for irradiation lengths of 2, 3, 4, and 5 cm by Nucletron Oncentra TPS. The same irradiation scenarios were simulated with Monte Carlo N-Particle (MCNP) Transport Code. The MC and TPS calculateddose rate profiles were compared for all settings.RESULTSThe difference between the calculated dose rate profiles by MC and TPS was found within 2% at aninterval of 2 mm–8 cm away from the source for all irradiation lengths. The largest discrepancy of 5.2%was computed at a distance of 10 cm from the source for 5 cm irradiation length.CONCLUSIONThere is a good consistency between the calculated dose rate profiles by MCNP and Oncentra TPS. Thedeviation between the calculated dose rate values slightly increases as the distance from the source center increases >5 cm from the source, and its quantity depends on the number of dwell positions.