Dergiler / Turkish Journal of Chemistry / 2019 / Cilt: 43 - Sayı: 6
An ultrasensitive electrochemical immunosensor platform based on disposable ITO electrode modified by 3-CPTMS for early detection of parathyroid hormone
- Sayfa
- 1697–1710
- DOI
- —
Abstract
Parathyroid hormone (PTH) is a polypeptide containing 84 amino acids secreted by cells of the parathyroidglands. Imbalances of parathyroid levels cause medical problems such as osteoporosis, mental disorders, pancreatitis,kidney stones, cancer, and other symptoms. In this study, we aimed to design an ultrasensitive electrochemicalimmunosensor for PTH detection. Indium tin oxide (ITO) was used as an electrode for electrochemical impedancespectroscopy (EIS) measurements. ITO sheets were modified by using 3-cyanopropyltrimethoxysilane (3-CPTMS) selfassembled monolayers (SAMs) for immobilizing the anti-PTH antibody via covalent interactions. Cyclic voltammetry (CV) and EIS methods were applied to characterize immobilization steps. Therefore, 1% was selected for an optimal concentration of silane, 10 ng/mL was selected as an optimal concentration of anti-PTH, and 30 and 45 min were selected as optimal incubation times for anti-PTH and PTH, respectively. PTH antigen was determined in the concentration range from 0.05 fg/mL to 150 fg/mL. To detect the analytical characterization of the 3-CPTMS modified immunosensor, linear range, repeatability, reproducibility, Kramers–Kronig transform, and regeneration studies were performed. Also, the shelf life of the developed biosensor was investigated. Finally, real human serum samples were analyzed with the PTH immunosensor. The results showed that the designed biosensor system has high potential for early detection for medical treatments.