Dergiler / European Endodontic Journal / 2019 / Cilt: 4 - Sayı: 1

Accuracy of Software-Based Three-Dimensional Root Canal Length Measurements Using Cone-Beam Computed Tomography

Sayfa
28–32
DOI
—

Abstract

Objective: This study aims to evaluate the accuracy of three-dimensional root canal length measurements performed by dentists with different experience levels using a special software based on cone beam com-puted tomography (CBCT).Methods: A CBCT scan of an artificial resin maxillary molar was used to train dentists (n=65) in using the software (3D Endo, DentsplySirona, Ballaigues, Switzerland) as part of a continuing education course. At the beginning, each participant completed a questionnaire on endodontic and CBCT experiences. After com-prehensive instructions, each participant performed an entire three-dimensional treatment plan by tracing the root canal anatomy between the apical foramen and the center of the canal orifice and simulating a straight-line access. The final root canal length was indicated after the virtual rubber stop of a simulated instrument was adapted to the adjoining cusp. To evaluate the individual accuracy in terms of trueness and precision, differences between the three-dimensional planning and the actual root canal length (ARCL) were calculated, and statistically analyzed.Results: Mean absolute differences between the measurements with the 3D Endo™ software (n=260) and the ARCL were 0.30±0.22 mm. All measurements were within a limit of ±1 mm. The accuracy of root canal length measurements was significantly influenced by the type of root canal (p