Dergiler / Turkish Journal of Chemistry / 2017 / Cilt: 41 - Sayı: 6

Cyclic voltammetry behavior of tamoxifen in different electrolytes, electrochemical characterization, and its measurement by differential pulse anodic voltammetry

Sayfa
826–844
DOI
—

Abstract

Cyclic voltammetry (CV) was performed on tamoxifen (tam) using different electrodes and in various acidicelectrolytes containing 10% v v1methanol (MeOH). Tam oxidation was found to be most favorable using 0.1 molL1H2SO4. To investigate the mechanism, we performed chronoamperometry, CV at different scan rates, differentialpulse anodic voltammetry (DPAV), and electrochemical impedance spectroscopy on tam-covered glassy carbon electrodes(GCEs). The electrode area, diffusion coefficient, and surface concentration () of tam were calculated to be 0.062 cm2,3.65 106cm2s1, and 3.2 1010mol cm2respectively. Then the effects of different parameters on theDPAV were optimized. The best conditions were 2.5% v v1cMeOH, 0.1 mol L1H2SO4, deposition potential 0.4 V,deposition time 30 s, and GCE rotating rate 400 rpm. Therefore, the calibration curve was plotted in the range of 0.5to 5 g mL1. The limits of detection and quantitation (LOD and LOQ) were found to be 0.008 and 0.025 g mL1,respectively. The relative bias and standard deviation of 2 g mL1tam were estimated to be 5% (n = 5) and 2.3%,respectively. Finally, the proposed method was successfully employed for the determination of tam in real samples.