Dergiler / Diagnostic and Interventional Radiology / 2016 / Cilt: 22 - Sayı: 5
Günther Tulip inferior vena cava filter retrieval using a bidirectional loop-snare technique
- Sayfa
- 460–462
- DOI
- —
Abstract
Many advanced techniques have been reported in the literature for difficult Günther Tulip filter removal. This report describes a bidirectional loop-snare technique in the setting of a fibrin scar formation around the filter leg anchors. The bidirectional loop-snare technique allows for maximal axial tension and alignment for stripping fibrin scar from the filter legs, a commonly encountered complication of prolonged dwell times.