Using Response Surface Methodology for Amperometric Glucose Biosensor Construction

Sayfa
1–15
Yayın Tarihi
2018-01-04
DOI
—
🏛️Bu makalenin bibliyografik verileri ve tam metin bağlantısı DergiPark üzerinden sağlanmıştır.
DergiPark Kaydı ↗
✨

Yapay Zeka Hızlı Bakış & Çıkarımlar

AI Insight
⚡ TL;DR(1 Cümlede Hızlı Çıkarım)

The highest current response was obsreved when pH of glucose solution and applied potential were 6.0 and 0.8, respectively.

📌 3 Maddede Bu Makale

🎯
Amaç & Hipotez:In this study, construction of amperometric glucose biosensor was carried out by immobilizing of glucose oxidase (GOD) on platinum electrode with 0.09 cm2 surface area which coated with polypyrrole (PPy) by cyclic voltammetry technique.
📊
Temel Bulgular & Veri:To this, State Ease Design Expert 8.0.7.1. (Serial Number:0021-6578) programe was used.
💡
Sonuç & Katkı:The highest current response was obsreved when pH of glucose solution and applied potential were 6.0 and 0.8, respectively.
🔬Tür:Araştırma Makalesi
📍Kapsam:Genişletilmiş Akademik Örneklem
⚙️Değişkenler:Response surface methodology · Pt electrode · Polypyrrole

Abstract

In this study, construction of amperometric glucose biosensor was carried out by immobilizing of glucose oxidase (GOD) on platinum electrode with 0.09 cm2 surface area which coated with polypyrrole (PPy) by cyclic voltammetry technique. Because measured current values in the presence of glucose would be affected from the electrode preparing and working conditions, experimental parameters should be optimized by response surface methodology (RSM). To this, State Ease Design Expert 8.0.7.1. (Serial Number:0021-6578) programe was used. PPy synthesis conditions of pyrrole (Py) monomer concentration and scan rate were optimized according to current response in presence of glucose. Optimal Py monomer concentration and scan rate for PPy synthesis were determined as 10 mM and 50 mV/s, respectively. Immobilization parameters such as concentrations of chitosan, GOD and glutaraldehyde (GAL) also were optimized by RSM as 1.0 %, 4 mg/ml and 0.0625 %, respectively. The digital photos of electrodes at each stage were obtained. All electrodes well characterized in absence and in the presence of glucose by cyclic voltammetry and impedance techniques and it was observed that electrodes were sensitive to glucose molecule. Finally the effect of working pH and applied potential on the current response was investigated by RSM. The highest current response was obsreved when pH of glucose solution and applied potential were 6.0 and 0.8, respectively.

🔖

Bu Makaleye Atıf Yap

Özyılmaz, G., Özyılmaz, A. T., Ağçam, S. (2018). Using Response Surface Methodology for Amperometric Glucose Biosensor Construction. Natural and Engineering Sciences, Cilt: 3 - Sayı: 1, 1–15. https://ajindex.com/articles/793240

📄 Tam Metin PDF Okuyucu

📄

PDF görüntüleyici hazırlanıyor...

Kaynakça (11)

  1. Ahuja, T., Mir, I.A., Kumar, D. & Rajesh, (2007). Biomolecular immobilization on conducting polymers for biosensing applications. Biomaterials, 28, 791-805.
  2. Bal, Ö. (2012).Preparation of a new biosensor for determination of glucose with immobilization on the polypyrrole film of glucose oxidase in biologically fluids. MSc. Thesis, Gazi University. Bezerra, M.A., Santelli, R.E., Oliveira, E.P., Villar L.S. & Escaleira, L.A. (2013). Highly selective electrochemical biosensor for the determination of folic acid based on DNA modified-pencil graphite electrode using response surface methodology, Materials Science and Engineering C, 33 1753–1758.
  3. Borole, D. D., Kapadi, U.R., Mahulikar, P.P. & Hundiwale, D.G. (2004). Glucose Oxidase Electrodes of Polyaniline, Poly(o-toluidine) and Their Copolymer as Biosensor: a Comparative Study. Polymer Advantage Technology, 15, 306-312.
  4. Chaichi, M.J. & Ehsani, M. (2016). A novel glucose sensor based on immobilization of glucose oxidase on the chitosan-coated Fe3O4 nanoparticles and the luminol-H2O2-gold nanoparticle chemiluminescence detection system, Sensors and Actuators B-Chemical, 223, 713-722.
  5. Chen, C., Jian, Y. & Kan, J. (2005). A noninterference polypyrole glucose biosensor. Biosensors Bioelectronics. 22, 639-643.
  6. Dong Y.P., Huang L., Chu X.F. & Pei L.Z. (2013). An Amperometric Glucose Biosensor Based on the Immobilization of Glucose Oxidase on the CuGeO Nanowire Modified Electrode Russian Journal of Electrochemistry, 49(6), 571–576.
  7. Ebrahimi, B., Shojaosadati, S.A., Daneshgar, P., Norouzi, P. & Mousavi S.M. (2011). Performance evaluation of fast Fourier-transform continuous cyclic-voltammetry pesticide biosensor Analytica Chimica Acta 687 168–176.
  8. Gouda, M. D., Thakur, M. S. & Karanth, N. G. (2001). Optimization of the multienzym system for sucrose biosensor by response surface methodology. World Journal of Microbiology & Biotechnology, 17, 595-600.
  9. Haighi, B., Nazari, L. & Sajjadi, S.M., (2012). Fabrication and application of a sensitive and highly stable copper hexacyanoferrate modified carbon ionic liquid paste electrode for hydrogen peroxide and glucose detection, Electroanalysis, 24, 2165-2175.
  10. Ivory, D.M., Miller, G.G., Sowa, J.M., Shacklette, L.W., Chance,R.R. & Baughman, R.H. (1979). Highly conducting chargetransfer complexes of poly(p-phenylene). The Journal of Chemical Physics, 71, 1506–1507.
  11. Kawai, T., Kuwabara, T., Wang, S. & Yoshino, K. (1990). Secondar battery characteristics poly(3-alkylthiophene). Japan