Dergiler / Diagnostic and Interventional Radiology / 2019 / Cilt: 25 - Sayı: 6
Assessment of the compliance with minimum acceptable technical parameters proposed by PI-RADS v2 guidelines in multiparametric prostate MRI acquisition in tertiary referral hospitals in the Republic of Turkey
- Sayfa
- 421–427
- DOI
- —
Özet
PURPOSEAlthough the clinical use of multiparametric prostate magnetic resonance imaging (mpMRI) isincreasing, the adherence to parameters for mpMRI, which had been described in the ProstateImaging-Reporting and Data System version 2 (PI-RADS v2) for an optimum image acquisition isunknown. In this paper, we aimed to determine the compliance with the minimum acceptabletechnical parameters for prostate mpMRI defined by PI-RADS v2 in tertiary care centers in Turkey.METHODSWe sent a survey to all radiology departments of tertiary referral hospitals in Turkey (n=120)to evaluate their adherence to PI-RADS v2 technical specifications. Statistical analysis was performed using chi-square, Fisher exact, ANOVA, and the Student t tests. The cutoff values for image acquisition times were also determined with receiver operating characteristics (ROC) analysis. P values <0.05 were considered statistically significant.RESULTSOne hundred and eleven clinics responded to our survey (response rate, 92.5%). Prostate MRIwas reported to be performed in 61 centers, of which 26 (42.6%) used 3 T (Tesla) scanner while35 (57.4%) used 1.5 T. The adherence to slice thickness, in-plane phase and frequency resolutionson T2-weighted imaging were 68.9%, 41%, and 9.8%, respectively. The adherence to the sameparameters on diffusion-weighted imaging (DWI) were higher compared with T2-weighted imaging (85.2%, 62.3%, and 78.7%, respectively). In comparative analysis, the adherence to slicethickness, field of view (FOV) and in-plane phase resolution on T2-weighted imaging were higherfor 3 T compared with 1.5 T scanners (P = 0.004, P = 0.041, and P = 0.001, respectively). T2-weighted imaging acquisition time was significantly longer for the centers that adhered to FOV (P =0.034) and in-plane T2-weighted imaging phase resolution (P = 0.028). The DWI scan time wassignificantly longer when they adhered to DWI-FOV (P = 0.014) and b value ≥1400 s/mm2 (P =0.008). The calculated cutoff of scan times were 220 s in T2-weighted imaging and 312 s in DWIto ensure the compliance with voxel sizes and b value criteria.CONCLUSIONThe tertiary referral centers in Turkey did not meet majority of the technical specifications of PI-RADSv2 during prostate MRI acquisition. Awareness to the minimum acceptable technical parameters ofmpMRI should be increased to potentially improve the quality of prostate cancer imaging.